arXiv:2207.11009 · Scientific Reports

A digital alkali spin maser

Stuart J. Ingleby, Paul F. Griffin, T. Dyer, Marcin S. Mrozowski, Erling Riis

Published Jul 28, 2022• 22 citations
arXivHTMLPDF

Abstract

Self-oscillating atomic magnetometers, in which the precession of atomic spins in a magnetic field is driven by resonant modulation, offer high sensitivity and dynamic range. Phase-coherent feedback from the detected signal to the applied modulation creates a resonant spin maser system, highly responsive to changes in the background magnetic field. Here we show a system in which the phase condition for resonant precession is met by digital signal processing integrated into the maser feedback loop. This system uses a modest chip-scale laser and mass-produced dual-pass caesium vapour cell and operates in a 50 [Formula: see text]T field, making it a suitable technology for portable measurements of the geophysical magnetic field. We demonstrate a Cramér-Rao lower bound-limited resolution of 50 fT at 1 s sampling cadence, and a sensor bandwidth of 10 kHz. This device also represents an important class of atomic system in which low-latency digital processing forms an integral part of a coherently-driven quantum system.

DOI
10.1038/s41598-022-16910-z

Related papers

Subpicotesla scalar atomic magnetometer with a microfabricated cell

Rui Zhang, T. Dyer, N. L. Brockie +2 more

We demonstrated a scalar atomic magnetometer using a microfabricated Cs vapor cell. The atomic spin precession is driven by an amplitude-modulated circularly-polarized pump laser resonant on the D1 transition of Cs atoms and detected by an off-resonant linearly-polarized probe laser using a balanced polarimeter setup. Under a magnetic field with amplitude in the Earth's magnetic field range, the magnetometer in the gradiometer mode can reach sensitivities below 150fT/√Hz, which shows that the magnetometer by itself can achieve sub-100fT/√Hz sensitivities. In addition to its high sensitivity, the magnetometer has a bandwidth close to 1 kHz due to the broad magnetic resonance inside the small vapor cell. Our experiment suggests the feasibility of a portable low-power and high-performance magnetometer which can be operated in the Earth's magnetic field. Such a device will greatly reduce the restrictions on the operating environment and expand the range of applications for atomic magnetometers, such as detection of nuclear magnetic resonance in low magnetic fields

•Sep 28, 2019•1 citation

Subpicotesla scalar atomic magnetometer with a microfabricated cell

Rui Zhang, Terry Dyer, Nathan Brockie +2 more

We demonstrated a scalar atomic magnetometer using a microfabricated Cs vapor cell. The atomic spin precession is driven by an amplitude-modulated circularly-polarized pump laser resonant on the D1 transition of Cs atoms and detected by an off-resonant linearly-polarized probe laser using a balanced polarimeter setup. Under a magnetic field with amplitude in the Earth's magnetic field range, the magnetometer in the gradiometer mode can reach sensitivities below 150fT/Hz, which shows that the magnetometer by itself can achieve sub-100fT/Hz sensitivities. In addition to its high sensitivity, the magnetometer has a bandwidth close to 1 kHz due to the broad magnetic resonance inside the small vapor cell. Our experiment suggests the feasibility of a portable low-power and high-performance magnetometer which can be operated in the Earth's magnetic field. Such a device will greatly reduce the restrictions on the operating environment and expand the range of applications for atomic magnetometers, such as detection of nuclear magnetic resonance in low magnetic fields.

arXiv:1910.01768•Journal of Applied Physics•Sep 26, 2019•17 citations

Self-Driven Hybrid Atomic Spin Oscillator

Erwei Li, Qianjin Ma, Guobin Liu +2 more

A self-driven hybrid atomic spin oscillator is demonstrated in theory and experiment with a vapor $\mathrm{Rb}$-$\mathrm{Xe}$ dual-spin system. The raw signal of $\mathrm{Rb}$ spin oscillation is amplified, phase-shifted, and sent back to drive the $\mathrm{Xe}$ spins coherently. By fine-tuning the driving-field strength and phase, a self-sustaining spin-oscillation signal with zero frequency shift is obtained. The effective coherence time is infinitely prolonged beyond the intrinsic coherence time of $\mathrm{Xe}$ spins, forming a hybrid atomic spin oscillator. Spectral analysis indicates that a frequency resolution of 13.1 nHz is achieved, increasing the detection sensitivity for a magnetic field. Allan-deviation analysis shows that the spin oscillator can operate in continuous-wave mode like a spin maser. The prototype spin oscillator can be easily incorporated into other hybrid spin systems and can increase the detection sensitivity of alkali-metal--noble-gas comagnetometers.

arXiv:2301.12121•Physical Review Applied•Jul 14, 2023•3 citations

Continuous High-Sensitivity and High-Bandwidth Atomic Magnetometer

Rujie Li, Fred N. Baynes, Andre Nicholas Luiten +1 more

Here we demonstrate an atomic magnetometer that has a bandwidth of more than 100 kHz and a sensitivity of $180\phantom{\rule{0.2em}{0ex}}\mathrm{fT/}\ensuremath{\surd}\mathrm{Hz}$ at a Fourier frequency of 8 Hz, and $0.7\phantom{\rule{0.2em}{0ex}}\mathrm{nT/}\ensuremath{\surd}\mathrm{Hz}$ at 100 kHz. These sensitivity measurements are achieved at geophysically useful magnetic field magnitudes (approximately $50\phantom{\rule{0.2em}{0ex}}\ensuremath{\mu}\mathrm{T}$ inside a three-layer $\ensuremath{\mu}$-metal shield) and are limited by the photon shot noise for frequencies above 8 Hz. This device is based on a nonlinear magneto-optical rotation sensor that is operated with an active feedback to the pump modulation frequency. We present a theoretical description of the response functions of the components in the magnetometer. We show that the range of operation of atomic magnetometers can thus be expanded beyond the conventional high-sensitivity, low-bandwidth domain, to provide a high-linearity, high-bandwidth, and high-sensitivity sensor.

Physical Review Applied•Dec 23, 2020•75 citations

Remote detection magnetometry

Simon Rochester, James Higbie, Bob J. Patton +3 more

Introduction Shortly after the inception of atomic magnetometry, alkali-vapor magnetometers were being used to measure the Earth's magnetic field to unprecedented precision. During the same era, Bell and Bloom first demonstrated all-optical atomic magnetometry through synchronous optical pumping [1] (see Chapters 1 and 6). In this approach, optical-pumping light is frequency- or amplitude-modulated at harmonics of the Larmor frequency ω L to generate a precessing spin polarization within an alkali vapor at finite magnetic field [2, 3]. Although this technique received considerable attention from the atomic physics community for its applicability to optical pumping experiments, Earth's-field alkali-vapor atomic magnetometers continued to rely on radiofrequency (RF) field excitation for several decades (see Chapter 4). Upon the advent of diode lasers addressing alkali and metastable helium transitions, synchronously pumped magnetometers experienced a revival beginning in the late 1980s. In recent years, such magnetometers have found applications in nuclear magnetic resonance detection [4] (see also Chapter 14), quantum control experiments [5], and chip-scale devices intended for spacecraft use [6] (see also Chapters 7 and 15). All-optical magnetometers possess several advantages over devices that employ RF coils. RF-driven magnetometers can suffer from cross-talk if two sensors are placed in close proximity, since the AC magnetic field driving resonance in one vapor cell can adversely affect the other. All-optical magnetometers are free from such interference. When operated in self-oscillating mode [7], RF-driven magnetometers require an added ±90° electronic phase shift in the feedback loop to counter the intrinsic phase shift between the RF field and the probe-beam modulation.

Cambridge University Press eBooks•Mar 7, 2013•0 citations

A remotely interrogated all-optical 87Rb magnetometer

Brian Patton, Oscar O. Versolato, D. C. Hovde +3 more

Atomic magnetometry was performed at Earth’s magnetic field over a free-space distance of ten meters. Two laser beams aimed at a distant alkali-vapor cell excited and detected the 87Rb magnetic resonance, allowing the magnetic field within the cell to be interrogated remotely. Operated as a driven oscillator, the magnetometer measured the geomagnetic field with ≲3.5 pT precision in a ∼2 s data acquisition; this precision was likely limited by ambient field fluctuations. The sensor was also operated in self-oscillating mode with a 5.3 pT/Hz noise floor. Further optimization will yield a high-bandwidth, fully remote magnetometer with sub-pT sensitivity.

arXiv:1208.1236•Applied Physics Letters•Aug 20, 2012•45 citations
Self-oscillating atomic magnetometers, in which the precession of atomic spins in a magnetic field is driven by resonant modulation, offer high sensitivity and dynamic range. Phase-coherent feedback from the detected signal to the applied modulation creates a resonant spin maser system, highly responsive to changes in the background magnetic field. Here we show a system in which the phase condition for resonant precession is met by digital signal processing integrated into the maser feedback loop. This system uses a modest chip-scale laser and mass-produced dual-pass caesium vapour cell and operates in a 50 [Formula: see text]T field, making it a suitable technology for portable measurements of the geophysical magnetic field. We demonstrate a Cramér-Rao lower bound-limited resolution of 50 fT at 1 s sampling cadence, and a sensor bandwidth of 10 kHz. This device also represents an important class of atomic system in which low-latency digital processing forms an integral part of a coherently-driven quantum system.
We demonstrated a scalar atomic magnetometer using a microfabricated Cs vapor cell. The atomic spin precession is driven by an amplitude-modulated circularly-polarized pump laser resonant on the D1 transition of Cs atoms and detected by an off-resonant linearly-polarized probe laser using a balanced polarimeter setup. Under a magnetic field with amplitude in the Earth's magnetic field range, the magnetometer in the gradiometer mode can reach sensitivities below 150fT/√Hz, which shows that the magnetometer by itself can achieve sub-100fT/√Hz sensitivities. In addition to its high sensitivity, the magnetometer has a bandwidth close to 1 kHz due to the broad magnetic resonance inside the small vapor cell. Our experiment suggests the feasibility of a portable low-power and high-performance magnetometer which can be operated in the Earth's magnetic field. Such a device will greatly reduce the restrictions on the operating environment and expand the range of applications for atomic magnetometers, such as detection of nuclear magnetic resonance in low magnetic fields
We demonstrated a scalar atomic magnetometer using a microfabricated Cs vapor cell. The atomic spin precession is driven by an amplitude-modulated circularly-polarized pump laser resonant on the D1 transition of Cs atoms and detected by an off-resonant linearly-polarized probe laser using a balanced polarimeter setup. Under a magnetic field with amplitude in the Earth's magnetic field range, the magnetometer in the gradiometer mode can reach sensitivities below 150fT/Hz, which shows that the magnetometer by itself can achieve sub-100fT/Hz sensitivities. In addition to its high sensitivity, the magnetometer has a bandwidth close to 1 kHz due to the broad magnetic resonance inside the small vapor cell. Our experiment suggests the feasibility of a portable low-power and high-performance magnetometer which can be operated in the Earth's magnetic field. Such a device will greatly reduce the restrictions on the operating environment and expand the range of applications for atomic magnetometers, such as detection of nuclear magnetic resonance in low magnetic fields.
A self-driven hybrid atomic spin oscillator is demonstrated in theory and experiment with a vapor Rb\mathrm{Rb}Rb-Xe\mathrm{Xe}Xe dual-spin system. The raw signal of Rb\mathrm{Rb}Rb spin oscillation is amplified, phase-shifted, and sent back to drive the Xe\mathrm{Xe}Xe spins coherently. By fine-tuning the driving-field strength and phase, a self-sustaining spin-oscillation signal with zero frequency shift is obtained. The effective coherence time is infinitely prolonged beyond the intrinsic coherence time of Xe\mathrm{Xe}Xe spins, forming a hybrid atomic spin oscillator. Spectral analysis indicates that a frequency resolution of 13.1 nHz is achieved, increasing the detection sensitivity for a magnetic field. Allan-deviation analysis shows that the spin oscillator can operate in continuous-wave mode like a spin maser. The prototype spin oscillator can be easily incorporated into other hybrid spin systems and can increase the detection sensitivity of alkali-metal--noble-gas comagnetometers.
Here we demonstrate an atomic magnetometer that has a bandwidth of more than 100 kHz and a sensitivity of $180\phantom{\rule{0.2em}{0ex}}\mathrm{fT/}\ensuremath{\surd}\mathrm{Hz}$ at a Fourier frequency of 8 Hz, and $0.7\phantom{\rule{0.2em}{0ex}}\mathrm{nT/}\ensuremath{\surd}\mathrm{Hz}$ at 100 kHz. These sensitivity measurements are achieved at geophysically useful magnetic field magnitudes (approximately $50\phantom{\rule{0.2em}{0ex}}\ensuremath{\mu}\mathrm{T}$ inside a three-layer $\ensuremath{\mu}$-metal shield) and are limited by the photon shot noise for frequencies above 8 Hz. This device is based on a nonlinear magneto-optical rotation sensor that is operated with an active feedback to the pump modulation frequency. We present a theoretical description of the response functions of the components in the magnetometer. We show that the range of operation of atomic magnetometers can thus be expanded beyond the conventional high-sensitivity, low-bandwidth domain, to provide a high-linearity, high-bandwidth, and high-sensitivity sensor.
Introduction Shortly after the inception of atomic magnetometry, alkali-vapor magnetometers were being used to measure the Earth's magnetic field to unprecedented precision. During the same era, Bell and Bloom first demonstrated all-optical atomic magnetometry through synchronous optical pumping [1] (see Chapters 1 and 6). In this approach, optical-pumping light is frequency- or amplitude-modulated at harmonics of the Larmor frequency ω L to generate a precessing spin polarization within an alkali vapor at finite magnetic field [2, 3]. Although this technique received considerable attention from the atomic physics community for its applicability to optical pumping experiments, Earth's-field alkali-vapor atomic magnetometers continued to rely on radiofrequency (RF) field excitation for several decades (see Chapter 4). Upon the advent of diode lasers addressing alkali and metastable helium transitions, synchronously pumped magnetometers experienced a revival beginning in the late 1980s. In recent years, such magnetometers have found applications in nuclear magnetic resonance detection [4] (see also Chapter 14), quantum control experiments [5], and chip-scale devices intended for spacecraft use [6] (see also Chapters 7 and 15). All-optical magnetometers possess several advantages over devices that employ RF coils. RF-driven magnetometers can suffer from cross-talk if two sensors are placed in close proximity, since the AC magnetic field driving resonance in one vapor cell can adversely affect the other. All-optical magnetometers are free from such interference. When operated in self-oscillating mode [7], RF-driven magnetometers require an added ±90° electronic phase shift in the feedback loop to counter the intrinsic phase shift between the RF field and the probe-beam modulation.
Atomic magnetometry was performed at Earth’s magnetic field over a free-space distance of ten meters. Two laser beams aimed at a distant alkali-vapor cell excited and detected the 87Rb magnetic resonance, allowing the magnetic field within the cell to be interrogated remotely. Operated as a driven oscillator, the magnetometer measured the geomagnetic field with ≲3.5 pT precision in a ∼2 s data acquisition; this precision was likely limited by ambient field fluctuations. The sensor was also operated in self-oscillating mode with a 5.3 pT/Hz noise floor. Further optimization will yield a high-bandwidth, fully remote magnetometer with sub-pT sensitivity.