A primary radiation standard based on quantum nonlinear optics

Lemieux, Samuel, Giese, Enno, Fickler, Robert, Chekhova, Maria V., Boyd, Robert W.

Published Jan 1, 2019• 0 citations
Publisher

Abstract

The black body remains the most prominent source of light for absolute radiometry. Its main alternative, synchrotron radiation, requires costly and large facilities. Quantum optics offers a new radiometric source: parametric down-conversion (PDC), a nonlinear optical process, in which pairwise photon correlations enable absolute calibration of photodetectors. Since the emission rate crucially depends on the brightness of the electromagnetic field, quantum-mechanical fluctuations of the vacuum7 can be seen as a seed of spontaneous PDC, and their amplitude is a natural radiometric standard. Thus, they allow for the calibration of the spectral radiance of light sources by measuring the ratio between seeded and unseeded PDC. Here, we directly use the frequency spectrum of the electromagnetic vacuum to trigger spontaneous PDC and employ the generated light to infer the spectral response of a spectrometer over a broad spectral range. Then, we deduce the absolute quantum efficiency from the spectral shape of PDC in the high-gain regime, without relying on a seed or reference detector. Our results compare well with the ones obtained with a reference lamp, demonstrating a promising primary radiation standard.

DOI
10.26083/tuprints-00027264

Related papers

Absolute self-calibration of single-photon detectors with cw and pulsed spontaneous parametric down-conversion

Ling-An Wu, Xi-Hao Chen, Yanhua Zhai +2 more

Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Aug 31, 2006•0 citations

Absolute measurements of radiation sources spectral brightness and detectors quantum efficiency

Alexander N. Penin, Давид Николаевич Клышко

In this paper we present schemes of experimental setups for the radiation spectral brightness measurements in the range of 0.6 - 5 (mu) for N varying from 10-1 to 102 (if, for example, (lambda) equals 1 (mu) , this range of N corresponds to the brightness temperature range from 6 (DOT) 103 to 106 K), and for photomultipliers quantum efficiency measurements in the range of 0.4 - 5 (mu) with a dynamical range 10 - 1012 photon/sec and accuracy not worse than 1%. The new measurement methods are based on the utilization of the parametric light scattering phenomenon which is a spontaneous decay of laser pump photons in correlated photon pairs in crystals with quadratic nonlinear susceptibility. The first of two methods allows measurement of the radiation spectral brightness N in absolute units ('photons per mode') in visible and infrared range. The quantity N is related to the energetic brightness spectral density B through the equation B equals (hc2/(lambda) 5)N, where h is the Plank constant, c - the light velocity, (lambda) - the wavelength. The method is absolute and does not require any reference source or detector of radiation. Quantum noise of a parametric down-convertor, caused by the zero vacuum fluctuations with an effective brightness Nvac equals 1 photon per mode, is the reference in this case. The second method concerns the quantum efficiency of photodetectors determination, and it is based on the connection between the statistics of photocurrent and the radiation which causes it. The parametric scattering is a unique source of rather intensive and directed radiation flow consisting of photon pairs. Such a flow can be used to determine the absolute quantum efficiency of photodetectors.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Dec 1, 1991•1 citation

Absolute standardless calibration of photodetectors based on quantum two-photon fields

Alexander N. Penin, Alexander V. Sergienko

A new method of absolute calibration of photodetectors based on a nonclassical effect in nonlinear optics is discussed. The combined influence of a monochromatic flux of pump photons and quantum vacuum noise on a nonlinear crystal with no center of symmetry results in the appearance of rigorously correlated pairs of photons with continuous spectral distribution in the spontaneous parametric scattering of light (spontaneous parametric downconversion) process. The presence of an optical field in a broad spectral range created by the two-photon states (biphotons) makes it possible to develop a new method of measuring the spectral distribution of the absolute value of the quantum efficiency of photodetectors, both in photon counting and in analog regimes, that does not use any calibrated standard light sources (étalons). The feasibility of the method is demonstrated on different types of photomultiplier tube.

Applied Optics•Sep 1, 1991•65 citations

Absolute self-calib ration of single-photon detectors with cw and pulsed spontaneous parametric down-conversion - art. no. 630505

La Wu, Xiangdong Chen, YH Zhai +2 more

Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.

•Jan 1, 2006•0 citations

A Primary Radiation Standard Based on Quantum Nonlinear Optics

Samuel Lemieux, Enno Giese, Robert Fickler +2 more

We introduce spontaneous parametric down-conversion as a light source for intensity calibration of spectrometers solely based on physical principles. The response function of the spectrometer obtained with our technique agrees with the one measured with a NIST-traceable calibration lamp.

Conference on Lasers and Electro-Optics•Jan 1, 2018•0 citations

Investigation of absolute measurement in quantum efficiency of photoelectric detectors by means of correlated photons

Yu Feng, Xiaobing Zheng, Yanli Qiao +1 more

There are two general methods for radiometric calibration of detectors. One is based on radiometric sources and the other is based on detectors. Because two methods need to establish primary standard of high precision and transfer chain of standard, the precision of standard will be reduced by extension of the chain. An absolute calibration of quantum efficiency of detectors can be realized by using correlated photons generated in spontaneous parametric down-conversion (SPDC) effect of nonlinear crystal. Generation mechanism of SPDC is interpreted with theories of quantum mechanics and classical theories. The theory of calibration based correlated photons are demonstrated. The project for proof of principle of absolute calibration of quantum efficiency of radiometric detectors is given in this paper. Comparing with traditional methods, the advantages of it are pointed out. A scheme using to calibrate quantum efficiency of infrared detectors is described here. The measurement of spatial and spectral response of detectors is mentioned too.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Feb 23, 2006•0 citations
The black body remains the most prominent source of light for absolute radiometry. Its main alternative, synchrotron radiation, requires costly and large facilities. Quantum optics offers a new radiometric source: parametric down-conversion (PDC), a nonlinear optical process, in which pairwise photon correlations enable absolute calibration of photodetectors. Since the emission rate crucially depends on the brightness of the electromagnetic field, quantum-mechanical fluctuations of the vacuum7 can be seen as a seed of spontaneous PDC, and their amplitude is a natural radiometric standard. Thus, they allow for the calibration of the spectral radiance of light sources by measuring the ratio between seeded and unseeded PDC. Here, we directly use the frequency spectrum of the electromagnetic vacuum to trigger spontaneous PDC and employ the generated light to infer the spectral response of a spectrometer over a broad spectral range. Then, we deduce the absolute quantum efficiency from the spectral shape of PDC in the high-gain regime, without relying on a seed or reference detector. Our results compare well with the ones obtained with a reference lamp, demonstrating a promising primary radiation standard.
Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.
In this paper we present schemes of experimental setups for the radiation spectral brightness measurements in the range of 0.6 - 5 (mu) for N varying from 10-1 to 102 (if, for example, (lambda) equals 1 (mu) , this range of N corresponds to the brightness temperature range from 6 (DOT) 103 to 106 K), and for photomultipliers quantum efficiency measurements in the range of 0.4 - 5 (mu) with a dynamical range 10 - 1012 photon/sec and accuracy not worse than 1%. The new measurement methods are based on the utilization of the parametric light scattering phenomenon which is a spontaneous decay of laser pump photons in correlated photon pairs in crystals with quadratic nonlinear susceptibility. The first of two methods allows measurement of the radiation spectral brightness N in absolute units ('photons per mode') in visible and infrared range. The quantity N is related to the energetic brightness spectral density B through the equation B equals (hc2/(lambda) 5)N, where h is the Plank constant, c - the light velocity, (lambda) - the wavelength. The method is absolute and does not require any reference source or detector of radiation. Quantum noise of a parametric down-convertor, caused by the zero vacuum fluctuations with an effective brightness Nvac equals 1 photon per mode, is the reference in this case. The second method concerns the quantum efficiency of photodetectors determination, and it is based on the connection between the statistics of photocurrent and the radiation which causes it. The parametric scattering is a unique source of rather intensive and directed radiation flow consisting of photon pairs. Such a flow can be used to determine the absolute quantum efficiency of photodetectors.
A new method of absolute calibration of photodetectors based on a nonclassical effect in nonlinear optics is discussed. The combined influence of a monochromatic flux of pump photons and quantum vacuum noise on a nonlinear crystal with no center of symmetry results in the appearance of rigorously correlated pairs of photons with continuous spectral distribution in the spontaneous parametric scattering of light (spontaneous parametric downconversion) process. The presence of an optical field in a broad spectral range created by the two-photon states (biphotons) makes it possible to develop a new method of measuring the spectral distribution of the absolute value of the quantum efficiency of photodetectors, both in photon counting and in analog regimes, that does not use any calibrated standard light sources (étalons). The feasibility of the method is demonstrated on different types of photomultiplier tube.
Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.
There are two general methods for radiometric calibration of detectors. One is based on radiometric sources and the other is based on detectors. Because two methods need to establish primary standard of high precision and transfer chain of standard, the precision of standard will be reduced by extension of the chain. An absolute calibration of quantum efficiency of detectors can be realized by using correlated photons generated in spontaneous parametric down-conversion (SPDC) effect of nonlinear crystal. Generation mechanism of SPDC is interpreted with theories of quantum mechanics and classical theories. The theory of calibration based correlated photons are demonstrated. The project for proof of principle of absolute calibration of quantum efficiency of radiometric detectors is given in this paper. Comparing with traditional methods, the advantages of it are pointed out. A scheme using to calibrate quantum efficiency of infrared detectors is described here. The measurement of spatial and spectral response of detectors is mentioned too.