A novel absorption spectrometer where an absorber is placed in one beam of a parametric oscillator is described. The sensitivity is not shot-noise limited and may be enhanced by the operation of the parametric oscillator close to threshold.
Nondegenerate optical parametric oscillators generate above-threshold signal and idler beams that have intensity fluctuations correlated at the quantum level (twin beams). We describe what is to our knowledge the first high-sensitivity spectroscopy experiment using twin beams emitted by a cw optical parametric oscillator: a very weak two-photon absorption signal, in the 10(-7) range, is recorded on the 4S(1/2)-5S(1/2) transition of atomic potassium with a noise background that is reduced by 1.9 dB with respect to the shot-noise limit of the light used in the experiment.
We investigate the limits of intracavity absorption spectroscopy with nonlinear media. Using a common theoretical framework, we compare the detection of a trace gas within (i) an undriven cavity with gain near and above threshold; (ii) a driven cavity with gain kept just below threshold; and (iii) a cavity driven close to the saturation point of a saturable absorber. These phase-transition-based metrology methods are typically quantum limited by spontaneous emission, and we compare them to the empty-cavity shot-noise-limited case. Although the fundamental limits achievable with nonlinear media do not surpass the empty-cavity limits, we show that nonlinear methods are more robust against certain technical noise models. This recognition may have applications in spectrometer design for devices operating in nonideal field environments.