arXiv:1911.13146

Dynamic decoupling of laser phase noise in compound atomic clocks

Sören Dörscher, Ali Al-Masoudi, Marcin Bober, R. Schwarz, Richard Hobson, Uwe Sterr, Christian Lisdat

Published Nov 29, 2019• 0 citations
arXivHTMLPDF

Abstract

The frequency stability achieved by an optical atomic clock ultimately depends on the coherence of its local oscillator. Even the best ultrastable lasers only allow interrogation times of a few seconds, at present. Here we present a universal measurement protocol that overcomes this limitation. Engineered dynamic decoupling of laser phase noise allows any optical atomic clock with high signal-to-noise ratio in a single interrogation to reconstruct the laser's phase well beyond its coherence limit. A compound clock is then formed in combination with another optical clock of any type, allowing the latter to achieve significantly higher frequency stability than on its own. We demonstrate implementation of the protocol in a realistic proof-of-principle experiment with a phase reconstruction fidelity of 99 %. The protocol enables minute-long interrogation for the best ultrastable laser systems. Likewise, it can improve clock performance where less stable local oscillators are used, such as in transortable systems.

DOI
10.48550/arxiv.1911.13146

Related papers

Dynamical decoupling of laser phase noise in compound atomic clocks

Sören Dörscher, Ali Al-Masoudi, Marcin Bober +4 more

Abstract The frequency stability of many optical atomic clocks is limited by the coherence of their local oscillator. Here, we present a measurement protocol that overcomes the laser coherence limit. It relies on engineered dynamical decoupling of laser phase noise and near-synchronous interrogation of two clocks. One clock coarsely tracks the laser phase using dynamical decoupling; the other refines this estimate using a high-resolution phase measurement. While the former needs to have a high signal-to-noise ratio, the latter clock may operate with any number of particles. The protocol effectively enables minute-long Ramsey interrogation for coherence times of few seconds as provided by the current best ultrastable laser systems. We demonstrate implementation of the protocol in a realistic proof-of-principle experiment, where we interrogate for 0.5 s at a laser coherence time of 77 ms. Here, a single lattice clock is used to emulate synchronous interrogation of two separate clocks in the presence of artificial laser frequency noise. We discuss the frequency instability of a single-ion clock that would result from using the protocol for stabilisation, under these conditions and for minute-long interrogation, and find expected instabilities of σy(τ) = 8 × 10−16(τ/s)−1/2 and σy(τ) = 5 × 10−17(τ/s)−1/2, respectively.

Communications Physics•Oct 20, 2020•23 citations

Leveraging correlated laser noise to improve optical frequency comparisons

Ethan Clements, M. Bohman, May E. Kim +14 more

Laser noise usually limits the stability of optical frequency ratio measurements, limiting the speed and precision one can compare two atomic frequency standards. In this talk I will describe two methods, correlation and differential spectroscopy, which utilize correlations in laser noise to increase the achievable interrogation time and thus increase the frequency comparison stability. Correlation spectroscopy is a technique which uses a parity measurement following a synchronized Ramsey interrogation to measure the relative frequency of two similar frequency atomic clocks. With this technique we achieve a measurement instability of (4×10^(-16))⁄√(τ⁄s) for a comparison of two single 27Al+ ion clocks. Differential spectroscopy uses an atomic clock with low projection noise, here a 171Yb lattice clock, to correct the phase noise of a second, higher frequency clock’s local oscillator thereby reducing the measurement instability to the level of the first. This can be further extended using two lattice clocks in a zero dead time configuration to correct the phase noise beyond the interrogation time reachable for a single Yb lattice clock. With these techniques we achieve measurement stabilities of (2.5×10^(-16))⁄√(τ⁄s) and (2×10^(-16))⁄√(τ⁄s) for a comparison between a single 27Al+ ion clock and a 171Yb lattice clock running as single clock and in a zero dead time configuration respectively. In addition to these techniques, I will also discuss recent progress towards characterizing the systematics of the NIST 40Ca+/27Al+ optical atomic clock.

Optical and Quantum Sensing and Precision Metrology II•Mar 9, 2022•0 citations

Probing beyond the laser coherence time in optical clock comparisons

David B. Hume, David R. Leibrandt

We develop differential measurement protocols that circumvent the laser noise limit in the stability of optical clock comparisons by synchronous probing of two clocks using phase-locked local oscillators. This allows for probe times longer than the laser coherence time, avoids the Dick effect, and supports Heisenberg-limited measurement precision. We present protocols for such frequency comparisons and develop numerical simulations of the protocols with realistic noise sources. These methods provide a route to reduce frequency ratio measurement durations by more than an order of magnitude.

arXiv:1508.05129•Physical Review A•Mar 30, 2016•43 citations

Improved absolute clock stability by the joint interrogation of two atomic ensembles

Weidong Li, Shuyuan Wu, Augusto Smerzi +1 more

Improving the clock stability is of fundamental importance for the development of quantum-enhanced metrology. One of the main limitations arises from the randomly fluctuating local oscillator (LO) frequency, which introduces ``phase slips'' for long interrogation times and hence the failure of the frequency-feedback loop. Here we propose a strategy to improve the stability of atomic clocks by interrogating two atomic ensembles sharing the same LO. The two ensembles are prepared in coherent spin states pointing along orthogonal directions in the Bloch sphere. While standard Ramsey interrogation can only determine phases unambiguously in the interval $[\ensuremath{-}\ensuremath{\pi}/2,\ensuremath{\pi}/2]$, the joint interrogation allows for an extension to $[\ensuremath{-}\ensuremath{\pi},\ensuremath{\pi}]$, resulting in a relaxed restriction of the Ramsey time and improvement of absolute clock stability. Theoretical predictions are supported by ab initio numerical simulation for white and correlated LO noise. While our basic protocol uses uncorrelated atoms, we further extended it to include spin-squeezing and further improving the scaling of clock stability with the number of atoms. Our protocol can be readily tested in current state-of-the-art experiments.

arXiv:2104.14309•Physical Review A•May 19, 2022•8 citations

Reducing the Instability of an Optical Lattice Clock Using Multiple Atomic Ensembles

Xin Zheng, Jonathan Dolde, Shimon Kolkowitz

The stability of an optical atomic clock is a critical figure of merit for almost all clock applications. To this end, much optical atomic clock research has focused on reducing clock instability by increasing the atom number, lengthening the coherent interrogation times, and introducing entanglement to push beyond the standard quantum limit. In this work, we experimentally demonstrate an alternative approach to reducing clock instability using a phase estimation approach based on individually controlled atomic ensembles in a strontium (Sr) optical lattice clock. We first demonstrate joint Ramsey interrogation of two spatially resolved atom ensembles that are out of phase with respect to each other, which we call “quadrature Ramsey spectroscopy,” resulting in a factor of 1.36(5) reduction in absolute clock instability as measured with interleaved self-comparisons. We then leverage the rich hyperfine structure of Sr87 to realize independent coherent control over multiple ensembles with only global laser addressing. Finally, we utilize this independent control over four atom ensembles to implement a form of phase estimation, achieving a factor of greater than 3 enhancement in coherent interrogation time and a factor of 2.08(6) reduction in instability over an otherwise identical single-ensemble clock with the same local oscillator and the same number of atoms. We expect that multiensemble protocols similar to those demonstrated here will result in reduction in the instability of any optical lattice clock with an interrogation time limited by the local oscillator. Published by the American Physical Society 2024

arXiv:2305.12315•Physical Review X•Jan 23, 2024•17 citations

Optical atomic clocks and the quest for ultrastable lasers

F. Riehle

Optical atomic clocks with fractional uncertainties around and below 10−17 [1] and fractional instabilities at one second interrogation time of below 10−15 [2] have outperformed the best microwave clocks like the cesium atomic clock. In optical atomic clocks a highly stable laser is used to interrogate optical transitions in laser-cooled atoms, ions, or molecules. The increase in accuracy and stability of the optical atomic clocks results from the increased line quality factor Q= ν/Δν given by the five orders of magnitude higher frequency of optical transitions as compared to the transition relevant in the microwave domain. To utilize this potential it is furthermore necessary to realize a suitable narrow linewidth of the interrogated line. The linewidth of any optical transition is given by the different broadening mechanisms of the respective quantum transition relevant for particular interrogation process as well as by the coherence properties of the interrogating laser.

Frontiers in Optics•Jan 1, 2013•1 citation
The frequency stability achieved by an optical atomic clock ultimately depends on the coherence of its local oscillator. Even the best ultrastable lasers only allow interrogation times of a few seconds, at present. Here we present a universal measurement protocol that overcomes this limitation. Engineered dynamic decoupling of laser phase noise allows any optical atomic clock with high signal-to-noise ratio in a single interrogation to reconstruct the laser's phase well beyond its coherence limit. A compound clock is then formed in combination with another optical clock of any type, allowing the latter to achieve significantly higher frequency stability than on its own. We demonstrate implementation of the protocol in a realistic proof-of-principle experiment with a phase reconstruction fidelity of 99 %. The protocol enables minute-long interrogation for the best ultrastable laser systems. Likewise, it can improve clock performance where less stable local oscillators are used, such as in transortable systems.
Abstract The frequency stability of many optical atomic clocks is limited by the coherence of their local oscillator. Here, we present a measurement protocol that overcomes the laser coherence limit. It relies on engineered dynamical decoupling of laser phase noise and near-synchronous interrogation of two clocks. One clock coarsely tracks the laser phase using dynamical decoupling; the other refines this estimate using a high-resolution phase measurement. While the former needs to have a high signal-to-noise ratio, the latter clock may operate with any number of particles. The protocol effectively enables minute-long Ramsey interrogation for coherence times of few seconds as provided by the current best ultrastable laser systems. We demonstrate implementation of the protocol in a realistic proof-of-principle experiment, where we interrogate for 0.5 s at a laser coherence time of 77 ms. Here, a single lattice clock is used to emulate synchronous interrogation of two separate clocks in the presence of artificial laser frequency noise. We discuss the frequency instability of a single-ion clock that would result from using the protocol for stabilisation, under these conditions and for minute-long interrogation, and find expected instabilities of σy(τ) = 8 × 10−16(τ/s)−1/2 and σy(τ) = 5 × 10−17(τ/s)−1/2, respectively.
Laser noise usually limits the stability of optical frequency ratio measurements, limiting the speed and precision one can compare two atomic frequency standards. In this talk I will describe two methods, correlation and differential spectroscopy, which utilize correlations in laser noise to increase the achievable interrogation time and thus increase the frequency comparison stability. Correlation spectroscopy is a technique which uses a parity measurement following a synchronized Ramsey interrogation to measure the relative frequency of two similar frequency atomic clocks. With this technique we achieve a measurement instability of (4×10^(-16))⁄√(τ⁄s) for a comparison of two single 27Al+ ion clocks. Differential spectroscopy uses an atomic clock with low projection noise, here a 171Yb lattice clock, to correct the phase noise of a second, higher frequency clock’s local oscillator thereby reducing the measurement instability to the level of the first. This can be further extended using two lattice clocks in a zero dead time configuration to correct the phase noise beyond the interrogation time reachable for a single Yb lattice clock. With these techniques we achieve measurement stabilities of (2.5×10^(-16))⁄√(τ⁄s) and (2×10^(-16))⁄√(τ⁄s) for a comparison between a single 27Al+ ion clock and a 171Yb lattice clock running as single clock and in a zero dead time configuration respectively. In addition to these techniques, I will also discuss recent progress towards characterizing the systematics of the NIST 40Ca+/27Al+ optical atomic clock.
We develop differential measurement protocols that circumvent the laser noise limit in the stability of optical clock comparisons by synchronous probing of two clocks using phase-locked local oscillators. This allows for probe times longer than the laser coherence time, avoids the Dick effect, and supports Heisenberg-limited measurement precision. We present protocols for such frequency comparisons and develop numerical simulations of the protocols with realistic noise sources. These methods provide a route to reduce frequency ratio measurement durations by more than an order of magnitude.
Improving the clock stability is of fundamental importance for the development of quantum-enhanced metrology. One of the main limitations arises from the randomly fluctuating local oscillator (LO) frequency, which introduces ``phase slips'' for long interrogation times and hence the failure of the frequency-feedback loop. Here we propose a strategy to improve the stability of atomic clocks by interrogating two atomic ensembles sharing the same LO. The two ensembles are prepared in coherent spin states pointing along orthogonal directions in the Bloch sphere. While standard Ramsey interrogation can only determine phases unambiguously in the interval $[\ensuremath{-}\ensuremath{\pi}/2,\ensuremath{\pi}/2]$, the joint interrogation allows for an extension to $[\ensuremath{-}\ensuremath{\pi},\ensuremath{\pi}]$, resulting in a relaxed restriction of the Ramsey time and improvement of absolute clock stability. Theoretical predictions are supported by ab initio numerical simulation for white and correlated LO noise. While our basic protocol uses uncorrelated atoms, we further extended it to include spin-squeezing and further improving the scaling of clock stability with the number of atoms. Our protocol can be readily tested in current state-of-the-art experiments.
The stability of an optical atomic clock is a critical figure of merit for almost all clock applications. To this end, much optical atomic clock research has focused on reducing clock instability by increasing the atom number, lengthening the coherent interrogation times, and introducing entanglement to push beyond the standard quantum limit. In this work, we experimentally demonstrate an alternative approach to reducing clock instability using a phase estimation approach based on individually controlled atomic ensembles in a strontium (Sr) optical lattice clock. We first demonstrate joint Ramsey interrogation of two spatially resolved atom ensembles that are out of phase with respect to each other, which we call “quadrature Ramsey spectroscopy,” resulting in a factor of 1.36(5) reduction in absolute clock instability as measured with interleaved self-comparisons. We then leverage the rich hyperfine structure of Sr87 to realize independent coherent control over multiple ensembles with only global laser addressing. Finally, we utilize this independent control over four atom ensembles to implement a form of phase estimation, achieving a factor of greater than 3 enhancement in coherent interrogation time and a factor of 2.08(6) reduction in instability over an otherwise identical single-ensemble clock with the same local oscillator and the same number of atoms. We expect that multiensemble protocols similar to those demonstrated here will result in reduction in the instability of any optical lattice clock with an interrogation time limited by the local oscillator. Published by the American Physical Society 2024
Optical atomic clocks with fractional uncertainties around and below 10−17 [1] and fractional instabilities at one second interrogation time of below 10−15 [2] have outperformed the best microwave clocks like the cesium atomic clock. In optical atomic clocks a highly stable laser is used to interrogate optical transitions in laser-cooled atoms, ions, or molecules. The increase in accuracy and stability of the optical atomic clocks results from the increased line quality factor Q= ν/Δν given by the five orders of magnitude higher frequency of optical transitions as compared to the transition relevant in the microwave domain. To utilize this potential it is furthermore necessary to realize a suitable narrow linewidth of the interrogated line. The linewidth of any optical transition is given by the different broadening mechanisms of the respective quantum transition relevant for particular interrogation process as well as by the coherence properties of the interrogating laser.