Measurement Science and Technology

Optimization of the alkali-metal density ratio in a hybrid optical pumping atomic magnetometer

Yang Li, Xue-Jing Liu, Hongwei Cai, Ming Ding, Jang-Cheng Fang

Published Nov 9, 2018• 19 citations
Publisher

Abstract

Abstract The alkali-metal density ratio is one of the primary design parameters in hybrid optical pumping magnetometers. Here, experiments were conducted with a hybrid optically-pumped atomic magnetometer using a cell containing potassium and rubidium. Potassium atoms are pumped by a laser and become polarized. Rubidium atoms are polarized via collisions with potassium atoms and can be probed. The light shift was eliminated by adjusting the density ratio of potassium to rubidium atoms and by optimizing the pump laser parameters. The effect of fluctuations in the laser power and frequency were minimized as a result. The optimal density ratio and magnetic field sensitivity were 1:277 and 0.68 fT Hz −1/2 , respectively.

DOI
10.1088/1361-6501/aaefe2

Related papers

Sensitivity Improvement of Spin-Exchange Relaxation Free Atomic Magnetometers by Hybrid Optical Pumping of Potassium and Rubidium

Yosuke Ito, Hiroyuki Ohnishi, Keigo Kamada +1 more

An optically pumped atomic magnetometer using a hybrid cell of potassium and rubidium atoms was demonstrated to yield high sensitivity to magnetic fields. We operated the magnetometer with the four possible combinations of optically pumped and optically probed atoms and found that the combination of optically pumped potassium and optically probed rubidium showed the highest sensitivity among the four combinations because the rubidium atoms were denser than those of potassium. Furthermore, we investigated the dependence of the sensitivity on the power densities of the pump and probe beams and the wavelength of the probe beam. The magnetometer using the hybrid cell required higher pump-beam power and had narrower magnetic linewidth than those of the single alkali-metal cell. However, the magnetic linewidth was larger than the theoretical value, ignoring the spin relaxation caused by the spin-exchange collisions. By adjusting the laser conditions, the highest sensitivity approached 30 fTrms/Hz1/2.

IEEE Transactions on Magnetics•Oct 1, 2011•53 citations

Rate-equation approach to optimal density ratio of K-Rb hybrid cells for optically pumped atomic magnetometers

Yosuke Ito, Hiroyuki Ohnishi, Keigo Kamada +1 more

The goal of this study is to determine optimal conditions of optically pumped atomic magnetometers using a hybrid cell of potassium and rubidium atoms. We theoretically investigated the properties of the magnetometers and considered the optimal density ratio of potassium and rubidium atoms. We found that the experimental results agreed well with the theoretical estimations and the density ratio of potassium and rubidium atoms should be around 200 to achieve the highest sensitivity.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)•Jul 1, 2013•5 citations

Optimizations of spin-exchange relaxation-free magnetometer based on potassium and rubidium hybrid optical pumping

Jiancheng Fang, Tao Wang, Hong Zhang +2 more

The hybrid optical pumping atomic magnetometers have not realized its theoretical sensitivity, the optimization is critical for optimal performance. The optimizations proposed in this paper are suitable for hybrid optical pumping atomic magnetometer, which contains two alkali species. To optimize the parameters, the dynamic equations of spin evolution with two alkali species were solved, whose steady-state solution is used to optimize the parameters. The demand of the power of the pump beam is large for hybrid optical pumping. Moreover, the sensitivity of the hybrid optical pumping magnetometer increases with the increase of the power density of the pump beam. The density ratio between the two alkali species is especially important for hybrid optical pumping magnetometer. A simple expression for optimizing the density ratio is proposed in this paper, which can help to determine the mole faction of the alkali atoms in fabricating the hybrid cell before the cell is sealed. The spin-exchange rate between the two alkali species is proportional to the saturated density of the alkali vapor, which is highly dependent on the temperature of the cell. Consequently, the sensitivity of the hybrid optical pumping magnetometer is dependent on the temperature of the cell. We proposed the thermal optimization of the hybrid cell for a hybrid optical pumping magnetometer, which can improve the sensitivity especially when the power of the pump beam is low. With these optimizations, a sensitivity of approximately 5 fT/Hz(1/2) is achieved with gradiometer arrangement.

Review of Scientific Instruments•Dec 1, 2014•38 citations

Research on the influence of alkali polarization on an all-optical atomic spin magnetometer based on hybrid pumping potassium and rubidium

Hong Zhang, Sheng Zou, Xiyuan Chen +1 more

An all-optical atomic spin magnetometer using a hybrid cell of K and Rb atoms is demonstrated in this paper. Via deriving and solving the Bloch equations, we established the models for the polarization rate of Rb and the magnetometer’s output signal, respectively. In order to achieve the maximum sensitivity, we carried out research on the influence of the pump power on the polarization rate and magnetometer’s output signal. The magnetometer’s sensitivity was found to be proportional to the polarization rate of Rb, which was confined by the pump power. In a certain context, theoretical predictions indicated that the larger the pump power was, the higher magnetometer’s sensitivity that could be realized. However, the theoretical estimate deviated from the experimental results. An analysis of the errors between the actual value and the theoretical value showed that the different molar ratio between the two species of alkali-metal atoms was the primary reason causing the deviation. By comparing the experimental measurement result with simulation result, we found the difference in the molar ratio to be approximate 4 times. After various parameters had been optimized, a sensitivity of 18 fT/Hz 1/2 was achieved by analyzing the signal-to-noise ratio.

Journal of the Korean Physical Society•Apr 1, 2015•9 citations

Multi-Parameter Optimization of Rubidium Laser Optically Pumped Magnetometers with Geomagnetic Field Intensity

Kun Xu, Xiuyan Ren, Yujie Xiang +7 more

Rubidium laser optically pumped magnetometers (OPMs) are widely used magnetic sensors based on the Zeeman effect, laser pumping, and magnetic resonance principles. They measure the magnetic field by measuring the magnetic resonance signal passing through a rubidium atomic gas cell. The quality of the magnetic resonance signal is a necessary condition for a magnetometer to achieve high sensitivity. In this research, to obtain the best magnetic resonance signal of rubidium laser OPMs in the Earth’s magnetic field intensity, the experiment system of rubidium laser OPMs is built with a rubidium atomic gas cell as the core component. The linewidth and amplitude ratio (LAR) of magnetic resonance signals is utilized as the optimization objective function. The magnetic resonance signals of the magnetometer experiment system are experimentally measured for different laser frequencies, radio frequency (RF) intensities, laser powers, and atomic gas cell temperatures in a background magnetic field of 50,765 nT. The experimental results indicate that optimizing these parameters can reduce the LAR by one order of magnitude. This shows that the optimal parameter combination can effectively improve the sensitivity of the magnetometer. The sensitivity defined using the noise spectral density measured under optimal experimental parameters is 1.5 pT/Hz1/2@1 Hz. This work will provide key technical support for rubidium laser OPMs’ product development.

Sensors•Nov 2, 2023•6 citations

Effects of pump laser power density on the hybrid optically pumped comagnetometer for rotation sensing

Liwei Jiang, Wei Quan, Yixiang Liang +3 more

We investigate the effects of pump laser power density on the hybrid optically pumped comagnetometer operated in the spin-exchange relaxation-free (SERF) regime. The analytic steady-state output model for the comagnetometer considering two alkali metal species and one nuclear species is presented for the first time. And the effects of pump laser power density on the rotation sensitivity, suppression of low-frequency magnetic noise and long-term stability of the comagnetometer are studied experimentally. The results indicate that when the product of pumping rate and density ratio of pumped atom to probed atom is equal to the spin relaxation rate of the probed atom, the maximum response and highest sensitivity of the comagnetometer are achieved. However, the suppression of low-frequency magnetic noise and long-term stability improve with the increasing of pump laser power density due to the increasing of nuclear spin polarization. Our focus is to optimize the performance of the comagnetometer for rotation sensing, but the theory and method presented here are relevant to all applications of the hybrid optical pumping technique.

Optics Express•Sep 16, 2019•62 citations
Abstract The alkali-metal density ratio is one of the primary design parameters in hybrid optical pumping magnetometers. Here, experiments were conducted with a hybrid optically-pumped atomic magnetometer using a cell containing potassium and rubidium. Potassium atoms are pumped by a laser and become polarized. Rubidium atoms are polarized via collisions with potassium atoms and can be probed. The light shift was eliminated by adjusting the density ratio of potassium to rubidium atoms and by optimizing the pump laser parameters. The effect of fluctuations in the laser power and frequency were minimized as a result. The optimal density ratio and magnetic field sensitivity were 1:277 and 0.68 fT Hz −1/2 , respectively.
An optically pumped atomic magnetometer using a hybrid cell of potassium and rubidium atoms was demonstrated to yield high sensitivity to magnetic fields. We operated the magnetometer with the four possible combinations of optically pumped and optically probed atoms and found that the combination of optically pumped potassium and optically probed rubidium showed the highest sensitivity among the four combinations because the rubidium atoms were denser than those of potassium. Furthermore, we investigated the dependence of the sensitivity on the power densities of the pump and probe beams and the wavelength of the probe beam. The magnetometer using the hybrid cell required higher pump-beam power and had narrower magnetic linewidth than those of the single alkali-metal cell. However, the magnetic linewidth was larger than the theoretical value, ignoring the spin relaxation caused by the spin-exchange collisions. By adjusting the laser conditions, the highest sensitivity approached 30 fTrms/Hz1/2.
The hybrid optical pumping atomic magnetometers have not realized its theoretical sensitivity, the optimization is critical for optimal performance. The optimizations proposed in this paper are suitable for hybrid optical pumping atomic magnetometer, which contains two alkali species. To optimize the parameters, the dynamic equations of spin evolution with two alkali species were solved, whose steady-state solution is used to optimize the parameters. The demand of the power of the pump beam is large for hybrid optical pumping. Moreover, the sensitivity of the hybrid optical pumping magnetometer increases with the increase of the power density of the pump beam. The density ratio between the two alkali species is especially important for hybrid optical pumping magnetometer. A simple expression for optimizing the density ratio is proposed in this paper, which can help to determine the mole faction of the alkali atoms in fabricating the hybrid cell before the cell is sealed. The spin-exchange rate between the two alkali species is proportional to the saturated density of the alkali vapor, which is highly dependent on the temperature of the cell. Consequently, the sensitivity of the hybrid optical pumping magnetometer is dependent on the temperature of the cell. We proposed the thermal optimization of the hybrid cell for a hybrid optical pumping magnetometer, which can improve the sensitivity especially when the power of the pump beam is low. With these optimizations, a sensitivity of approximately 5 fT/Hz(1/2) is achieved with gradiometer arrangement.
An all-optical atomic spin magnetometer using a hybrid cell of K and Rb atoms is demonstrated in this paper. Via deriving and solving the Bloch equations, we established the models for the polarization rate of Rb and the magnetometer’s output signal, respectively. In order to achieve the maximum sensitivity, we carried out research on the influence of the pump power on the polarization rate and magnetometer’s output signal. The magnetometer’s sensitivity was found to be proportional to the polarization rate of Rb, which was confined by the pump power. In a certain context, theoretical predictions indicated that the larger the pump power was, the higher magnetometer’s sensitivity that could be realized. However, the theoretical estimate deviated from the experimental results. An analysis of the errors between the actual value and the theoretical value showed that the different molar ratio between the two species of alkali-metal atoms was the primary reason causing the deviation. By comparing the experimental measurement result with simulation result, we found the difference in the molar ratio to be approximate 4 times. After various parameters had been optimized, a sensitivity of 18 fT/Hz 1/2 was achieved by analyzing the signal-to-noise ratio.
Rubidium laser optically pumped magnetometers (OPMs) are widely used magnetic sensors based on the Zeeman effect, laser pumping, and magnetic resonance principles. They measure the magnetic field by measuring the magnetic resonance signal passing through a rubidium atomic gas cell. The quality of the magnetic resonance signal is a necessary condition for a magnetometer to achieve high sensitivity. In this research, to obtain the best magnetic resonance signal of rubidium laser OPMs in the Earth’s magnetic field intensity, the experiment system of rubidium laser OPMs is built with a rubidium atomic gas cell as the core component. The linewidth and amplitude ratio (LAR) of magnetic resonance signals is utilized as the optimization objective function. The magnetic resonance signals of the magnetometer experiment system are experimentally measured for different laser frequencies, radio frequency (RF) intensities, laser powers, and atomic gas cell temperatures in a background magnetic field of 50,765 nT. The experimental results indicate that optimizing these parameters can reduce the LAR by one order of magnitude. This shows that the optimal parameter combination can effectively improve the sensitivity of the magnetometer. The sensitivity defined using the noise spectral density measured under optimal experimental parameters is 1.5 pT/Hz1/2@1 Hz. This work will provide key technical support for rubidium laser OPMs’ product development.
We investigate the effects of pump laser power density on the hybrid optically pumped comagnetometer operated in the spin-exchange relaxation-free (SERF) regime. The analytic steady-state output model for the comagnetometer considering two alkali metal species and one nuclear species is presented for the first time. And the effects of pump laser power density on the rotation sensitivity, suppression of low-frequency magnetic noise and long-term stability of the comagnetometer are studied experimentally. The results indicate that when the product of pumping rate and density ratio of pumped atom to probed atom is equal to the spin relaxation rate of the probed atom, the maximum response and highest sensitivity of the comagnetometer are achieved. However, the suppression of low-frequency magnetic noise and long-term stability improve with the increasing of pump laser power density due to the increasing of nuclear spin polarization. Our focus is to optimize the performance of the comagnetometer for rotation sensing, but the theory and method presented here are relevant to all applications of the hybrid optical pumping technique.