Evaluation of an In+ Coulomb crystal clock with relative uncertainties in the low 10−18 range

H. N. Hausser

Published Jan 1, 2025• 0 citations
Publisher

Abstract

Clocks have been among the most accurate man-made devices for centuries. Atomic clocks, which have been defining time by means of a microwave transition in the 133Cs atom since 1967, represent a major step in this development. A redefinition of the SI unit of time based on optical atomic clocks is currently under discussion. This forthcoming redefinition requires, among other criteria, the repeated demonstration of the consistency of different optical frequency standards in the low 10−18 range. In this work, the first evaluation of the systematic shifts and their uncertainties in a 115In+–172Yb+ Coulomb crystal clock is presented. This clock utilizes the 1S0 ↔ 3P0 transition in 115In+. 172Yb+ ions are co-trapped with the 115In+ ions in a segmented Paul trap and used for sympathetic cooling. With a systematic uncertainty of 2.5×10−18 when operating with a 1In+–3Yb+ crystal and 2.6×10−18 when operating with a 2In+–4Yb+ crystal, this clock is one of the most accurate clocks in the world. In addition, the operation of the clock with multiple clock ions is the first of its kind at this level of uncertainty. Measurements with up to four clock ions show a reduction in instability from 1.6×10−15/√ τ for a single indium ion to 9.2(4)×10−16/√τ when operating with four indium ions. By using an additional quench laser, the instability in another series of measurements was reduced from a single-ion instability of 1.5×10−15/√ τ to a four-ion instability of 7.1(15)×10−16/√τ , which is consistent with the expected scaling of 1/√N with the number of clock ions N. In this work, the hyperfine structure of the 3P0 ↔ 1P1 transition used for quenching was also resolved for the first time and the hyperfine constants were determined to be AHFS = 265.42(75)MHz and BHFS = 291.4(135) MHz. In frequency comparisons with other clocks, the optical frequency ratios 115In+/171Yb+(E3) and 115In+/87Sr were determined. With a relative total uncertainty of 4.2 × 10−18, the former ratio is the most accurate frequency ratio of different species published to date. It is also the first to meet one of the requirements defined on the roadmap for redefining the SI unit of the second: Measurements of frequency ratios with relative total measurement uncertainties below 5 × 10−18. The determination of the absolute frequency results in an uncertainty of 1.3 × 10−16 for the 1S0 ↔ 3P0 transition in 115In+, which is limited by that of the cesium fountains. The results obtained summarized: ν115In+ ν171Yb+(E3) =1.973 773 591 557 215 788(8) ν115In+ ν87Sr =2.952 748 749 874 860 781(27) ν115In+ =1 267 402 452 901 038.87(16) Hz

DOI
10.15488/19232

Related papers

In + 115 − Yb + 172 Coulomb Crystal Clock with 2.5 × 10 − 18 Systematic Uncertainty

H. N. Hausser, J. Keller, T. Nordmann +17 more

We present a scalable mixed-species Coulomb crystal clock based on the ^{1}S_{0}↔^{3}P_{0} transition in ^{115}In^{+}. ^{172}Yb^{+} ions are cotrapped and used for sympathetic cooling. Reproducible interrogation conditions for mixed-species Coulomb crystals are ensured by a conditional preparation sequence with permutation control. We demonstrate clock operation with a 1In^{+}-3Yb^{+} crystal, achieving a relative systematic uncertainty of 2.5×10^{-18} and a relative frequency instability of 1.6×10^{-15}/sqrt[τ/1 s]. We report on absolute frequency measurements with an uncertainty of 1.3×10^{-16} and optical frequency comparisons with clocks based on ^{171}Yb^{+} (E3) and ^{87}Sr. With a fractional uncertainty of 4.4×10^{-18}, the former is-to our knowledge-the most accurate frequency ratio value reported to date. For the ^{115}In^{+}/^{87}Sr ratio, we improve upon the best previous measurement by more than an order of magnitude. We also demonstrate operation with four ^{115}In^{+} clock ions, which reduces the instability to 9.2×10^{-16}/sqrt[τ/1 s].

arXiv:2402.16807•Physical Review Letters•Jan 16, 2025•42 citations

In+ 115 - Yb+ 172 Coulomb Crystal Clock with 2.5×10-18 Systematic Uncertainty

H. N. Hausser, Jonas Keller, T. Nordmann +12 more

We present a scalable mixed-species Coulomb crystal clock based on the S01↔P30 transition in In+115. Yb+172 ions are cotrapped and used for sympathetic cooling. Reproducible interrogation conditions for mixed-species Coulomb crystals are ensured by a conditional preparation sequence with permutation control. We demonstrate clock operation with a 1In+-3Yb+ crystal, achieving a relative systematic uncertainty of 2.5×10-18 and a relative frequency instability of 1.6×10-15/τ/1 s. We report on absolute frequency measurements with an uncertainty of 1.3×10-16 and optical frequency comparisons with clocks based on Yb+171 (E3) and Sr87. With a fractional uncertainty of 4.4×10-18, the former is - to our knowledge - the most accurate frequency ratio value reported to date. For the In+115/Sr87 ratio, we improve upon the best previous measurement by more than an order of magnitude. We also demonstrate operation with four In+115 clock ions, which reduces the instability to 9.2×10-16/τ/1 s.

•Jan 1, 2025•0 citations

High-accuracy multi-ion spectroscopy with mixed-species Coulomb crystals

Jonas Keller, H. N. Hausser, Ingo Richter +14 more

Abstract Multi-ion optical clocks offer the possibility of overcoming the low signal-to-noise ratio of single-ion clocks, while still providing low systematic uncertainties. We present simultaneous spectroscopy of up to four 115 In + clock ions in a linear Coulomb crystal, sympathetically cooled with 172 Yb + ions. In first clock comparisons, we see agreement below 1 × 10 −17 with results obtained using a single In + ion, for which we have evaluated the systematic uncertainty to be 2.5 × 10 −18 . Operation with four clock ions reduces the instability from 1.6 × 10 − 15 / t / ( 1 s ) to 9.2 × 10 − 16 / t / ( 1 s ) . We derive a model for decay-related dead time during state preparation, which matches the observed scaling of instability with clock ion number N , and indicates that 1 / N scaling can be achieved with the addition of a repump laser.

arXiv:2501.09615•Journal of Physics Conference Series•Nov 1, 2024•1 citation

Linear Paul trap design for an optical clock with Coulomb crystals

Norbert Herschbach, Karsten Pyka, J. Keller +1 more

We report on the design of a segmented linear Paul trap for optical clock applications using trapped ion Coulomb crystals. For an optical clock with an improved short-term stability and a fractional frequency uncertainty of 10−18, we propose 115In+ ions sympathetically cooled by 172Yb+. We discuss the systematic frequency shifts of such a frequency standard. In particular, we elaborate on high-precision calculations of the electric radiofrequency field of the ion trap using the finite element method. These calculations are used to find a scalable design with minimized excess micromotion of the ions at a level at which the corresponding second-order Doppler shift contributes less than 10−18 to the relative uncertainty of the frequency standard.

•Aug 17, 2016•34 citations

Indium single-ion optical frequency standard

Thomas P. Becker, M. Eichenseer, Alexander Yu. Nevsky +5 more

Single laser-cooled ions in radiofrequency traps can serve as the basis for highly stable optical frequency standards. Here we present recent results on In+, using the 1S0 yields 3P0 line at 236.5 nm wavelength as the clock transition. This resonance has a natural linewidth of only 0.82 Hz and systematic shifts should be controllable at the mHz level. A single indium ion is stored in a miniature Paul- Straubel trap and laser cooled to a temperature of about 100 (mu) K. The clock transition is excited using a frequency quadrupled 946 nm Nd:YAG laser. A fractional resolution of 1.3 (DOT) 10-13 has been achieved so far (linewidth of 170 Hz at 1267 THz). The absolute optical frequency of the clock transition has been measured with an uncertainty being less than 3 (DOT) 10-13 using a frequency chain and a methane-stabilized HeNe laser that was calibrated with a cesium clock.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•May 3, 2001•0 citations

A dual species optical atomic clock

Martin Steinel

Frequency standards based on optical transitions between electronic states of trapped ions have reached uncertainties below 10−18 [1]. The Stark shift caused by room-temperature blackbody radiation (BBR) causes the largest correction for most high-performance optical clock. The systematic uncertainty of 171Yb+ clocks based on the 2S1/2 (F = 0) → 2D3/2 (F = 2) electric quadrupole (E2) and 2S1/2 (F = 0) → 2F7/2 (F = 3) electric octupole (E3) transitions is currently limited to 28×10−18 and 1.5×10−18, respectively, by the fractional accuracy of the differential polarizability Δαdc of approximately 2% [2]. In the scope of this thesis, a dual species clock based on 171Yb+ and 88Sr+ ions was set up, operated and evaluated, to overcome this barrier. Since Δαdc is known with a smaller fractional uncertainty of 0.04% for the 2S1/2 → 2D5/2 E2 transition of 88Sr+ [3], this allows for a transfer of Δαdc from 88Sr+ to 171Yb+ following a method proposed in [4]. The optical intensity of a laser is calibrated with the Stark shift it causes on the 88Sr+ clock transition, allowing for the determination of Δαdc of the 171Yb+ clock transitions. This allows for a reduction of the fractional uncertainty of Δαdc for both 171Yb+ clock transitions below 0.4% and clock operation on the E2 and E3 transition with a fractional BBR shift uncertainty of 0.84×10−18 and 0.24×10−18 is enabled, respectively. However the value of Δαdc determined in this thesis does not agree with [2], a fractional offset of approximately 14% is observed. To investigate this discrepancy further, Δαdc is measured in the same manner as in [2] for the 88Sr+ clock transition. The same offset of approximately 14% is found, calling the accuracy of at least one of these methods into question. Finally, the first optical frequency ratio ℛ between the 88Sr+ clock transition and the 171Yb+ E3 transition is measured. The fractional uncertainty of ℛ is 23×10−18. ℛ in combination with the absolute frequency of the E3 transition allows for the determination of the 88Sr+ absolute frequency νSr+ with an uncertainty limited by the caesium references. The value of νSr+ falls outside of the recommended range, but is corroborated by measurements of the same quantity by other institutes [5, 6]. [1] M. Marshall et al., Phys. Rev. Lett. 135, 033201 (2025) [2] N. Huntemann et al., Phys. Rev. Lett. 116, 063001 (2016) [3] T. Lindvall et al., Phys. Rev. Lett. 135, 043402 (2025) [4] M. D. Barrett et al., Phys. Rev. A 100, 043418 (2019) [5] C. Marceau et al., Metrologia 62.4, 045001 (2025) [6] T. Lindvall et al., Phys. Rev. Appl. 24, 044082 (2025)

•Dec 16, 2025•0 citations
Clocks have been among the most accurate man-made devices for centuries. Atomic clocks, which have been defining time by means of a microwave transition in the 133Cs atom since 1967, represent a major step in this development. A redefinition of the SI unit of time based on optical atomic clocks is currently under discussion. This forthcoming redefinition requires, among other criteria, the repeated demonstration of the consistency of different optical frequency standards in the low 10−18 range. In this work, the first evaluation of the systematic shifts and their uncertainties in a 115In+–172Yb+ Coulomb crystal clock is presented. This clock utilizes the 1S0 ↔ 3P0 transition in 115In+. 172Yb+ ions are co-trapped with the 115In+ ions in a segmented Paul trap and used for sympathetic cooling. With a systematic uncertainty of 2.5×10−18 when operating with a 1In+–3Yb+ crystal and 2.6×10−18 when operating with a 2In+–4Yb+ crystal, this clock is one of the most accurate clocks in the world. In addition, the operation of the clock with multiple clock ions is the first of its kind at this level of uncertainty. Measurements with up to four clock ions show a reduction in instability from 1.6×10−15/√ τ for a single indium ion to 9.2(4)×10−16/√τ when operating with four indium ions. By using an additional quench laser, the instability in another series of measurements was reduced from a single-ion instability of 1.5×10−15/√ τ to a four-ion instability of 7.1(15)×10−16/√τ , which is consistent with the expected scaling of 1/√N with the number of clock ions N. In this work, the hyperfine structure of the 3P0 ↔ 1P1 transition used for quenching was also resolved for the first time and the hyperfine constants were determined to be AHFS = 265.42(75)MHz and BHFS = 291.4(135) MHz. In frequency comparisons with other clocks, the optical frequency ratios 115In+/171Yb+(E3) and 115In+/87Sr were determined. With a relative total uncertainty of 4.2 × 10−18, the former ratio is the most accurate frequency ratio of different species published to date. It is also the first to meet one of the requirements defined on the roadmap for redefining the SI unit of the second: Measurements of frequency ratios with relative total measurement uncertainties below 5 × 10−18. The determination of the absolute frequency results in an uncertainty of 1.3 × 10−16 for the 1S0 ↔ 3P0 transition in 115In+, which is limited by that of the cesium fountains. The results obtained summarized: ν115In+ ν171Yb+(E3) =1.973 773 591 557 215 788(8) ν115In+ ν87Sr =2.952 748 749 874 860 781(27) ν115In+ =1 267 402 452 901 038.87(16) Hz
We present a scalable mixed-species Coulomb crystal clock based on the ^{1}S_{0}↔^{3}P_{0} transition in ^{115}In^{+}. ^{172}Yb^{+} ions are cotrapped and used for sympathetic cooling. Reproducible interrogation conditions for mixed-species Coulomb crystals are ensured by a conditional preparation sequence with permutation control. We demonstrate clock operation with a 1In^{+}-3Yb^{+} crystal, achieving a relative systematic uncertainty of 2.5×10^{-18} and a relative frequency instability of 1.6×10^{-15}/sqrt[τ/1 s]. We report on absolute frequency measurements with an uncertainty of 1.3×10^{-16} and optical frequency comparisons with clocks based on ^{171}Yb^{+} (E3) and ^{87}Sr. With a fractional uncertainty of 4.4×10^{-18}, the former is-to our knowledge-the most accurate frequency ratio value reported to date. For the ^{115}In^{+}/^{87}Sr ratio, we improve upon the best previous measurement by more than an order of magnitude. We also demonstrate operation with four ^{115}In^{+} clock ions, which reduces the instability to 9.2×10^{-16}/sqrt[τ/1 s].
We present a scalable mixed-species Coulomb crystal clock based on the S01↔P30 transition in In+115. Yb+172 ions are cotrapped and used for sympathetic cooling. Reproducible interrogation conditions for mixed-species Coulomb crystals are ensured by a conditional preparation sequence with permutation control. We demonstrate clock operation with a 1In+-3Yb+ crystal, achieving a relative systematic uncertainty of 2.5×10-18 and a relative frequency instability of 1.6×10-15/τ/1 s. We report on absolute frequency measurements with an uncertainty of 1.3×10-16 and optical frequency comparisons with clocks based on Yb+171 (E3) and Sr87. With a fractional uncertainty of 4.4×10-18, the former is - to our knowledge - the most accurate frequency ratio value reported to date. For the In+115/Sr87 ratio, we improve upon the best previous measurement by more than an order of magnitude. We also demonstrate operation with four In+115 clock ions, which reduces the instability to 9.2×10-16/τ/1 s.
Abstract Multi-ion optical clocks offer the possibility of overcoming the low signal-to-noise ratio of single-ion clocks, while still providing low systematic uncertainties. We present simultaneous spectroscopy of up to four 115 In + clock ions in a linear Coulomb crystal, sympathetically cooled with 172 Yb + ions. In first clock comparisons, we see agreement below 1 × 10 −17 with results obtained using a single In + ion, for which we have evaluated the systematic uncertainty to be 2.5 × 10 −18 . Operation with four clock ions reduces the instability from 1.6 × 10 − 15 / t / ( 1 s ) to 9.2 × 10 − 16 / t / ( 1 s ) . We derive a model for decay-related dead time during state preparation, which matches the observed scaling of instability with clock ion number N , and indicates that 1 / N scaling can be achieved with the addition of a repump laser.
We report on the design of a segmented linear Paul trap for optical clock applications using trapped ion Coulomb crystals. For an optical clock with an improved short-term stability and a fractional frequency uncertainty of 10−18, we propose 115In+ ions sympathetically cooled by 172Yb+. We discuss the systematic frequency shifts of such a frequency standard. In particular, we elaborate on high-precision calculations of the electric radiofrequency field of the ion trap using the finite element method. These calculations are used to find a scalable design with minimized excess micromotion of the ions at a level at which the corresponding second-order Doppler shift contributes less than 10−18 to the relative uncertainty of the frequency standard.
Single laser-cooled ions in radiofrequency traps can serve as the basis for highly stable optical frequency standards. Here we present recent results on In+, using the 1S0 yields 3P0 line at 236.5 nm wavelength as the clock transition. This resonance has a natural linewidth of only 0.82 Hz and systematic shifts should be controllable at the mHz level. A single indium ion is stored in a miniature Paul- Straubel trap and laser cooled to a temperature of about 100 (mu) K. The clock transition is excited using a frequency quadrupled 946 nm Nd:YAG laser. A fractional resolution of 1.3 (DOT) 10-13 has been achieved so far (linewidth of 170 Hz at 1267 THz). The absolute optical frequency of the clock transition has been measured with an uncertainty being less than 3 (DOT) 10-13 using a frequency chain and a methane-stabilized HeNe laser that was calibrated with a cesium clock.
Frequency standards based on optical transitions between electronic states of trapped ions have reached uncertainties below 10−18 [1]. The Stark shift caused by room-temperature blackbody radiation (BBR) causes the largest correction for most high-performance optical clock. The systematic uncertainty of 171Yb+ clocks based on the 2S1/2 (F = 0) → 2D3/2 (F = 2) electric quadrupole (E2) and 2S1/2 (F = 0) → 2F7/2 (F = 3) electric octupole (E3) transitions is currently limited to 28×10−18 and 1.5×10−18, respectively, by the fractional accuracy of the differential polarizability Δαdc of approximately 2% [2]. In the scope of this thesis, a dual species clock based on 171Yb+ and 88Sr+ ions was set up, operated and evaluated, to overcome this barrier. Since Δαdc is known with a smaller fractional uncertainty of 0.04% for the 2S1/2 → 2D5/2 E2 transition of 88Sr+ [3], this allows for a transfer of Δαdc from 88Sr+ to 171Yb+ following a method proposed in [4]. The optical intensity of a laser is calibrated with the Stark shift it causes on the 88Sr+ clock transition, allowing for the determination of Δαdc of the 171Yb+ clock transitions. This allows for a reduction of the fractional uncertainty of Δαdc for both 171Yb+ clock transitions below 0.4% and clock operation on the E2 and E3 transition with a fractional BBR shift uncertainty of 0.84×10−18 and 0.24×10−18 is enabled, respectively. However the value of Δαdc determined in this thesis does not agree with [2], a fractional offset of approximately 14% is observed. To investigate this discrepancy further, Δαdc is measured in the same manner as in [2] for the 88Sr+ clock transition. The same offset of approximately 14% is found, calling the accuracy of at least one of these methods into question. Finally, the first optical frequency ratio ℛ between the 88Sr+ clock transition and the 171Yb+ E3 transition is measured. The fractional uncertainty of ℛ is 23×10−18. ℛ in combination with the absolute frequency of the E3 transition allows for the determination of the 88Sr+ absolute frequency νSr+ with an uncertainty limited by the caesium references. The value of νSr+ falls outside of the recommended range, but is corroborated by measurements of the same quantity by other institutes [5, 6]. [1] M. Marshall et al., Phys. Rev. Lett. 135, 033201 (2025) [2] N. Huntemann et al., Phys. Rev. Lett. 116, 063001 (2016) [3] T. Lindvall et al., Phys. Rev. Lett. 135, 043402 (2025) [4] M. D. Barrett et al., Phys. Rev. A 100, 043418 (2019) [5] C. Marceau et al., Metrologia 62.4, 045001 (2025) [6] T. Lindvall et al., Phys. Rev. Appl. 24, 044082 (2025)