A measurement method for responsivity of microwave kinetic inductance detector by changing power of readout microwaves

Hiroki Kutsuma, M. Hattori, Ryo Koyano, S. Mima, S. Oguri, Chiko Otani, Tohru Taino, O. Tajima

Published Jul 15, 2019• 5 citations
arXivHTMLPDF

Abstract

Superconducting detectors are a modern technology applied in various fields. The microwave kinetic inductance detector (MKID) is one of the cutting-edge superconducting detectors. It is based on the principle of a superconducting resonator circuit. A radiation entering the MKID breaks the Cooper pairs in the superconducting resonator, and the intensity of the radiation is detected as a variation of the resonant condition. Therefore, calibration of the detector responsivity, i.e., the variation of the resonant phase with respect to the number of Cooper-pair breaks (quasiparticles), is important. We propose a method for responsivity calibration. Microwaves used for the detector readout locally raise the temperature in each resonator, which increases the number of quasiparticles. Since the magnitude of the temperature rise depends on the power of readout microwaves, the number of quasiparticles also depends on the power of microwaves. By changing the power of the readout microwaves, we simultaneously measure the phase difference and lifetime of quasiparticles. We calculate the number of quasiparticles from the measured lifetime and by using a theoretical formula. This measurement yields a relation between the phase responses as a function of the number of quasiparticles. We demonstrate this responsivity calibration using the MKID maintained at 285 mK. We also confirm the consistency between the results obtained using this method and conventional calibration methods in terms of the accuracy.

DOI
10.1063/1.5110692

Related papers

A method to measure superconducting transition temperature of microwave kinetic inductance detector by changing power of readout microwaves

Hiroki Kutsuma, Yoshinori Sueno, M. Hattori +5 more

A microwave kinetic inductance detector (MKID) is a cutting-edge superconducting detector, and its principle is based on a superconducting resonator circuit. The superconducting transition temperature (Tc) of the MKID is an important parameter because various MKID characterization parameters depend on it. In this paper, we propose a method to measure the Tc of the MKID by changing the applied power of the readout microwaves. A small fraction of the readout power is deposited on the MKID, and the number of quasiparticles in the MKID increases with this power. Furthermore, the quasiparticle lifetime decreases with the number of quasiparticles. Therefore, we can measure the relation between the quasiparticle lifetime and the detector response by rapidly varying the readout power. From this relation, we evaluate the intrinsic quasiparticle lifetime. This lifetime is theoretically modeled by Tc, the physical temperature of the MKID device, and other known parameters. We obtain Tc by comparing the measured lifetime with that acquired using the theoretical model. Using an MKID fabricated with aluminum, we demonstrate this method at a 0.3 K operation. The results are consistent with those obtained by Tc measured by monitoring the transmittance of the readout microwaves with the variation in the device temperature. The method proposed in this paper is applicable to other types, such as a hybrid-type MKID.

arXiv:2005.08765•AIP Advances•Sep 1, 2020•2 citations

Calibration Scheme for Large Kinetic Inductance Detector Arrays Based on Readout Frequency Response

Laura Bisigello, S. J. C. Yates, Vignesh Murugesan +2 more

Microwave kinetic inductance detector (MKID) provides a way to build large ground-based sub-mm instruments such as NIKA and A-MKID. For such instruments, therefore, it is important to understand and characterize the response to ensure good linearity and calibration over a wide dynamic range. We propose to use the MKID readout frequency response to determine the MKID responsivity to an input optical source power. A signal can be measured in a KID as a change in the phase of the readout signal with respect to the KID resonant circle. Fundamentally, this phase change is due to a shift in the KID resonance frequency, in turn due to a radiation induced change in the quasiparticle number in the superconducting resonator. We show that the shift in resonant frequency can be determined from the phase shift by using KID phase versus frequency dependence using a previously measured resonant frequency. Working in this calculated resonant frequency, we gain near linearity and constant calibration to a constant optical signal applied in a wide range of operating points on the resonance and readout powers. This calibration method has three particular advantages: first, it is fast enough to be used to calibrate large arrays, with pixel counts in the thousands of pixels; second, it is based on data that are already necessary to determine KID positions; third, it can be done without applying any optical source in front of the array.

arXiv:1602.00442•Journal of Low Temperature Physics•Feb 9, 2016•14 citations

Characterization of the quasiparticle lifetime of microwave kinetic inductance detector

Zhi Li, Wei-Tao Lv, Qing Shi +3 more

We design and fabricate a 350GHz 8×8 Al Microwave Kinetic Inductance Detector (MKIDs) array for the demonstration of its characteristics, mainly focus on the quasiparticle lifetime of the resonators. The quasiparticle lifetime data is collected by measuring a resonator’s phase response to a LED pulse at resonant frequency in a dilution refrigerator cooled to mK stage. We also measure and discuss the influence of various parameters on the change of quasiparticle lifetime, including different LED voltage supply, bath temperature of the MKIDs, and superconducting film thickness.

•Nov 24, 2021•0 citations

Validity of analytical Microwave Kinetic Inductance Detector responsivity expressions

Elmore Vaal

Astrophysical photons with a sub-millimetre wavelength are of great value to research star and galaxy formation. Superconductors have a small energy gap with which these sub-millimetre photons can be measured. Superconductors can absorb photons so that quasiparticles are created, which will in turn change the complex conductivity: incoming photons change the electrical response of the superconductor. This principle can be used to realise a detector. This is done with the Microwave Kinetic Inductance Detectors (MKIDs) that consist of a superconducting film embedded in a resonant circuit, so that photons falling in will change the resonant frequency. This can be measured with a readout signal. To enhance the sensitivity of the MKID, we want to increase the responsivity of an MKID as much as possible.A previous study has proposed a figure of merit regarding the material properties of a superconductor to achieve a high responsivity. However, the readout frequency used is not included in this figure. The goal of this work is to extend the figure of merit so that the readout frequency of the MKID can be optimised for a higher responsivity.

•Jan 1, 2016•0 citations

Design and measurement of superconducting microwave kinetic inductance detectors

Florian Blauth

Microwave kinetic inductance detectors (MKID) are superconducting broadband direct detectors. If a superconducting thin film patterned into a resonator is illuminated with photons, Cooper pairs break up and the kinetic inductance of the film changes. This shifts the resonance frequency downwards and the dissipation of the resonator increases, which can be detected in the phase and amplitude of a transmitted signal. \n \nIn this thesis, a waveguide coupled MKID is designed and simulated. 350 GHz radiation is coupled into a copper waveguide via a horn antenna. A waveguide probe leads the signal onto a 9 µm silicon membrane where it is absorbed in a 40 nm superconducting aluminum film which is the inductive part of a superconducting resonator. The capacitor of the resonator is made of niobium. Two versions of the device are developed, a transmission line resonator and a lumped element resonator design. The new design will enable an accurate understanding of the optical sensitivity of MKIDs and allow comparison of different resonator materials. Since two different resonator types are used, the direct comparison of transmission line and lumped resonators will also be possible. \n \nA measurement setup is developed including an ADR cryostat to cool MKID devices down to below 40 mK. An analog readout system using either a vector network analyzer or an IQ mixer is designed and built. The necessary software to control the ADR, the temperature monitoring, the readout signal generation and detection and the signal acquisition of the setup is also developed. \n \nThe waveguide coupled MKID design is fabricated in our in-house microfabrication laboratory and measured in the new setup. In my measurements, the devices do not show a resonance feature. The VNA and DC measurements are analyzed to determine a plausible hypothesis of the failure. The devices are diagnosed for critical issues during fabrication which will be addressed in a future iteration. \n \nTo verify the measurement setup, a test device of the IRAM NIKA 2 MKID camera is measured and the complete analysis pipeline is exercised. The quasi-particle lifetime in the aluminum thin films is measured and a dark and optical noise equivalent power are determined. The results are within the uncertainty comparable to measurements of IRAM themselves. \n \nA test chip designed in my master thesis that contains eight NbTiN resonators is measured in the new setup and non-linear effects under strong readout power are studied. Parametric amplification and oscillation of a resonator are shown qualitatively.

•Oct 1, 2019•0 citations

Ultra High Kinetic Inductance Detectors

Stefan Verheul

Microwave Kinetic Inductance Detectors (MKIDs) are extremely sensitive radiation detectors based on superconducting resonators that can be combined in large arrays on a single readout line within a limited frequency bandwidth. This makes MKIDs ideal detectors for the ultimate far-infrared observatory: a future space-based actively cooled telescope with its performance solely limited by the low universe background radiation. However, to reach these detector requirements, state-of-the-art MKIDs still need a order of magnitude improvement in device sensitivity. In this work, the MKID sensitivity is improved by reducing the aluminium volume that absorbs pair-breaking radiation into quasiparticle excitations, while making sure all radiation is still absorbed. Furthermore, a key requirement is sufficient reduction of excess noise as to keep the device intrinsically limited by thermally driven random fluctuations in the number of quasiparticles in absence of radiation, or Generation-Recombination (G-R) noise. To this end, a model is developed that describes the noise contributions as function of device geometry, readout power, material properties and radiation power. Subsequently, a realistic MKID design is presented and tested that reduces excess noise and maximises the sensitivity, expressed as Noise Equivalent Power (NEP). At high temperatures, good overall agreement is found between the measured noise spectra and the model. At low temperature T = 120 mK, the measurement results give an optical NEP similar to current state-of-the-art MKIDs. The NEP is not as low as expected due to short quasiparticle lifetimes, an unexpected decrease in the G-R noise level and a very high excess noise attributed to Two-Level Systems (TLS) noise that starts to dominate the already low G-R noise spectrum at low temperatures. Possibly, the quick quasiparticle lifetime saturation and noise level drop are caused by a strong readout power effect, as the readout power is known to create excess quasiparticles and to cause a strongly non-thermal electron energy distribution in the aluminium strip of the MKID. However, the exact microscopic details of these effects are unknown and not studied in this project. Based on the current chip design, a straightforward way to improve device performance and study the readout power effect in more detail is a reduction of the high TLS noise levels, which is possibly fabrication related. This would allow an unobstructed view of the G-R noise spectrum at low temperatures, thereby allowing both a study of the readout power effect on the quasiparticle system, and ultimately achieving the improvement in NEP needed reach the detector requirements for the ultimate space-based far-infrared observatory.

•Jan 1, 2019•0 citations
Superconducting detectors are a modern technology applied in various fields. The microwave kinetic inductance detector (MKID) is one of the cutting-edge superconducting detectors. It is based on the principle of a superconducting resonator circuit. A radiation entering the MKID breaks the Cooper pairs in the superconducting resonator, and the intensity of the radiation is detected as a variation of the resonant condition. Therefore, calibration of the detector responsivity, i.e., the variation of the resonant phase with respect to the number of Cooper-pair breaks (quasiparticles), is important. We propose a method for responsivity calibration. Microwaves used for the detector readout locally raise the temperature in each resonator, which increases the number of quasiparticles. Since the magnitude of the temperature rise depends on the power of readout microwaves, the number of quasiparticles also depends on the power of microwaves. By changing the power of the readout microwaves, we simultaneously measure the phase difference and lifetime of quasiparticles. We calculate the number of quasiparticles from the measured lifetime and by using a theoretical formula. This measurement yields a relation between the phase responses as a function of the number of quasiparticles. We demonstrate this responsivity calibration using the MKID maintained at 285 mK. We also confirm the consistency between the results obtained using this method and conventional calibration methods in terms of the accuracy.
A microwave kinetic inductance detector (MKID) is a cutting-edge superconducting detector, and its principle is based on a superconducting resonator circuit. The superconducting transition temperature (Tc) of the MKID is an important parameter because various MKID characterization parameters depend on it. In this paper, we propose a method to measure the Tc of the MKID by changing the applied power of the readout microwaves. A small fraction of the readout power is deposited on the MKID, and the number of quasiparticles in the MKID increases with this power. Furthermore, the quasiparticle lifetime decreases with the number of quasiparticles. Therefore, we can measure the relation between the quasiparticle lifetime and the detector response by rapidly varying the readout power. From this relation, we evaluate the intrinsic quasiparticle lifetime. This lifetime is theoretically modeled by Tc, the physical temperature of the MKID device, and other known parameters. We obtain Tc by comparing the measured lifetime with that acquired using the theoretical model. Using an MKID fabricated with aluminum, we demonstrate this method at a 0.3 K operation. The results are consistent with those obtained by Tc measured by monitoring the transmittance of the readout microwaves with the variation in the device temperature. The method proposed in this paper is applicable to other types, such as a hybrid-type MKID.
Microwave kinetic inductance detector (MKID) provides a way to build large ground-based sub-mm instruments such as NIKA and A-MKID. For such instruments, therefore, it is important to understand and characterize the response to ensure good linearity and calibration over a wide dynamic range. We propose to use the MKID readout frequency response to determine the MKID responsivity to an input optical source power. A signal can be measured in a KID as a change in the phase of the readout signal with respect to the KID resonant circle. Fundamentally, this phase change is due to a shift in the KID resonance frequency, in turn due to a radiation induced change in the quasiparticle number in the superconducting resonator. We show that the shift in resonant frequency can be determined from the phase shift by using KID phase versus frequency dependence using a previously measured resonant frequency. Working in this calculated resonant frequency, we gain near linearity and constant calibration to a constant optical signal applied in a wide range of operating points on the resonance and readout powers. This calibration method has three particular advantages: first, it is fast enough to be used to calibrate large arrays, with pixel counts in the thousands of pixels; second, it is based on data that are already necessary to determine KID positions; third, it can be done without applying any optical source in front of the array.
We design and fabricate a 350GHz 8×8 Al Microwave Kinetic Inductance Detector (MKIDs) array for the demonstration of its characteristics, mainly focus on the quasiparticle lifetime of the resonators. The quasiparticle lifetime data is collected by measuring a resonator’s phase response to a LED pulse at resonant frequency in a dilution refrigerator cooled to mK stage. We also measure and discuss the influence of various parameters on the change of quasiparticle lifetime, including different LED voltage supply, bath temperature of the MKIDs, and superconducting film thickness.
Astrophysical photons with a sub-millimetre wavelength are of great value to research star and galaxy formation. Superconductors have a small energy gap with which these sub-millimetre photons can be measured. Superconductors can absorb photons so that quasiparticles are created, which will in turn change the complex conductivity: incoming photons change the electrical response of the superconductor. This principle can be used to realise a detector. This is done with the Microwave Kinetic Inductance Detectors (MKIDs) that consist of a superconducting film embedded in a resonant circuit, so that photons falling in will change the resonant frequency. This can be measured with a readout signal. To enhance the sensitivity of the MKID, we want to increase the responsivity of an MKID as much as possible.A previous study has proposed a figure of merit regarding the material properties of a superconductor to achieve a high responsivity. However, the readout frequency used is not included in this figure. The goal of this work is to extend the figure of merit so that the readout frequency of the MKID can be optimised for a higher responsivity.
Microwave kinetic inductance detectors (MKID) are superconducting broadband direct detectors. If a superconducting thin film patterned into a resonator is illuminated with photons, Cooper pairs break up and the kinetic inductance of the film changes. This shifts the resonance frequency downwards and the dissipation of the resonator increases, which can be detected in the phase and amplitude of a transmitted signal. \n \nIn this thesis, a waveguide coupled MKID is designed and simulated. 350 GHz radiation is coupled into a copper waveguide via a horn antenna. A waveguide probe leads the signal onto a 9 µm silicon membrane where it is absorbed in a 40 nm superconducting aluminum film which is the inductive part of a superconducting resonator. The capacitor of the resonator is made of niobium. Two versions of the device are developed, a transmission line resonator and a lumped element resonator design. The new design will enable an accurate understanding of the optical sensitivity of MKIDs and allow comparison of different resonator materials. Since two different resonator types are used, the direct comparison of transmission line and lumped resonators will also be possible. \n \nA measurement setup is developed including an ADR cryostat to cool MKID devices down to below 40 mK. An analog readout system using either a vector network analyzer or an IQ mixer is designed and built. The necessary software to control the ADR, the temperature monitoring, the readout signal generation and detection and the signal acquisition of the setup is also developed. \n \nThe waveguide coupled MKID design is fabricated in our in-house microfabrication laboratory and measured in the new setup. In my measurements, the devices do not show a resonance feature. The VNA and DC measurements are analyzed to determine a plausible hypothesis of the failure. The devices are diagnosed for critical issues during fabrication which will be addressed in a future iteration. \n \nTo verify the measurement setup, a test device of the IRAM NIKA 2 MKID camera is measured and the complete analysis pipeline is exercised. The quasi-particle lifetime in the aluminum thin films is measured and a dark and optical noise equivalent power are determined. The results are within the uncertainty comparable to measurements of IRAM themselves. \n \nA test chip designed in my master thesis that contains eight NbTiN resonators is measured in the new setup and non-linear effects under strong readout power are studied. Parametric amplification and oscillation of a resonator are shown qualitatively.
Microwave Kinetic Inductance Detectors (MKIDs) are extremely sensitive radiation detectors based on superconducting resonators that can be combined in large arrays on a single readout line within a limited frequency bandwidth. This makes MKIDs ideal detectors for the ultimate far-infrared observatory: a future space-based actively cooled telescope with its performance solely limited by the low universe background radiation. However, to reach these detector requirements, state-of-the-art MKIDs still need a order of magnitude improvement in device sensitivity. In this work, the MKID sensitivity is improved by reducing the aluminium volume that absorbs pair-breaking radiation into quasiparticle excitations, while making sure all radiation is still absorbed. Furthermore, a key requirement is sufficient reduction of excess noise as to keep the device intrinsically limited by thermally driven random fluctuations in the number of quasiparticles in absence of radiation, or Generation-Recombination (G-R) noise. To this end, a model is developed that describes the noise contributions as function of device geometry, readout power, material properties and radiation power. Subsequently, a realistic MKID design is presented and tested that reduces excess noise and maximises the sensitivity, expressed as Noise Equivalent Power (NEP). At high temperatures, good overall agreement is found between the measured noise spectra and the model. At low temperature T = 120 mK, the measurement results give an optical NEP similar to current state-of-the-art MKIDs. The NEP is not as low as expected due to short quasiparticle lifetimes, an unexpected decrease in the G-R noise level and a very high excess noise attributed to Two-Level Systems (TLS) noise that starts to dominate the already low G-R noise spectrum at low temperatures. Possibly, the quick quasiparticle lifetime saturation and noise level drop are caused by a strong readout power effect, as the readout power is known to create excess quasiparticles and to cause a strongly non-thermal electron energy distribution in the aluminium strip of the MKID. However, the exact microscopic details of these effects are unknown and not studied in this project. Based on the current chip design, a straightforward way to improve device performance and study the readout power effect in more detail is a reduction of the high TLS noise levels, which is possibly fabrication related. This would allow an unobstructed view of the G-R noise spectrum at low temperatures, thereby allowing both a study of the readout power effect on the quasiparticle system, and ultimately achieving the improvement in NEP needed reach the detector requirements for the ultimate space-based far-infrared observatory.