Physical Review Letters
Xe 129 -Rb Spin-Exchange Optical Pumping with High Photon Efficiency
Graham Norquay, Guilhem J. Collier, Madhwesha R. Rao, Neil J. Stewart, Jim M. Wild
Abstract
Here we present a $\mathrm{Rb}\text{\ensuremath{-}}^{129}\mathrm{Xe}$ spin-exchange optical pumping polarizer capable of rapid generation of large volumes of highly polarized $^{129}\mathrm{Xe}$ gas. Through modeling and measurements we maximize the $^{129}\mathrm{Xe}$ nuclear spin polarization output to enable the generation of polarized $^{129}\mathrm{Xe}$ gas imaging volumes ($300\text{ }\text{ }{\mathrm{cm}}^{3}$) every 5 min within a clinical setting. Our model is verified by experiment to correctly predict the optimum Rb vapor density for maximum $^{129}\mathrm{Xe}$ nuclear polarization for a flux $3.4\text{ }\text{ }\mathrm{W}/{\mathrm{cm}}^{2}$ of circularly polarized Rb ${D}_{1}$ photons incident on an 80 cm long cylindrical optical cell. We measure a $^{129}\mathrm{Xe}$ magnetization production efficiency of ${\ensuremath{\eta}}_{\mathrm{pr}}=1.8%$, which approaches the photon efficiency limit ${\ensuremath{\eta}}_{\ensuremath{\gamma}}=3.3%$ of this system and enables the polarization of $2.72\ifmmode\times\else\texttimes\fi{}{10}^{22}$ $^{129}\mathrm{Xe}$ spins per hour, corresponding to $1013\text{ }\text{ }{\mathrm{cm}}^{3}$ of 100% polarized $^{129}\mathrm{Xe}$ at STP. This magnetization production rate is threefold higher than the highest previously published $^{129}\mathrm{Xe}$ magnetization production rate and has enabled routine clinical lung magnetic resonance imaging (MRI) with hyperpolarized $^{129}\mathrm{Xe}$ doses available on demand at run time, as well as high-SNR $^{129}\mathrm{Xe}$ MRI of the human brain and kidneys.