Measurement Science and Technology
Dual radiofrequency drive quantum voltage standard with nanovolt resolution based on a closed-loop refrigeration cycle
D. Georgakopoulos, Ilya F. Budovsky, Thomas Hagen, Hajime SASAKI, Hirotake Yamamori
Abstract
We have developed a programmable Josephson voltage standard that can produce voltages up to 20 V with a resolution of better than 0.1 µV over the whole voltage range and better than 1 nV for voltages up to 10 mV. The standard has two superconductor–normal metal–superconductor junction arrays connected in series and driven by two radiofrequency oscillators. The cryogenic part of the standard is based on a cryocooler. The new standard agrees with the primary quantum voltage standard maintained at the National Measurement Institute, Australia, within 10 nV and forms the basis of an automated calibration system for digital multimeters and voltage references.