arXiv:2008.05601 · Review of Scientific Instruments

Ultra-low-vibration closed-cycle cryogenic surface-electrode ion trap apparatus

T. Dubielzig, S. Halama, H. Hahn, G. Zarantonello, M. Niemann, A. Bautista-Salvador, C. Ospelkaus

Published Apr 1, 2021• 35 citations
arXivHTMLPDF

Abstract

We describe the design, commissioning, and operation of an ultra-low-vibration closed-cycle cryogenic ion trap apparatus. One hundred lines for low-frequency signals and eight microwave/radio frequency coaxial feed-lines offer the possibility of implementing a small-scale ion-trap quantum processor or simulator. With all supply cables attached, more than 1.3 W of cooling power at 5 K is still available for absorbing energy from electrical pulses introduced to control ions. The trap itself is isolated from vibrations induced by the cold head using a helium exchange gas interface. The performance of the vibration isolation system has been characterized using a Michelson interferometer, finding residual vibration amplitudes on the order of 10 nm rms. Trapping of 9Be+ ions has been demonstrated using a combination of laser ablation and photoionization.

DOI
10.1063/5.0024423

Related papers

Ultra-low vibration closed-cycle cryogenic surface-electrode ion trap apparatus

T. Dubielzig

Trapped ions are one of the leading platforms for building scalable quantum information processors. Current ion traps run on less than 100 qubits. A useful quantum computer needs to operate thousands or even millions of qubits. If systems reach gate fidelities that allow for quantum error correction, gate errors below the physical fidelity can be reached, which is important for scaling. The most promising platform for scaling up ion-based quantum computers are surface electrode traps in connection with the so-called QCCD (Quantum Charge Coupled Device) architecture. In an evolution of the standard (laser-based) approach to quantum logic gates, gates can be performed by applying an oscillating magnetic near-field gradient to the ions. Current fidelities are about one order of magnitude away from reaching the fault tolerant threshold with no fundamental limitations in sight. Scaling QCCDs is impeded by heating rates of the ions as well as vacuum quality. Ions have to be stored for at least as long as it takes for an algorithm to run and subsequent readout. Collisions with background gas and heating can expel ions from the trap. The microwave near-field approach of driving gates benefits from ions that are trapped close to the trap surface. Unfortunately, heating becomes more prominent near the surface. A cryogenic environment reduces the background pressure by several orders of magnitude and the kinetic energy of gas molecules by two orders of magnitude. Heating rates in a cryogenic trap are about two orders of magnitude lower than in a room temperature trap with otherwise equal physical properties. To operate a QCCD processor, it is required to employ phase-stable Raman laser beams. The cryogenic design of the apparatus needs to take into account the resulting requirements for ultra-low vibration amplitudes. This is particularly relevant for closed-cycle cooling systems, which are desirable from an economic point of view compared to liquid cryostats that boil off helium, but which also feature moving mechanical parts that can introduce a serious amount of vibrations. Here, we present a vibration isolated closed cycle Gifford-McMahon cryocooler with record low vibration amplitudes of 29 nm (RMS=7.8 nm) in the vertical and 51 nm (RMS=13.5 nm) in the horizontal direction. It contains a self-sustained inner vacuum chamber that houses the trap. It features 100 DC connections, 10 of which are able to carry up to 1 A of current and 8 high frequency lines. At 5 K, we can apply continuous 3.4 W of microwave power continuously at 1 GHz to the trap electrodes without heating the system. This is about 5.8 times more than state of the art near-field gates require. We also present the laser systems required to for single-shot ablation loading and Doppler cooling of Be ions: a 70 mW 313 nm Dopper cooling laser, a 1064 nm ns-pulse laser for ablation of neutral beryllium atoms and a 235 nm laser for ionizing them. The detection system features an in vacuum imaging system on a cryogenic 3D translation stage that is able to scan and image the complete surface of an ion trap. The light can be detected on a photomultiplier tube or spatially resolved on an EMCCD camera. Our trap features a meander structure for driving gates and can trap up to 5 ions of which we, so far, did not lose any due to background gas collisions. This hints at a vacuum pressure of 1e-12 mbar or lower. The demonstrated system could hold a cryogenic QCCD chip with a few 10 qubits

•Jan 1, 2021•25 citations

Cryogenic ion trapping systems with surface-electrode traps

Paul Antohi, David Schuster, Gleb M. Akselrod +5 more

We present two simple cryogenic rf ion trap systems in which cryogenic temperatures and ultra high vacuum pressures can be reached in as little as 12 h. The ion traps are operated either in a liquid helium bath cryostat or in a low vibration closed cycle cryostat. The fast turn around time and availability of buffer gas cooling made the systems ideal for testing surface-electrode ion traps. The vibration amplitude of the closed cycled cryostat was found to be below 106 nm. We evaluated the systems by loading surface-electrode ion traps with (88)Sr(+) ions using laser ablation, which is compatible with the cryogenic environment. Using Doppler cooling we observed small ion crystals in which optically resolved ions have a trapped lifetime over 2500 min.

arXiv:0807.4763•Review of Scientific Instruments•Jan 1, 2009•47 citations

Stability of ion chains in a cryogenic surface-electrode ion trap

Grahame Vittorini

Cold, trapped atomic ions have enabled the investigation of fundamental physics and generated a rich field of applications. Foremost among these is quantum computation which has recently driven the development of the sophisticated, scalable surface-electrode trap. Despite the many advantages of surface-electrode traps, the typically smaller ion-electrode distance, d, in these traps results in an increased ion heating rate that is proportional to d^(-4) and a decreased trap well-depth that is proportional to d^(-2). These shortcomings can be simultaneously addressed by installing the trap into a cryogenic environment. With this in mind, a closed-cycle, cryogenic ion trapping apparatus that maintains excellent vacuum, is highly modular, has increased optical access, and uses a simple vibration isolation system has been developed. \n \nSingle ions are trapped and used to characterize system properties such as the motion of the vibration isolation stage. In order to compare this system to a similar room temperature apparatus, the ion trapping lifetime and heating rate are determined. A single ion also serves as a sensitive electric field probe that is used to measure and compensate stray electric fields across the trap. \n \nDue to the long dark ion lifetimes in this system, it is well-suited to probing the stability of small, linear ion crystals. Linear ion crystals of arbitrary length are built in an automated fashion using transport waveforms and the scaling of dark lifetime with ion number for N <= 6 is investigated. These data are then used to consider the relevance of various loss channels.

Bulletin of the American Physical Society•Sep 10, 2013•1 citation

High-Stability Cryogenic System for Quantum Computing With Compact Packaged Ion Traps

Robert Spivey, I. V. Inlek, Zhubing Jia +9 more

Cryogenic environments benefit ion trapping experiments by offering lower motional heating rates, collision energies, and an ultrahigh vacuum (UHV) environment for maintaining long ion chains for extended periods of time. Mechanical vibrations caused by compressors in closed-cycle cryostats can introduce relative motion between the ion and the wavefronts of lasers used to manipulate the ions. Here, we present a novel ion trapping system where a commercial low-vibration closed-cycle cryostat is used in a custom monolithic enclosure. We measure mechanical vibrations of the sample stage using an optical interferometer, and observe a root-mean-square relative displacement of 2.4 nm and a peak-to-peak displacement of 17 nm between free-space beams and the trapping location. We packaged a surface ion trap in a cryopackage assembly that enables easy handling while creating a UHV environment for the ions. The trap cryopackage contains activated carbon getter material for enhanced sorption pumping near the trapping location, and source material for ablation loading. Using$^{171}$Yb$^{+}$as our ion, we estimate the operating pressure of the trap as a function of package temperature using phase transitions of zig-zag ion chains as a probe. We measured the radial mode heating rate of a single ion to be 13 quanta/s on average. The Ramsey coherence measurements yield 330-ms coherence time for counter-propagating Raman carrier transitions using a 355-nm mode-locked pulse laser, demonstrating the high optical stability.

arXiv:2108.05290•IEEE Transactions on Quantum Engineering•Nov 8, 2021•33 citations

Evaluation and Suppression of Vibrational Noise for Cryogenic Surface Electrode Ion Traps

Qingqing Qin, Bao-Quan Ou, Wei Wu +4 more

Abstract Cryogenic ion traps offer substantial advantages over their room-temperature counterparts, particularly in terms of stable ion confinement and cooling. However, this technological advancement has introduced a significant challenge, namely the generation of additional vibrational noise. This noise is pronounced during the compression and expansion phases of the Gifford–McMahon cycle refrigerator and stems from the mechanical operation of the cold head. The residual vibrational noise can modulate the Rabi oscillation frequency, thereby compromising the fidelity of the quantum logic gate. In this study, we quantitatively assess the impact of residual vibrational noise and introduce an innovative strategy for its mitigation. This strategy was designed to reduce superimposed potential fluctuations within a cryogenic surface electrode ion trap. The measured residual vibrational noise was used for real-time feedback control. Addressing the inverse determination of the compensating potential for a stable superimposed trapping potential is an inherently ill-posed problem considering known measurement noise results. By applying quadratic programming, we numerically calculated the optimal time series for the compensation potential. An ensemble of these compensation voltages was applied in real time to the static compensation electrodes. Although this method is in the theoretical stage, we believe that it can effectively suppress the translation induced by vibrational noise and foster the creation of a stable superimposed harmonic potential. The deployment of this technique is expected to substantially improve the accuracy of quantum operations, thereby fueling the evolution of quantum technologies.

Chinese Physics Letters•Feb 12, 2025•0 citations

A Cryogenic Linear Paul Trap for Quantum Simulation

Paul Hess, Harvey Kaplan, Aaron Lee +5 more

statistics. We present progress towards the construction of a cryogenic ion trap apparatus, designed to cryopump background gas within the 4K chamber. The resulting reduction in pressure will allow robust trapping of up to 100 ions in a single chain. Cooling is provided by a closed cycle cryostat with a gas mediated thermal linkage which mechanically decouples the ion trap from the vibrating cold head. A spherical octagon surrounding the ion trap allows optical access for global and individual addressing beams and high numerical aperture uorescence collection.

Bulletin of the American Physical Society•Jun 11, 2015•0 citations
We describe the design, commissioning, and operation of an ultra-low-vibration closed-cycle cryogenic ion trap apparatus. One hundred lines for low-frequency signals and eight microwave/radio frequency coaxial feed-lines offer the possibility of implementing a small-scale ion-trap quantum processor or simulator. With all supply cables attached, more than 1.3 W of cooling power at 5 K is still available for absorbing energy from electrical pulses introduced to control ions. The trap itself is isolated from vibrations induced by the cold head using a helium exchange gas interface. The performance of the vibration isolation system has been characterized using a Michelson interferometer, finding residual vibration amplitudes on the order of 10 nm rms. Trapping of 9Be+ ions has been demonstrated using a combination of laser ablation and photoionization.
Trapped ions are one of the leading platforms for building scalable quantum information processors. Current ion traps run on less than 100 qubits. A useful quantum computer needs to operate thousands or even millions of qubits. If systems reach gate fidelities that allow for quantum error correction, gate errors below the physical fidelity can be reached, which is important for scaling. The most promising platform for scaling up ion-based quantum computers are surface electrode traps in connection with the so-called QCCD (Quantum Charge Coupled Device) architecture. In an evolution of the standard (laser-based) approach to quantum logic gates, gates can be performed by applying an oscillating magnetic near-field gradient to the ions. Current fidelities are about one order of magnitude away from reaching the fault tolerant threshold with no fundamental limitations in sight. Scaling QCCDs is impeded by heating rates of the ions as well as vacuum quality. Ions have to be stored for at least as long as it takes for an algorithm to run and subsequent readout. Collisions with background gas and heating can expel ions from the trap. The microwave near-field approach of driving gates benefits from ions that are trapped close to the trap surface. Unfortunately, heating becomes more prominent near the surface. A cryogenic environment reduces the background pressure by several orders of magnitude and the kinetic energy of gas molecules by two orders of magnitude. Heating rates in a cryogenic trap are about two orders of magnitude lower than in a room temperature trap with otherwise equal physical properties. To operate a QCCD processor, it is required to employ phase-stable Raman laser beams. The cryogenic design of the apparatus needs to take into account the resulting requirements for ultra-low vibration amplitudes. This is particularly relevant for closed-cycle cooling systems, which are desirable from an economic point of view compared to liquid cryostats that boil off helium, but which also feature moving mechanical parts that can introduce a serious amount of vibrations. Here, we present a vibration isolated closed cycle Gifford-McMahon cryocooler with record low vibration amplitudes of 29 nm (RMS=7.8 nm) in the vertical and 51 nm (RMS=13.5 nm) in the horizontal direction. It contains a self-sustained inner vacuum chamber that houses the trap. It features 100 DC connections, 10 of which are able to carry up to 1 A of current and 8 high frequency lines. At 5 K, we can apply continuous 3.4 W of microwave power continuously at 1 GHz to the trap electrodes without heating the system. This is about 5.8 times more than state of the art near-field gates require. We also present the laser systems required to for single-shot ablation loading and Doppler cooling of Be ions: a 70 mW 313 nm Dopper cooling laser, a 1064 nm ns-pulse laser for ablation of neutral beryllium atoms and a 235 nm laser for ionizing them. The detection system features an in vacuum imaging system on a cryogenic 3D translation stage that is able to scan and image the complete surface of an ion trap. The light can be detected on a photomultiplier tube or spatially resolved on an EMCCD camera. Our trap features a meander structure for driving gates and can trap up to 5 ions of which we, so far, did not lose any due to background gas collisions. This hints at a vacuum pressure of 1e-12 mbar or lower. The demonstrated system could hold a cryogenic QCCD chip with a few 10 qubits
We present two simple cryogenic rf ion trap systems in which cryogenic temperatures and ultra high vacuum pressures can be reached in as little as 12 h. The ion traps are operated either in a liquid helium bath cryostat or in a low vibration closed cycle cryostat. The fast turn around time and availability of buffer gas cooling made the systems ideal for testing surface-electrode ion traps. The vibration amplitude of the closed cycled cryostat was found to be below 106 nm. We evaluated the systems by loading surface-electrode ion traps with (88)Sr(+) ions using laser ablation, which is compatible with the cryogenic environment. Using Doppler cooling we observed small ion crystals in which optically resolved ions have a trapped lifetime over 2500 min.
Cold, trapped atomic ions have enabled the investigation of fundamental physics and generated a rich field of applications. Foremost among these is quantum computation which has recently driven the development of the sophisticated, scalable surface-electrode trap. Despite the many advantages of surface-electrode traps, the typically smaller ion-electrode distance, d, in these traps results in an increased ion heating rate that is proportional to d^(-4) and a decreased trap well-depth that is proportional to d^(-2). These shortcomings can be simultaneously addressed by installing the trap into a cryogenic environment. With this in mind, a closed-cycle, cryogenic ion trapping apparatus that maintains excellent vacuum, is highly modular, has increased optical access, and uses a simple vibration isolation system has been developed. \n \nSingle ions are trapped and used to characterize system properties such as the motion of the vibration isolation stage. In order to compare this system to a similar room temperature apparatus, the ion trapping lifetime and heating rate are determined. A single ion also serves as a sensitive electric field probe that is used to measure and compensate stray electric fields across the trap. \n \nDue to the long dark ion lifetimes in this system, it is well-suited to probing the stability of small, linear ion crystals. Linear ion crystals of arbitrary length are built in an automated fashion using transport waveforms and the scaling of dark lifetime with ion number for N <= 6 is investigated. These data are then used to consider the relevance of various loss channels.
Cryogenic environments benefit ion trapping experiments by offering lower motional heating rates, collision energies, and an ultrahigh vacuum (UHV) environment for maintaining long ion chains for extended periods of time. Mechanical vibrations caused by compressors in closed-cycle cryostats can introduce relative motion between the ion and the wavefronts of lasers used to manipulate the ions. Here, we present a novel ion trapping system where a commercial low-vibration closed-cycle cryostat is used in a custom monolithic enclosure. We measure mechanical vibrations of the sample stage using an optical interferometer, and observe a root-mean-square relative displacement of 2.4 nm and a peak-to-peak displacement of 17 nm between free-space beams and the trapping location. We packaged a surface ion trap in a cryopackage assembly that enables easy handling while creating a UHV environment for the ions. The trap cryopackage contains activated carbon getter material for enhanced sorption pumping near the trapping location, and source material for ablation loading. Using171^{171}171Yb+^{+}+as our ion, we estimate the operating pressure of the trap as a function of package temperature using phase transitions of zig-zag ion chains as a probe. We measured the radial mode heating rate of a single ion to be 13 quanta/s on average. The Ramsey coherence measurements yield 330-ms coherence time for counter-propagating Raman carrier transitions using a 355-nm mode-locked pulse laser, demonstrating the high optical stability.
Abstract Cryogenic ion traps offer substantial advantages over their room-temperature counterparts, particularly in terms of stable ion confinement and cooling. However, this technological advancement has introduced a significant challenge, namely the generation of additional vibrational noise. This noise is pronounced during the compression and expansion phases of the Gifford–McMahon cycle refrigerator and stems from the mechanical operation of the cold head. The residual vibrational noise can modulate the Rabi oscillation frequency, thereby compromising the fidelity of the quantum logic gate. In this study, we quantitatively assess the impact of residual vibrational noise and introduce an innovative strategy for its mitigation. This strategy was designed to reduce superimposed potential fluctuations within a cryogenic surface electrode ion trap. The measured residual vibrational noise was used for real-time feedback control. Addressing the inverse determination of the compensating potential for a stable superimposed trapping potential is an inherently ill-posed problem considering known measurement noise results. By applying quadratic programming, we numerically calculated the optimal time series for the compensation potential. An ensemble of these compensation voltages was applied in real time to the static compensation electrodes. Although this method is in the theoretical stage, we believe that it can effectively suppress the translation induced by vibrational noise and foster the creation of a stable superimposed harmonic potential. The deployment of this technique is expected to substantially improve the accuracy of quantum operations, thereby fueling the evolution of quantum technologies.
statistics. We present progress towards the construction of a cryogenic ion trap apparatus, designed to cryopump background gas within the 4K chamber. The resulting reduction in pressure will allow robust trapping of up to 100 ions in a single chain. Cooling is provided by a closed cycle cryostat with a gas mediated thermal linkage which mechanically decouples the ion trap from the vibrating cold head. A spherical octagon surrounding the ion trap allows optical access for global and individual addressing beams and high numerical aperture uorescence collection.