arXiv:1403.7796 · Review of Scientific Instruments
Shot-noise-limited magnetometer with sub-picotesla sensitivity at room temperature
Vito Giovanni Lucivero, Pawel Anielski, Wojciech Gawlik, Morgan W. Mitchell
Abstract
We report a photon shot-noise-limited (SNL) optical magnetometer based on amplitude modulated optical rotation using a room-temperature (85)Rb vapor in a cell with anti-relaxation coating. The instrument achieves a room-temperature sensitivity of 70 fT / √Hz at 7.6 μT. Experimental scaling of noise with optical power, in agreement with theoretical predictions, confirms the SNL behaviour from 5 μT to 75 μT. The combination of best-in-class sensitivity and SNL operation makes the system a promising candidate for application of squeezed light to a state-of-the-art atomic sensor.