arXiv:2611.00277

Finesse evolution in a bichromatic cavity under blue and red resonant illumination in vacuum

M. S. Caracas Nuñez, A. Kadakkavattathu, G. Gatta, L. Duca, N. Mizukami, C. Sias

physics.atom-phcond-mat.quant-gasphysics.opticsquant-phSubmitted Oct 30, 2026• Announced Nov 3, 2026
arXivHTMLPDF

Abstract

The long-term performance of high-finesse optical cavities depends on the stability of their mirror coatings under vacuum and under continuous exposure to high circulating power. In cavities supporting multiple wavelengths, a central question is whether degradation at one wavelength compromises performance at another. Understanding this interplay is essential for those cavity quantum electrodynamics experiments that need diverse optical fields for cavity stabilization, trapping, and light--matter coupling. Here, we investigate wavelength-dependent degradation and recovery of a bichromatic cavity with high-reflectivity coating at 455 nm and 671 nm in vacuum. Cavity ring-down measurements reveal a decrease in the 455 nm finesse from 7.7*10^3 to 1.9*10^3 after 11 hours of continuous exposure to high circulating power. Simultaneous measurements show no measurable degradation at 671 nm. Continuous operation at 671 nm, instead, produces no measurable degradation at either wavelength under the tested conditions. Comparison of oxygen depletion and Avrami-type loss models indicates kinetics compatible with nucleation and growth, without uniquely identifying the microscopic mechanism. We demonstrate partial recovery of the finesse with illumination at 455 nm in air. Our results indicate a wavelength dependence of laser-induced optical damage in bichromatic cavities and provide an estimate of the blue light intensity that can be safely used in such cavities.

AI Summary

This paper tests whether illumination-induced loss in one reflection band of a bichromatic optical cavity affects its other band, and whether red illumination damages blue-band performance. A vacuum Fabry–Pérot cavity with $\mathrm{SiO_2}/\mathrm{Ta_2O_5}$ coatings supports resonances at 455 and 671 nm. Finesse is tracked by cavity ring-down during prolonged resonant illumination and subsequent recovery in air. Under 455 nm illumination, the blue finesse falls from $(7.7\pm0.7)\times10^3$ to $(1.9\pm0.2)\times10^3$ after 11 hours of accumulated illumination. A second run monitoring both wavelengths shows blue finesse decreasing from $(4.82\pm0.07)\times10^3$ to $(2.4\pm0.2)\times10^3$ over 9.3 hours, while red finesse remains near $1.6\times10^4$. No measurable degradation occurs during intervals without circulating light. Continuous 671 nm illumination for 11.3 and 13.5 hours produces no measurable degradation at either wavelength. Blue illumination in air partially restores blue finesse, from approximately $2.0\times10^3$ to $2.8\times10^3$ over 20.4 hours, without a measurable change at 671 nm. An Avrami-type loss model with fixed exponent $n=3$ describes blue degradation better than the tested first-order oxygen-depletion model. This is compatible with nucleation-and-growth kinetics but does not identify the microscopic mechanism; hydrocarbon contamination is suggested rather than established. Recovery kinetics are likewise phenomenological. Stability conclusions apply to the tested coating and illumination conditions. The proposed blue operating intensity of roughly $5\,\mathrm{kW/cm^2}$ is inferred from the degraded steady state rather than established by a systematic damage-threshold measurement.

Little sign of AI-generated textIncrementalOverclaims

Error found: The claimed safe blue operating intensity of approximately 5 kW/cm² is not adequately substantiated: its circulating-power calculation is not reproducible from the reported measurements, and a finesse plateau after degradation does not establish damage-free operation at that intensity.

Written by gpt-6.1-sol from the paper's full text. Automated: it can be wrong.

Comments
12 pages, 5 figures, comments welcome

Related papers

Stability of a high-finesse optical cavity at 493 nm in vacuum for cavity QED with barium ions

Diptaranjan Das, Ezra Kassa, Hiroki Takahashi

We explore the stability of a high-finesse optical cavity at 493 nm in vacuum for cavity QED with barium ions. A high-finesse Fabry-Perot cavity is built using mirrors with high-reflectivity (HR) coatings that are implemented by stacking multiple thin films of low-loss dielectrics on substrates. Applications of such HR mirrors in the near ultraviolet (UV) range have been hampered by degradation of coatings in vacuum. Here, we explore the degradation of mirrors with HR coatings at 493 nm in vacuum. We study both vacuum-induced and laser-induced effects on oxide-coated cavity mirrors by probing changes in cavity loss using cavity lifetime measurements. We investigate the role of circulating power in the rate of increase in cavity loss and demonstrate methods of reversal of cavity degradation. While we observe no degradation without long exposure or with short exposures at lower circulating powers, we find evidence of degradation on long exposure to high circulating powers. We discuss potential causes and conclude that laser-induced deposition is the likely cause while ruling out thermally activated processes due to laser-induced heating.

Applied Optics•Apr 27, 2026•2 citations

Stability of a high-finesse optical cavity at 493 nm in vacuum for cavity QED with Barium ions

Diptaranjan Das, Ezra Kassa, Hiroki Takahashi

We explore the stability of a high-finesse optical cavity at 493 nm in vacuum for cavity QED with Barium ions. A high-finesse Fabry-Perot cavity is built using mirrors with high-reflectivity (HR) coatings that are implemented by stacking multiple thin films of low-loss dielectrics on substrates. Applications of such HR mirrors in the near ultraviolet (UV) range have been hampered by degradation of coatings in vacuum. Here, we explore the degradation of mirrors with HR coatings at 493 nm in vacuum. We study both vacuum-induced and laser-induced effects on oxide-coated cavity mirrors by probing changes in cavity loss using cavity lifetime measurements. We investigate the role of circulating power in the rate of increase in cavity loss and demonstrate methods of reversal of cavity degradation. While we observe no degradation without long exposure or with short exposures at lower circulating powers, we find evidence of degradation on long exposure to high circulating powers. We discuss potential causes and conclude that laser-induced deposition is the likely cause while ruling out thermally activated processes due to laser-induced heating.

physics.opticsCross-listedarXiv:2603.17638•Mar 19, 2026•0 citations

Preventing and reversing vacuum-induced optical losses in high-finesse tantalum (V) oxide mirror coatings

Dorian A. Gangloff, Molu Shi, Tailin Wu +11 more

High-finesse optical cavities placed under vacuum are foundational platforms in quantum information science with photons and atoms. We study the vacuum-induced degradation of high-finesse optical cavities with mirror coatings composed of SiO₂-Ta₂O₅ dielectric stacks, and present methods to protect these coatings and to recover their initial low loss levels. For separate coatings with reflectivities centered at 370 nm and 422 nm, a vacuum-induced continuous increase in optical loss occurs if the surface-layer coating is made of Ta₂O₅, while it does not occur if it is made of SiO₂. The incurred optical loss can be reversed by filling the vacuum chamber with oxygen at atmospheric pressure, and the recovery rate can be strongly accelerated by continuous laser illumination at 422 nm. Both the degradation and the recovery processes depend strongly on temperature. We find that a 1 nm-thick layer of SiO₂ passivating the Ta₂O₅ surface layer is sufficient to reduce the degradation rate by more than a factor of 10, strongly supporting surface oxygen depletion as the primary degradation mechanism.

arXiv:1505.03381•Optics Express•Jul 1, 2015•33 citations

Multi-Watt cavity for 266 nm light in vacuum

Christian Brand, Christian Knobloch, Ksenija Simonović +1 more

Abstract Intense coherent ultraviolet radiation is gaining increasing importance in advanced quantum technologies—from optical clocks and quantum computers to matter-wave interferometry—as well as in photochemistry, life sciences, semiconductor industry, and space applications. Since the preparation of multi-Watt light sources is still an open challenge for many ultraviolet wavelengths, resonant enhancement in a cavity is an attractive alternative. However, many experiments with atoms, molecules or nanoparticles require isolation in high vacuum where UV optics often show fast degradation. Here, we present stable performance of a cavity for 266 nm light with several Watt of intra-cavity power in high vacuum despite the presence of hydrocarbons. Comparing two sets of cavity mirrors indicates that this feat is connected to the micro-chemical environment at the topmost coating layer. Our study emphasizes the need for further developments in this direction to facilitate robust, compact, and high-performing devices employing UV radiation.

Physica Scripta•Jun 29, 2023•1 citation

Novel characterization of an optical cavity with small mode volume

Moonjoo Lee, Myunghun Kim, Jungsoo Hong +4 more

We characterize a high-finesse Fabry-Perot resonator for coupling with single neutral atoms. Our cavity consists of two mirrors with different reflectivities: One has minimal optical loss, and the other high transmission loss where more than 90% of the intracavity photons would be emitted. Cavity finesse, birefringent effects, and mechanical resonances are measured using the lasers at 780, 782, and 795 nm. In order to obtain cavity geometric parameters, we drive the adjacent longitudinal or transverse modes with two lasers simultaneously, and measure those frequencies using a precision wavelength meter (WLM). A major novelty of this method is that the parameters' uncertainty is solely determined by the resolution of the WLM, eliminating all of the temporal environment fluctuations. Moreover, the technique with two lasers consists of a vital approach for determining geometric parameters of a short cavity, with a free spectral range on the order of THz. Our system operates in the strong atom-cavity coupling regime that allows us to explore fundamental quantum optics and implement quantum network protocols.

•Feb 11, 2021•1 citation

Precision measurement of resonate frequency and the effective cavity length of the high finesse optical micro-cavity

Jinjin Du, LI Wen-fang, Ruijuan Wen +2 more

Ultra-high finesse micro-resonator plays an important role in realizing the interaction between atoms and cavity field in the study of cavity quantum electrodynamics (QED) system, weak optical nonlinear effects and micro-optic devices. By measuring basic parameters of the microcavity, the atom-cavity coupling coefficient and the cavity decay rate can be determined precisely. It is also useful for exploring the dynamic characteristics of the system. However, it has difficulty in determining resonate frequency and effective cavity length due to the structure of the ultra-high finesse optical microcavity itself and the characteristics of multilayer coating. In this paper, we demonstrate the precision measurement of effective cavity length under different resonant frequencies which our cavity mirror is coated with 37 layers of dielectric film. The theoretical expectation when using the revised model of the multilayer coating agrees well with that of the experiment; and the measurement error for longitudinal mode interval is below 0.004 nm which is two orders of magnitude better than that obtained in previous unrevised model. The tiny depths into mirror coatings that the standing-wave light field inside the cavity penetrates are given for different mode numbers. This method may be applied to other micro resonator in the precision measurement.

Acta Physica Sinica•Jan 1, 2013•8 citations
The long-term performance of high-finesse optical cavities depends on the stability of their mirror coatings under vacuum and under continuous exposure to high circulating power. In cavities supporting multiple wavelengths, a central question is whether degradation at one wavelength compromises performance at another. Understanding this interplay is essential for those cavity quantum electrodynamics experiments that need diverse optical fields for cavity stabilization, trapping, and light--matter coupling. Here, we investigate wavelength-dependent degradation and recovery of a bichromatic cavity with high-reflectivity coating at 455 nm and 671 nm in vacuum. Cavity ring-down measurements reveal a decrease in the 455 nm finesse from 7.710^3 to 1.910^3 after 11 hours of continuous exposure to high circulating power. Simultaneous measurements show no measurable degradation at 671 nm. Continuous operation at 671 nm, instead, produces no measurable degradation at either wavelength under the tested conditions. Comparison of oxygen depletion and Avrami-type loss models indicates kinetics compatible with nucleation and growth, without uniquely identifying the microscopic mechanism. We demonstrate partial recovery of the finesse with illumination at 455 nm in air. Our results indicate a wavelength dependence of laser-induced optical damage in bichromatic cavities and provide an estimate of the blue light intensity that can be safely used in such cavities.
This paper tests whether illumination-induced loss in one reflection band of a bichromatic optical cavity affects its other band, and whether red illumination damages blue-band performance. A vacuum Fabry–Pérot cavity with SiO2/Ta2O5\mathrm{SiO_2}/\mathrm{Ta_2O_5}SiO2​/Ta2​O5​ coatings supports resonances at 455 and 671 nm. Finesse is tracked by cavity ring-down during prolonged resonant illumination and subsequent recovery in air.
Under 455 nm illumination, the blue finesse falls from (7.7±0.7)×103(7.7\pm0.7)\times10^3(7.7±0.7)×103 to (1.9±0.2)×103(1.9\pm0.2)\times10^3(1.9±0.2)×103 after 11 hours of accumulated illumination. A second run monitoring both wavelengths shows blue finesse decreasing from (4.82±0.07)×103(4.82\pm0.07)\times10^3(4.82±0.07)×103 to (2.4±0.2)×103(2.4\pm0.2)\times10^3(2.4±0.2)×103 over 9.3 hours, while red finesse remains near 1.6×1041.6\times10^41.6×104. No measurable degradation occurs during intervals without circulating light. Continuous 671 nm illumination for 11.3 and 13.5 hours produces no measurable degradation at either wavelength. Blue illumination in air partially restores blue finesse, from approximately 2.0×1032.0\times10^32.0×103 to 2.8×1032.8\times10^32.8×103 over 20.4 hours, without a measurable change at 671 nm.
An Avrami-type loss model with fixed exponent n=3n=3n=3 describes blue degradation better than the tested first-order oxygen-depletion model. This is compatible with nucleation-and-growth kinetics but does not identify the microscopic mechanism; hydrocarbon contamination is suggested rather than established. Recovery kinetics are likewise phenomenological. Stability conclusions apply to the tested coating and illumination conditions. The proposed blue operating intensity of roughly 5 kW/cm25\,\mathrm{kW/cm^2}5kW/cm2 is inferred from the degraded steady state rather than established by a systematic damage-threshold measurement.
We explore the stability of a high-finesse optical cavity at 493 nm in vacuum for cavity QED with barium ions. A high-finesse Fabry-Perot cavity is built using mirrors with high-reflectivity (HR) coatings that are implemented by stacking multiple thin films of low-loss dielectrics on substrates. Applications of such HR mirrors in the near ultraviolet (UV) range have been hampered by degradation of coatings in vacuum. Here, we explore the degradation of mirrors with HR coatings at 493 nm in vacuum. We study both vacuum-induced and laser-induced effects on oxide-coated cavity mirrors by probing changes in cavity loss using cavity lifetime measurements. We investigate the role of circulating power in the rate of increase in cavity loss and demonstrate methods of reversal of cavity degradation. While we observe no degradation without long exposure or with short exposures at lower circulating powers, we find evidence of degradation on long exposure to high circulating powers. We discuss potential causes and conclude that laser-induced deposition is the likely cause while ruling out thermally activated processes due to laser-induced heating.
We explore the stability of a high-finesse optical cavity at 493 nm in vacuum for cavity QED with Barium ions. A high-finesse Fabry-Perot cavity is built using mirrors with high-reflectivity (HR) coatings that are implemented by stacking multiple thin films of low-loss dielectrics on substrates. Applications of such HR mirrors in the near ultraviolet (UV) range have been hampered by degradation of coatings in vacuum. Here, we explore the degradation of mirrors with HR coatings at 493 nm in vacuum. We study both vacuum-induced and laser-induced effects on oxide-coated cavity mirrors by probing changes in cavity loss using cavity lifetime measurements. We investigate the role of circulating power in the rate of increase in cavity loss and demonstrate methods of reversal of cavity degradation. While we observe no degradation without long exposure or with short exposures at lower circulating powers, we find evidence of degradation on long exposure to high circulating powers. We discuss potential causes and conclude that laser-induced deposition is the likely cause while ruling out thermally activated processes due to laser-induced heating.
High-finesse optical cavities placed under vacuum are foundational platforms in quantum information science with photons and atoms. We study the vacuum-induced degradation of high-finesse optical cavities with mirror coatings composed of SiO₂-Ta₂O₅ dielectric stacks, and present methods to protect these coatings and to recover their initial low loss levels. For separate coatings with reflectivities centered at 370 nm and 422 nm, a vacuum-induced continuous increase in optical loss occurs if the surface-layer coating is made of Ta₂O₅, while it does not occur if it is made of SiO₂. The incurred optical loss can be reversed by filling the vacuum chamber with oxygen at atmospheric pressure, and the recovery rate can be strongly accelerated by continuous laser illumination at 422 nm. Both the degradation and the recovery processes depend strongly on temperature. We find that a 1 nm-thick layer of SiO₂ passivating the Ta₂O₅ surface layer is sufficient to reduce the degradation rate by more than a factor of 10, strongly supporting surface oxygen depletion as the primary degradation mechanism.
Abstract Intense coherent ultraviolet radiation is gaining increasing importance in advanced quantum technologies—from optical clocks and quantum computers to matter-wave interferometry—as well as in photochemistry, life sciences, semiconductor industry, and space applications. Since the preparation of multi-Watt light sources is still an open challenge for many ultraviolet wavelengths, resonant enhancement in a cavity is an attractive alternative. However, many experiments with atoms, molecules or nanoparticles require isolation in high vacuum where UV optics often show fast degradation. Here, we present stable performance of a cavity for 266 nm light with several Watt of intra-cavity power in high vacuum despite the presence of hydrocarbons. Comparing two sets of cavity mirrors indicates that this feat is connected to the micro-chemical environment at the topmost coating layer. Our study emphasizes the need for further developments in this direction to facilitate robust, compact, and high-performing devices employing UV radiation.
We characterize a high-finesse Fabry-Perot resonator for coupling with single neutral atoms. Our cavity consists of two mirrors with different reflectivities: One has minimal optical loss, and the other high transmission loss where more than 90% of the intracavity photons would be emitted. Cavity finesse, birefringent effects, and mechanical resonances are measured using the lasers at 780, 782, and 795 nm. In order to obtain cavity geometric parameters, we drive the adjacent longitudinal or transverse modes with two lasers simultaneously, and measure those frequencies using a precision wavelength meter (WLM). A major novelty of this method is that the parameters' uncertainty is solely determined by the resolution of the WLM, eliminating all of the temporal environment fluctuations. Moreover, the technique with two lasers consists of a vital approach for determining geometric parameters of a short cavity, with a free spectral range on the order of THz. Our system operates in the strong atom-cavity coupling regime that allows us to explore fundamental quantum optics and implement quantum network protocols.
Ultra-high finesse micro-resonator plays an important role in realizing the interaction between atoms and cavity field in the study of cavity quantum electrodynamics (QED) system, weak optical nonlinear effects and micro-optic devices. By measuring basic parameters of the microcavity, the atom-cavity coupling coefficient and the cavity decay rate can be determined precisely. It is also useful for exploring the dynamic characteristics of the system. However, it has difficulty in determining resonate frequency and effective cavity length due to the structure of the ultra-high finesse optical microcavity itself and the characteristics of multilayer coating. In this paper, we demonstrate the precision measurement of effective cavity length under different resonant frequencies which our cavity mirror is coated with 37 layers of dielectric film. The theoretical expectation when using the revised model of the multilayer coating agrees well with that of the experiment; and the measurement error for longitudinal mode interval is below 0.004 nm which is two orders of magnitude better than that obtained in previous unrevised model. The tiny depths into mirror coatings that the standing-wave light field inside the cavity penetrates are given for different mode numbers. This method may be applied to other micro resonator in the precision measurement.