Measurement Sensors

Frequency shift compensation for single and dual laser beam pass sensors of a coherent population trapping resonance based coupled dark state magnetometer

Michaela Ellmeier, Christoph Amtmann, Andreas Pollinger, Werner Magnes, Christian Hagen, Alexander Betzler, I. Jernej, Martín Agú, Laurentius Windholz, Roland Lammegger

Published Nov 26, 2022• 11 citations
Publisher

Abstract

The Coupled Dark State Magnetometer is an atomic magnetometer designed for magnetic field measurements in space. Its measurement principle is based on the excitation of coupled dark state resonances prepared via the coherent population trapping effect. The coupled dark state resonances enable the compensation of disturbing frequency shifts which degrade the measurement accuracy. Results with a sensor design based on a single laser beam pass through the sensor unit reveal a reduced ability to compensate these disturbing effects compared to theoretical predictions. Therefore, a new sensor design is presented which enables a dual laser beam pass through the sensor unit. For almost all angles between the external magnetic field and the light propagation direction, the dual pass sensor design enables a better compensation of disturbing frequency shifts than the single pass sensor design - up to a factor of nine.

DOI
10.1016/j.measen.2022.100606

Related papers

Control loops for a Coupled Dark State Magnetometer

Andreas Pollinger, Roland Lammegger, Werner Magnes +3 more

The Coupled Dark State Magnetometer (CDSM) is a new type of scalar magnetometer, which is based on two-photon spectroscopy of free alkali atoms. CDSM measurements do not depend on the sensor temperature. Furthermore, it has no dead zones, no moving parts and no excitation coils at the sensor. This paper deals with the control loops which are developed for tracking the magnetic field dependent resonances and for avoiding drifts of the carrier frequency of the laser diode used as excitation light source. It is the first time that a magnetometer of this type is operated in a control loop. The implementation of several functional blocks in a Field Programmable Gate Array (FPGA) is a first step towards miniaturization for a future space application. The magnetic field magnitude is measured by a frequency. Therefore, the Direct Digital Synthesis (DDS) principle is used for accurate signal generation. A noise floor measurement has shown no 1/f dependence near DC and 70 pT/√Hz up to the corner frequency of 3 Hz for which the CDSM is designed.

Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ...•Nov 1, 2010•12 citations

Sensitive magnetometers based on dark states

L. Moi, S. Cartaleva

24 In recent years interest has been growing into the fascinating properties of dark resonances. Most of the experiments have been carried out in alkali atoms due to their favourable level structure and the commercially available diode lasers, resulting in numerous applications. Teams from Siena and Sofia developed an all-optical magnetometer based on such dark resonances. The dark-state effect was discovered in 1976 by G. Alzetta, A. Gozzini, L. Moi and G. Orriols [1], while they were working on optical-pumping experiments with a multimode dye laser. They wanted to make the observation of multi-photon radio-frequency resonances in sodium atoms easy, i.e., observable by the naked eye. To this end, they introduced a magnetic field (MF) that was constant in time but spatially inhomogeneous along the laser beam, so as to produce Zeeman splitting of groundstate hyperfine levels (Fig.1a, left). In this way, an applied radio-frequency field was resonant only if it matched the Zeeman splitting. As a result, a bright radio-frequency resonance appeared (Fig.1a, right). During these experiments, besides the bright resonances, a new “dark” resonance was observed, which does not require any radio-frequency field. In the introductory illustration only the dark resonance is shown. The new dark state (DS) effect appears in a position along the laser beam where the frequency difference (ω10-ω20) between two optical frequencies is equal to the frequency splitting ω12 between two Zeeman sublevels (Fig.1b, left). This is the resonant condition for the stimulated Raman transitions between the two longliving ground levels. Theoretical studies described the dark resonance as a coherent superposition of two long-living ground levels prepared by the bi-chromatic laser field, and ‘atomic population trapping’ was introduced, leading to the term Coherent Population Trapping (CPT) [2]. l Luigi Moi1, Stefka Cartaleva2 DOI: 10.1051/epn/2012603

Europhysics news•Jul 1, 2012•8 citations

Dual-Axis Measurement and Suppression Method for Light Shift Based on a Single-Beam Compact SERF Atomic Magnetometer

Kexuan Miao, Tao Hu, Yan Chang +2 more

The light shift in a spin-exchange relaxation-free (SERF) atomic magnetometer (AM) generated by detuning of circularly polarized pump light, results in deviations of orthogonality measurement and attenuation of magnetometer response. In this paper, we propose and systematically investigate a dual-axis method to suppress light shift based on a self-developed single-beam compact SERF AM capable of measuring magnetic field components along two orthogonal sensitive axes. By sequentially observing the non-sensitive axis response and applying a magnetic field along the pump light direction, the virtual magnetic field induced by light shift is compensated and is rigorously demonstrated as a dominant component in this direction. Experimental results show a significant reduction in crosstalk between the dual sensitive axes, improving accuracy in resonant frequency determination of87Rb atoms and orthogonality evaluation of triaxial coils. Furthermore, the performance of dual-axis measurement in the SERF AM is tested, showing a more notable enhancement than that of single-axis measurement.

IEEE Sensors Journal•Jan 7, 2025•2 citations

Light shift suppression in a CPT magnetometer using linear polarization and double frequency interrogation

Mario Alberto González Maldonado, Yang Li, James McKelvy +5 more

We demonstrate a suppression of the light shift in a Coherent-Population-Trapping (CPT) atomic magnetometer by using linearly polarized light and a differential measurement between magnetic resonances. The radio frequency that creates the optical sidebands for CPT quickly switches between two magnetic sensitive transitions, and the magnetic field is extrapolated from the difference of the center frequencies of the magnetic resonances. Light shifts and common drifts like collisional shifts can be suppressed through careful choice of measured resonances, and we show the light shift reduction by more than a factor of 20 compared to excitation with circularly-polarized light. Various limitations of the method are discussed.

arXiv:2510.16159•Optics Letters•Dec 8, 2025•3 citations

Enable the inherent omni-directionality of an absolute coupled dark state magnetometer for e.g. scientific space applications

Andreas Pollinger, Michaela Ellmeier, Werner Magnes +4 more

The Coupled Dark State Magnetometer (CDSM) is an optically-pumped absolute scalar magnetometer which is based on two-photon spectroscopy of free alkali atoms. Several magnetic field dependent resonances occur in the presence of an external magnetic field, reaching their maximal strengths at different angles between the magnetic field direction and the reference axis of the sensor defined by the optical path of the laser excitation field. Dependent on this angle the strongest available resonances are selected. This enables an omni-directional dead zone free measurement. The paper discusses the measurement principle, describes the technical realization of the resonance switching, investigates the influence of this transition on the magnetic field measurement and proposes a method to detect the angle of the magnetic field direction in relation to the optical axis. The results presented within this paper show that dead zone free magnetic field measurements are possible with the CDSM without the need of a complex sensor design, e.g. additional moving parts or excitation coils.

IEEE International Instrumentation and Measurement Technology Conference•May 1, 2012•18 citations

Dead-zone free multi-beam atomic sensor for 4He magnetometer

Wei Gong, Haidong Wang, Teng Wu +5 more

A multi-beam atomic sensor with three orthogonal helium vapor cells to guarantee the availability of4He magnetometer under different directions of magnetic field is proposed and tested, which demonstrates experimentally that dead zones of magnetic field measurement can be eliminated. Furthermore, with the special optical design, the multi-beam atomic sensor is also used for laser frequency stabilization simultaneously, which shows the possibility of reducing heading errors caused by the light shift effect.

Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS)•May 1, 2016•2 citations
The Coupled Dark State Magnetometer is an atomic magnetometer designed for magnetic field measurements in space. Its measurement principle is based on the excitation of coupled dark state resonances prepared via the coherent population trapping effect. The coupled dark state resonances enable the compensation of disturbing frequency shifts which degrade the measurement accuracy. Results with a sensor design based on a single laser beam pass through the sensor unit reveal a reduced ability to compensate these disturbing effects compared to theoretical predictions. Therefore, a new sensor design is presented which enables a dual laser beam pass through the sensor unit. For almost all angles between the external magnetic field and the light propagation direction, the dual pass sensor design enables a better compensation of disturbing frequency shifts than the single pass sensor design - up to a factor of nine.
The Coupled Dark State Magnetometer (CDSM) is a new type of scalar magnetometer, which is based on two-photon spectroscopy of free alkali atoms. CDSM measurements do not depend on the sensor temperature. Furthermore, it has no dead zones, no moving parts and no excitation coils at the sensor. This paper deals with the control loops which are developed for tracking the magnetic field dependent resonances and for avoiding drifts of the carrier frequency of the laser diode used as excitation light source. It is the first time that a magnetometer of this type is operated in a control loop. The implementation of several functional blocks in a Field Programmable Gate Array (FPGA) is a first step towards miniaturization for a future space application. The magnetic field magnitude is measured by a frequency. Therefore, the Direct Digital Synthesis (DDS) principle is used for accurate signal generation. A noise floor measurement has shown no 1/f dependence near DC and 70 pT/√Hz up to the corner frequency of 3 Hz for which the CDSM is designed.
24 In recent years interest has been growing into the fascinating properties of dark resonances. Most of the experiments have been carried out in alkali atoms due to their favourable level structure and the commercially available diode lasers, resulting in numerous applications. Teams from Siena and Sofia developed an all-optical magnetometer based on such dark resonances. The dark-state effect was discovered in 1976 by G. Alzetta, A. Gozzini, L. Moi and G. Orriols [1], while they were working on optical-pumping experiments with a multimode dye laser. They wanted to make the observation of multi-photon radio-frequency resonances in sodium atoms easy, i.e., observable by the naked eye. To this end, they introduced a magnetic field (MF) that was constant in time but spatially inhomogeneous along the laser beam, so as to produce Zeeman splitting of groundstate hyperfine levels (Fig.1a, left). In this way, an applied radio-frequency field was resonant only if it matched the Zeeman splitting. As a result, a bright radio-frequency resonance appeared (Fig.1a, right). During these experiments, besides the bright resonances, a new “dark” resonance was observed, which does not require any radio-frequency field. In the introductory illustration only the dark resonance is shown. The new dark state (DS) effect appears in a position along the laser beam where the frequency difference (ω10-ω20) between two optical frequencies is equal to the frequency splitting ω12 between two Zeeman sublevels (Fig.1b, left). This is the resonant condition for the stimulated Raman transitions between the two longliving ground levels. Theoretical studies described the dark resonance as a coherent superposition of two long-living ground levels prepared by the bi-chromatic laser field, and ‘atomic population trapping’ was introduced, leading to the term Coherent Population Trapping (CPT) [2]. l Luigi Moi1, Stefka Cartaleva2 DOI: 10.1051/epn/2012603
The light shift in a spin-exchange relaxation-free (SERF) atomic magnetometer (AM) generated by detuning of circularly polarized pump light, results in deviations of orthogonality measurement and attenuation of magnetometer response. In this paper, we propose and systematically investigate a dual-axis method to suppress light shift based on a self-developed single-beam compact SERF AM capable of measuring magnetic field components along two orthogonal sensitive axes. By sequentially observing the non-sensitive axis response and applying a magnetic field along the pump light direction, the virtual magnetic field induced by light shift is compensated and is rigorously demonstrated as a dominant component in this direction. Experimental results show a significant reduction in crosstalk between the dual sensitive axes, improving accuracy in resonant frequency determination of87Rb atoms and orthogonality evaluation of triaxial coils. Furthermore, the performance of dual-axis measurement in the SERF AM is tested, showing a more notable enhancement than that of single-axis measurement.
We demonstrate a suppression of the light shift in a Coherent-Population-Trapping (CPT) atomic magnetometer by using linearly polarized light and a differential measurement between magnetic resonances. The radio frequency that creates the optical sidebands for CPT quickly switches between two magnetic sensitive transitions, and the magnetic field is extrapolated from the difference of the center frequencies of the magnetic resonances. Light shifts and common drifts like collisional shifts can be suppressed through careful choice of measured resonances, and we show the light shift reduction by more than a factor of 20 compared to excitation with circularly-polarized light. Various limitations of the method are discussed.
The Coupled Dark State Magnetometer (CDSM) is an optically-pumped absolute scalar magnetometer which is based on two-photon spectroscopy of free alkali atoms. Several magnetic field dependent resonances occur in the presence of an external magnetic field, reaching their maximal strengths at different angles between the magnetic field direction and the reference axis of the sensor defined by the optical path of the laser excitation field. Dependent on this angle the strongest available resonances are selected. This enables an omni-directional dead zone free measurement. The paper discusses the measurement principle, describes the technical realization of the resonance switching, investigates the influence of this transition on the magnetic field measurement and proposes a method to detect the angle of the magnetic field direction in relation to the optical axis. The results presented within this paper show that dead zone free magnetic field measurements are possible with the CDSM without the need of a complex sensor design, e.g. additional moving parts or excitation coils.
A multi-beam atomic sensor with three orthogonal helium vapor cells to guarantee the availability of4He magnetometer under different directions of magnetic field is proposed and tested, which demonstrates experimentally that dead zones of magnetic field measurement can be eliminated. Furthermore, with the special optical design, the multi-beam atomic sensor is also used for laser frequency stabilization simultaneously, which shows the possibility of reducing heading errors caused by the light shift effect.