Challenges to miniaturizing cold atom technology for deployable vacuum metrology

Stephen Eckel, Daniel S. Barker, James A. Fedchak, Nikolai N. Klimov, Eric B. Norrgard, Julia K. Scherschligt, Constantinos Makrides, Eite Tiesinga

Published Aug 21, 2018• 51 citations
arXivHTMLPDF

Abstract

Cold atoms are excellent metrological tools; they currently realize SI time and, soon, SI pressure in the ultra-high (UHV) and extreme high vacuum (XHV) regimes. The development of primary, vacuum metrology based on cold atoms currently falls under the purview of national metrology institutes. Under the emerging paradigm of the "quantum-SI", these technologies become deployable (relatively easy-to-use sensors that integrate with other vacuum chambers), providing a primary realization of the pascal in the UHV and XHV for the end-user. Here, we discuss the challenges that this goal presents. We investigate, for two different modes of operation, the expected corrections to the ideal cold-atom vacuum gauge and estimate the associated uncertainties. Finally, we discuss the appropriate choice of sensor atom, the light Li atom rather than the heavier Rb.

DOI
10.1088/1681-7575/aadbe4

Related papers

Precise quantum measurement of vacuum with cold atoms

Daniel S. Barker, B. P. Acharya, James A. Fedchak +5 more

We describe the cold-atom vacuum standards (CAVS) under development at the National Institute of Standards and Technology (NIST). The CAVS measures pressure in the ultra-high and extreme-high vacuum regimes by measuring the loss rate of sub-millikelvin sensor atoms from a magnetic trap. Ab initio quantum scattering calculations of cross sections and rate coefficients relate the density of background gas molecules or atoms to the loss rate of ultra-cold sensor atoms. The resulting measurement of pressure through the ideal gas law is traceable to the second and the kelvin, making it a primary realization of the pascal. At NIST, two versions of the CAVS have been constructed: a laboratory standard used to achieve the lowest possible uncertainties and pressures, and a portable version that is a potential replacement for the Bayard-Alpert ionization gauge. Both types of CAVSs are connected to a combined extreme-high vacuum flowmeter and dynamic expansion system to enable sensing of a known pressure of gas. In the near future, we anticipate being able to compare the laboratory scale CAVS, the portable CAVS, and the flowmeter/dynamic expansion system to validate the operation of the CAVS as both a standard and vacuum gauge.

Review of Scientific Instruments•Dec 1, 2022•15 citations

Roadmap on Quantum Sensors for Vacuum Metrology

D. S. Barker, J. L. Booth, S. P. Eckel +7 more

We describe how laser-cooled and trapped alkali-metal atoms have been used as pressure sensors in the high- and ultra-high-vacuum regimes operating with relative uncertainties of just over a percent at pressures from a few nPa and up to a few tens of $μ$Pa. These sensors can serve as primary standards of pressure as the determination of pressure requires only knowledge of collision cross sections between atoms and molecules, eliminating the need for calibration with other devices. The collision cross sections can be measured independently or computed with rigorous quantum methods at percent relative accuracies. We argue that this technology has advanced sufficiently to provide an alternative to the existing, expensive pressure standard in the ultra-high vacuum regime, the orifice flow standard. Experiments with trapped atoms can be used to calibrate cheaper pressure sensors (such as devices based on ionization gauges). However, the technology has not yet matured enough to be of use in other research fields because of the complexity and cost of the experiments. We propose a roadmap for this new pressure standard for vacuum metrology.

physics.atom-phCross-listedarXiv:2609.06626•Sep 8, 2026•0 citations

Vacuum pressure measurement of cold 7Li atoms in the magneto-optical and magnetic trap

Xuejiao Zhang, R. X. Li, Jingyi Zhang +2 more

Cold atoms play a very important role in metrology. The ultra-high or extreme-high vacuum pressure can be accurately measured by investigating the collision of cold atoms and residual atoms in the environment. We report a vacuum pressure metrometer using cold 7Li atoms as sensor atoms in a magneto-optical trap and a magnetic trap. The accuracy, uncertainty, and feasibility of the measurements in such two different traps are studied and compared in the pressure range from 10−5 to 10−8 Pa. It is found that the vacuum pressure measurement in a magneto-optical trap is more advantageous with a higher accuracy, higher signal-to-noise ratio, and shorter measurement time under current conditions. The results may be used to realize an applicable cold atom quantum vacuum standard.

AIP Advances•May 1, 2023•1 citation

Ultra-high vacuum measurement based on 7Li cold atoms manipulation

Yongjun Cheng, Meng Dong, Wen-Jun Sun +5 more

The redefinition of the International System of Units (SI) promotes the transformation of the vacuum measurement system toward quantization, and the quantization of vacuum parameters is one of the most leading, prospective and subversive research directions in the field of international vacuum metrology, and the quantum vacuum measurement is based on the quantum effect of the microscopic particle system, and the use of optical means and the theory of quantum mechanics to realize the precision measurement of the vacuum parameters. We develop a lithium-cooled atom vacuum measurement apparatus, which mainly consists of a 7Li atom trap system and a continuous expansion vacuum system. In this work, an experimental study of ultrahigh vacuum measurement is carried out by manipulating 7Li atoms and utilizing the loss characteristics of lithium cold atoms in magneto-optical and magnetic traps, and the results show that for the four commonly used gas molecules in vacuum, namely N2, Ar, He, and H2, in the vacuum range of (3×10–8–4×10–5) Pa, the maximum measurement uncertainty is 7.6%–6.0% (k = 2) based on 7Li cold atoms, and the cold atom vacuum measurement results are in good agreement with those of the traditional ionization vacuum gauges, and their relative sensitivities are in good agreement with those of the ionization vacuum gauges, and the maximal deviation of the relative sensitivity factor is less than 8%, which verifies the accuracy and reliability of the cold-atom quantum vacuum measurements. The research results are of great significance in promoting the development of new cross-generation vacuum measurement technology and meeting the needs of space science exploration, ultra-precision measurement and high-end equipment manufacturing.

Acta Physica Sinica•Jan 1, 2024•1 citation

Integrated atomic quantum technologies in demanding environments: Development and qualification of miniaturized optical setups and integration technologies for UHV and space operation

Marc Christ, Alexander Kassner, Robert Smol +6 more

Employing compact quantum sensors in field or in space (e.g., small satellites) implies demanding requirements on components and integration technologies. Within our work on integrated sensors, we develop miniaturized, ultra-stable optical setups for optical cooling and trapping of cold atomic gases. Besides challenging demands on alignment precision, and thermo-mechanical durability, we specifically address ultra-high vacuum (UHV) compatibility of our integration technologies and optical components. A prototype design of an UHV-compatible, crossed beam optical dipole trap setup and its application within a cold atomic quantum sensor is described. First qualification efforts on adhesive micro-integration technologies are presented. These tests are conducted in application-relevant geometries and material combinations common for micro-integrated optical setups. Adhesive aging will be investigated by thermal cycling or gamma radiation exposure. For vacuum compatibility testing, a versatile UHV testing system is currently being set up, enabling residual gas analysis and measurement of total gas rates down to 5•10-10mbar l/s at a base pressure of 10-11 mbar, exceeding the common ASTM E595 test.

•Jan 1, 2018•0 citations

Integrated atomic quantum technologies in demanding environments: development and qualification of miniaturized optical setups and integration technologies for UHV and space operation

Marc Christ, Alexander Kassner, Robert Smol +6 more

Employing compact quantum sensors in field or in space (e.g., small satellites) implies demanding requirements on components and integration technologies. Within our work on integrated sensors, we develop miniaturized, ultra-stable optical setups for optical cooling and trapping of cold atomic gases. Besides challenging demands on alignment precision, and thermo-mechanical durability, we specifically address ultra-high vacuum (UHV) compatibility of our integration technologies and optical components. A prototype design of an UHV-compatible, crossed beam optical dipole trap setup and its application within a cold atomic quantum sensor is described. First qualification efforts on adhesive micro-integration technologies are presented. These tests are conducted in application-relevant geometries and material combinations common for micro-integrated optical setups. Adhesive aging will be investigated by thermal cycling or gamma radiation exposure. For vacuum compatibility testing, a versatile UHV testing system is currently being set up, enabling residual gas analysis and measurement of total gas rates down to 5•10-10mbar l/s at a base pressure of 10-11 mbar, exceeding the common ASTM E595 test.

International Conference on Space Optics (ICSO)•Jul 12, 2019•1 citation
Cold atoms are excellent metrological tools; they currently realize SI time and, soon, SI pressure in the ultra-high (UHV) and extreme high vacuum (XHV) regimes. The development of primary, vacuum metrology based on cold atoms currently falls under the purview of national metrology institutes. Under the emerging paradigm of the "quantum-SI", these technologies become deployable (relatively easy-to-use sensors that integrate with other vacuum chambers), providing a primary realization of the pascal in the UHV and XHV for the end-user. Here, we discuss the challenges that this goal presents. We investigate, for two different modes of operation, the expected corrections to the ideal cold-atom vacuum gauge and estimate the associated uncertainties. Finally, we discuss the appropriate choice of sensor atom, the light Li atom rather than the heavier Rb.
We describe the cold-atom vacuum standards (CAVS) under development at the National Institute of Standards and Technology (NIST). The CAVS measures pressure in the ultra-high and extreme-high vacuum regimes by measuring the loss rate of sub-millikelvin sensor atoms from a magnetic trap. Ab initio quantum scattering calculations of cross sections and rate coefficients relate the density of background gas molecules or atoms to the loss rate of ultra-cold sensor atoms. The resulting measurement of pressure through the ideal gas law is traceable to the second and the kelvin, making it a primary realization of the pascal. At NIST, two versions of the CAVS have been constructed: a laboratory standard used to achieve the lowest possible uncertainties and pressures, and a portable version that is a potential replacement for the Bayard-Alpert ionization gauge. Both types of CAVSs are connected to a combined extreme-high vacuum flowmeter and dynamic expansion system to enable sensing of a known pressure of gas. In the near future, we anticipate being able to compare the laboratory scale CAVS, the portable CAVS, and the flowmeter/dynamic expansion system to validate the operation of the CAVS as both a standard and vacuum gauge.
We describe how laser-cooled and trapped alkali-metal atoms have been used as pressure sensors in the high- and ultra-high-vacuum regimes operating with relative uncertainties of just over a percent at pressures from a few nPa and up to a few tens of μμμPa. These sensors can serve as primary standards of pressure as the determination of pressure requires only knowledge of collision cross sections between atoms and molecules, eliminating the need for calibration with other devices. The collision cross sections can be measured independently or computed with rigorous quantum methods at percent relative accuracies. We argue that this technology has advanced sufficiently to provide an alternative to the existing, expensive pressure standard in the ultra-high vacuum regime, the orifice flow standard. Experiments with trapped atoms can be used to calibrate cheaper pressure sensors (such as devices based on ionization gauges). However, the technology has not yet matured enough to be of use in other research fields because of the complexity and cost of the experiments. We propose a roadmap for this new pressure standard for vacuum metrology.
Cold atoms play a very important role in metrology. The ultra-high or extreme-high vacuum pressure can be accurately measured by investigating the collision of cold atoms and residual atoms in the environment. We report a vacuum pressure metrometer using cold 7Li atoms as sensor atoms in a magneto-optical trap and a magnetic trap. The accuracy, uncertainty, and feasibility of the measurements in such two different traps are studied and compared in the pressure range from 10−5 to 10−8 Pa. It is found that the vacuum pressure measurement in a magneto-optical trap is more advantageous with a higher accuracy, higher signal-to-noise ratio, and shorter measurement time under current conditions. The results may be used to realize an applicable cold atom quantum vacuum standard.
The redefinition of the International System of Units (SI) promotes the transformation of the vacuum measurement system toward quantization, and the quantization of vacuum parameters is one of the most leading, prospective and subversive research directions in the field of international vacuum metrology, and the quantum vacuum measurement is based on the quantum effect of the microscopic particle system, and the use of optical means and the theory of quantum mechanics to realize the precision measurement of the vacuum parameters. We develop a lithium-cooled atom vacuum measurement apparatus, which mainly consists of a 7Li atom trap system and a continuous expansion vacuum system. In this work, an experimental study of ultrahigh vacuum measurement is carried out by manipulating 7Li atoms and utilizing the loss characteristics of lithium cold atoms in magneto-optical and magnetic traps, and the results show that for the four commonly used gas molecules in vacuum, namely N2, Ar, He, and H2, in the vacuum range of (3×10–8–4×10–5) Pa, the maximum measurement uncertainty is 7.6%–6.0% (k = 2) based on 7Li cold atoms, and the cold atom vacuum measurement results are in good agreement with those of the traditional ionization vacuum gauges, and their relative sensitivities are in good agreement with those of the ionization vacuum gauges, and the maximal deviation of the relative sensitivity factor is less than 8%, which verifies the accuracy and reliability of the cold-atom quantum vacuum measurements. The research results are of great significance in promoting the development of new cross-generation vacuum measurement technology and meeting the needs of space science exploration, ultra-precision measurement and high-end equipment manufacturing.
Employing compact quantum sensors in field or in space (e.g., small satellites) implies demanding requirements on components and integration technologies. Within our work on integrated sensors, we develop miniaturized, ultra-stable optical setups for optical cooling and trapping of cold atomic gases. Besides challenging demands on alignment precision, and thermo-mechanical durability, we specifically address ultra-high vacuum (UHV) compatibility of our integration technologies and optical components. A prototype design of an UHV-compatible, crossed beam optical dipole trap setup and its application within a cold atomic quantum sensor is described. First qualification efforts on adhesive micro-integration technologies are presented. These tests are conducted in application-relevant geometries and material combinations common for micro-integrated optical setups. Adhesive aging will be investigated by thermal cycling or gamma radiation exposure. For vacuum compatibility testing, a versatile UHV testing system is currently being set up, enabling residual gas analysis and measurement of total gas rates down to 5•10-10mbar l/s at a base pressure of 10-11 mbar, exceeding the common ASTM E595 test.
Employing compact quantum sensors in field or in space (e.g., small satellites) implies demanding requirements on components and integration technologies. Within our work on integrated sensors, we develop miniaturized, ultra-stable optical setups for optical cooling and trapping of cold atomic gases. Besides challenging demands on alignment precision, and thermo-mechanical durability, we specifically address ultra-high vacuum (UHV) compatibility of our integration technologies and optical components. A prototype design of an UHV-compatible, crossed beam optical dipole trap setup and its application within a cold atomic quantum sensor is described. First qualification efforts on adhesive micro-integration technologies are presented. These tests are conducted in application-relevant geometries and material combinations common for micro-integrated optical setups. Adhesive aging will be investigated by thermal cycling or gamma radiation exposure. For vacuum compatibility testing, a versatile UHV testing system is currently being set up, enabling residual gas analysis and measurement of total gas rates down to 5•10-10mbar l/s at a base pressure of 10-11 mbar, exceeding the common ASTM E595 test.