Journal of Physics Conference Series
Influence of Cell Temperature on the Signal-To-Noise Ratio of a Single-Beam Optically Pumped Atomic Magnetometer
Yuntian Zou, Xiangyang Zhou, Jiali Liu, Wei Quan, Liwei Jiang
Abstract
Abstract The single-beam optically pumped atomic magnetometer (OPAM) provides an effective solution for high-precision magnetic field measurements under nonzero-field conditions, including the geomagnetic field. However, the magnetometer performance is strongly influenced by the cell temperature, and the underlying mechanism by which the cell temperature affects the performance of single-beam OPAMs needs to be clarified. Based on the principle of the single-beam OPAM, a theoretical model for the scale factor is proposed, and the temperature dependence of the magnetometer signal-to-noise ratio (SNR) is investigated theoretically and experimentally. Both the theoretical analysis and experimental investigation indicate that an optimal temperature exists at which the SNR reaches its maximum, and this optimal temperature is determined for the constructed single-beam OPAM system. This work facilitates the optimization of magnetometer sensitivity and supports wearable magnetocardiography and magnetoencephalography measurements in geomagnetic field environments.