arXiv:2606.23151

An Al$^+$ clock with $1.6\times10^{-18}$ systematic uncertainty and its frequency ratios

Fabian Dawel, Johannes Krämer, D. Drapier, Lennart Pelzer, Kai Dietze, M Haider Ali, Marek B. Hild, Vincent Barbé, S A King, J. Klose, Kilian Stahl, Johannes Rahm, Navraj Poudel, Sören Dörscher, S. Weyers, Erik Benkler, Christian Lisdat, Piet O. Schmidt

Published Jun 22, 2026• 0 citations
arXivHTMLPDF

Abstract

Advances in optical clocks motivate a redefinition of the second, requiring rigorous evaluations of systematic uncertainties and robust consistency among the clocks. Here, we report the full evaluation of the systematic frequency shifts of an $^{27}\mathrm{Al}^+$ single-ion clock, and the measurement of its absolute frequency and frequency ratio with a $^{87}$Sr optical lattice clock at PTB. The evaluated total systematic fractional frequency uncertainty is $1.6\times10^{-18}$, mainly limited by the accuracy of the relevant atomic coefficients and by background gas collisions. The absolute frequency of the clock has been measured to be $ν_{\mathrm{Al}^+}=1 121 015 393 207 859.19(24)\,$Hz, obtained by comparison with two primary caesium fountain clocks at PTB. The frequency ratio between the Al and Sr optical clocks has been determined to be $ν_{\mathrm{Al}^+}/ν_{^{87}\mathrm{Sr}}=2.611 701 431 781 462 668(36)$, limited by the accuracy of the Sr clock. This ratio differs by $8.6σ$ and $1.2σ$ from the 2021 and 2025 frequency ratio published by the BACON collaboration, respectively. These results represent an important contribution toward a future redefinition of the second using optical clocks, and underscore the importance of independent measurements of clock-candidate frequency ratios across different institutions.

DOI
10.48550/arxiv.2606.23151

Related papers

Atomic Clock Frequency Ratios with Fractional Uncertainty ≤ 3.2 × 10 − 18

Alexander Aeppli, Willa J. Arthur-Dworschack, K. Beloy +24 more

We report high-precision frequency ratio measurements between optical atomic clocks based on ^{27}Al^{+}, ^{171}Yb, and ^{87}Sr. With total fractional uncertainties at or below 3.2×10^{-18}, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in ^{87}Sr ratios at approximately 1×10^{-16} and the Al^{+}/Yb ratio at 1.6×10^{-17} in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key upgrade from our previous work is the use of a common ultrastable reference delivered to all clocks via a 3.6-km phase-stabilized fiber link between two institutions, enabling better accuracy and stability in frequency transfer. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2-3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of 1.3×10^{-16} at 1 s. Identifying individual clock stabilities with a three-corner-hat measurement, we demonstrate the most stable optical lattice and trapped-ion clocks used in a multispecies comparison. By enabling faster comparisons, this stability will improve sensitivity to nonwhite noise processes and other underlying limits of state-of-the-art optical frequency standards.

Physical Review Letters•May 28, 2026•2 citations

Atomic clock frequency ratios with fractional uncertainty $\leq 3.2 \times 10^{-18}$

Alexander Aeppli, Willa J. Arthur-Dworschack, Kyle Beloy +24 more

We report high-precision frequency ratio measurements between optical atomic clocks based on $^{27}$Al$^+$, $^{171}$Yb, and $^{87}$Sr. With total fractional uncertainties at or below $3.2 \times 10^{-18}$, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in $^{87}$Sr ratios at approximately $1\times10^{-16}$ and the Al$^+$/Yb ratio at $1.6\times10^{-17}$ in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key innovation in this work is the use of a common ultrastable reference delivered to all clocks via a 3.6 km phase-stabilized fiber link between two institutions. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2 to 3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of $1.3 \times 10^{-16}$ at 1 second. By enabling faster comparisons, this stability will improve sensitivity to non-white noise processes and other underlying limits of state-of-the-art optical frequency standards.

physics.atom-phCross-listedarXiv:2512.21428•Dec 26, 2025

Frequency ratio measurement of Al⁺ at the 10⁻¹⁸ level

Fabian Dawel

The invention of the atomic clock has enabled applications like the global navigation satellite system, are important for telecommunication and are used for the definition of the SI second. Nowadays, optical clocks allow for measurement of height differences on the centimeter scale and can be employed for tests of the standard model of particle physics. Among the various implementations of optical clocks, this thesis focuses on an Al⁺ clock. For this work an optical resonator is examined, which enables the simultaneous frequency stabilization of two laser on the same mirror pair through the use of a dual-wavelength coating. Various contributions on the frequency stability are characterized and correlations are investigated. This cavity enables high fidelities in coherent manipulation of trapped ions, as required to operate the Al⁺ clock. This thesis partly builds upon the findings presented in Ref. Kramer, 2023. The setup and the systematic shifts are revisited and a detailed evaluation of the shifts is conducted, resulting in a total fractional frequency uncertainty of 1.6×10⁻¹⁸. The biggest contribution is from an ac-Stark shift induced by cooling during the clock interrogation, resulting in a shift of −9.3(1.1) × 10⁻¹⁸. The cooling reduces the second order Doppler shift to −1.6(0.2)×10⁻¹⁸. Additionally, a position dependent micromotion of the ion was investigated, which contributes an uncertainty of 0.38 × 10⁻¹⁸. Moreover, the first-order Doppler shift, servo error, ac-Stark shift of the clock laser and the phase chirp are determined for the first time, showing uncertainties below 10⁻¹⁸. The first frequency comparison between the Al⁺ clock and the ¹¹⁵In⁺, ¹⁷¹Yb⁺ and ⁸⁷Sr clocks at PTB are demonstrated. Measurements against Sr reveal a stability of the Al+ clock of 5.9 × 10⁻¹⁶ at 1s , which is one of the best single ion clock stabilities to date. All measured frequency ratios fall within the 10⁻¹⁸ uncertainty regime. The ratio between Sr/Al⁺ exhibits a significant deviation of more than 8.6𝜎 from published measurement data, underscoring the importance of inter-institutional frequency ratio measurements to verify the uncertainty budgets and reproducibility of optical clocks.

•Jan 1, 2026•0 citations

Frequency Ratio Measurements with 18-digit Accuracy Using a Network of Optical Clocks

Boulder Atomic Clock Optical Network, Collaboration, : +35 more

Atomic clocks occupy a unique position in measurement science, exhibiting higher accuracy than any other measurement standard and underpinning six out of seven base units in the SI system. By exploiting higher resonance frequencies, optical atomic clocks now achieve greater stability and lower frequency uncertainty than existing primary standards. Here, we report frequency ratios of the $^{27}$Al$^+$, $^{171}$Yb and $^{87}$Sr optical clocks in Boulder, Colorado, measured across an optical network spanned by both fiber and free-space links. These ratios have been evaluated with measurement uncertainties between $6\times10^{-18}$ and $8\times10^{-18}$, making them the most accurate reported measurements of frequency ratios to date. This represents a critical step towards redefinition of the SI second and future applications such as relativistic geodesy and tests of fundamental physics.

arXiv:2005.14694•May 29, 2020•13 citations

Frequency Ratio of Hg 199 and Sr 87 Optical Lattice Clocks beyond the SI Limit

K. Yamanaka, Noriaki Ohmae, Ichiro Ushijima +2 more

We report on a frequency ratio measurement of a $^{199}\mathrm{Hg}$-based optical lattice clock referencing a $^{87}\mathrm{Sr}$-based clock. Evaluations of lattice light shift, including atomic-motion-dependent shift, enable us to achieve a total systematic uncertainty of $7.2\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$ for the Hg clock. The frequency ratio is measured to be ${\ensuremath{\nu}}_{\mathrm{Hg}}/{\ensuremath{\nu}}_{\mathrm{Sr}}=2.629\text{ }314\text{ }209\text{ }898\text{ }909\text{ }60(22)$ with a fractional uncertainty of $8.4\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$, which is smaller than the uncertainty of the realization of the International System of Units (SI) second, i.e., the SI limit.

Physical Review Letters•Jun 10, 2015•104 citations

Direct comparison of multi-ion optical clocks based on $^{40}$Ca$^+$ and $^{88}$Sr$^+$

Yosef Sokolik, Roee Ozeri, Nitzan Akerman

We report the first direct frequency comparison between two multi-ion optical clocks based on the S$_{1/2}$ to D$_{5/2}$ transition in \Ca and \Sr ions. Using linear chains of up to nine \Ca ions and six \Sr ions, we demonstrate improved stability as a function of the number of ions that are contributing to the laser frequency stabilization servo. The measured joint fractional frequency stability of the two clocks reaches $1.37(12)\times 10^{-15}$ at one second, placing an upper bound on the same stability of one of the clocks at $9.6(8)\times 10^{-16}$ in one second. We measured the frequency ratio of the two clocks to be $R_{\text{Sr/Ca}}=1.082076536381896986(18)$, where the systematic uncertainty is primarily limited by the room temperature blackbody radiation. Our direct measurement represents an order of magnitude improvement compared to existing indirect frequency ratio measurements. Furthermore, by combining our results with recent absolute frequency measurements of the \Sr transition, referenced to a primary frequency standard, we refined the absolute frequency of the \Ca transition to $ν_{\text{Ca}^+}=411042129776400.21(4)$ Hz, reducing its uncertainty by a factor of three. This study presents the first direct comparison between two multi-ion optical clocks, highlighting their significant potential for future applications in fundamental physics tests, geodesy, and precision metrology.

arXiv:2602.05626•Feb 5, 2026•0 citations
An Al+^++ clock with 1.6×10−181.6\times10^{-18}1.6×10−18 systematic uncertainty and its frequency ratios
Advances in optical clocks motivate a redefinition of the second, requiring rigorous evaluations of systematic uncertainties and robust consistency among the clocks. Here, we report the full evaluation of the systematic frequency shifts of an 27Al+^{27}\mathrm{Al}^+27Al+ single-ion clock, and the measurement of its absolute frequency and frequency ratio with a 87^{87}87Sr optical lattice clock at PTB. The evaluated total systematic fractional frequency uncertainty is 1.6×10−181.6\times10^{-18}1.6×10−18, mainly limited by the accuracy of the relevant atomic coefficients and by background gas collisions. The absolute frequency of the clock has been measured to be νAl+=1121015393207859.19(24) ν_{\mathrm{Al}^+}=1 121 015 393 207 859.19(24)\,νAl+​=1121015393207859.19(24)Hz, obtained by comparison with two primary caesium fountain clocks at PTB. The frequency ratio between the Al and Sr optical clocks has been determined to be νAl+/ν87Sr=2.611701431781462668(36)ν_{\mathrm{Al}^+}/ν_{^{87}\mathrm{Sr}}=2.611 701 431 781 462 668(36)νAl+​/ν87Sr​=2.611701431781462668(36), limited by the accuracy of the Sr clock. This ratio differs by 8.6σ8.6σ8.6σ and 1.2σ1.2σ1.2σ from the 2021 and 2025 frequency ratio published by the BACON collaboration, respectively. These results represent an important contribution toward a future redefinition of the second using optical clocks, and underscore the importance of independent measurements of clock-candidate frequency ratios across different institutions.
We report high-precision frequency ratio measurements between optical atomic clocks based on ^{27}Al^{+}, ^{171}Yb, and ^{87}Sr. With total fractional uncertainties at or below 3.2×10^{-18}, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in ^{87}Sr ratios at approximately 1×10^{-16} and the Al^{+}/Yb ratio at 1.6×10^{-17} in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key upgrade from our previous work is the use of a common ultrastable reference delivered to all clocks via a 3.6-km phase-stabilized fiber link between two institutions, enabling better accuracy and stability in frequency transfer. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2-3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of 1.3×10^{-16} at 1 s. Identifying individual clock stabilities with a three-corner-hat measurement, we demonstrate the most stable optical lattice and trapped-ion clocks used in a multispecies comparison. By enabling faster comparisons, this stability will improve sensitivity to nonwhite noise processes and other underlying limits of state-of-the-art optical frequency standards.
Atomic clock frequency ratios with fractional uncertainty ≤3.2×10−18\leq 3.2 \times 10^{-18}≤3.2×10−18
We report high-precision frequency ratio measurements between optical atomic clocks based on 27^{27}27Al+^++, 171^{171}171Yb, and 87^{87}87Sr. With total fractional uncertainties at or below 3.2×10−183.2 \times 10^{-18}3.2×10−18, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in 87^{87}87Sr ratios at approximately 1×10−161\times10^{-16}1×10−16 and the Al+^++/Yb ratio at 1.6×10−171.6\times10^{-17}1.6×10−17 in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key innovation in this work is the use of a common ultrastable reference delivered to all clocks via a 3.6 km phase-stabilized fiber link between two institutions. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2 to 3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of 1.3×10−161.3 \times 10^{-16}1.3×10−16 at 1 second. By enabling faster comparisons, this stability will improve sensitivity to non-white noise processes and other underlying limits of state-of-the-art optical frequency standards.
The invention of the atomic clock has enabled applications like the global navigation satellite system, are important for telecommunication and are used for the definition of the SI second. Nowadays, optical clocks allow for measurement of height differences on the centimeter scale and can be employed for tests of the standard model of particle physics. Among the various implementations of optical clocks, this thesis focuses on an Al⁺ clock. For this work an optical resonator is examined, which enables the simultaneous frequency stabilization of two laser on the same mirror pair through the use of a dual-wavelength coating. Various contributions on the frequency stability are characterized and correlations are investigated. This cavity enables high fidelities in coherent manipulation of trapped ions, as required to operate the Al⁺ clock. This thesis partly builds upon the findings presented in Ref. Kramer, 2023. The setup and the systematic shifts are revisited and a detailed evaluation of the shifts is conducted, resulting in a total fractional frequency uncertainty of 1.6×10⁻¹⁸. The biggest contribution is from an ac-Stark shift induced by cooling during the clock interrogation, resulting in a shift of −9.3(1.1) × 10⁻¹⁸. The cooling reduces the second order Doppler shift to −1.6(0.2)×10⁻¹⁸. Additionally, a position dependent micromotion of the ion was investigated, which contributes an uncertainty of 0.38 × 10⁻¹⁸. Moreover, the first-order Doppler shift, servo error, ac-Stark shift of the clock laser and the phase chirp are determined for the first time, showing uncertainties below 10⁻¹⁸. The first frequency comparison between the Al⁺ clock and the ¹¹⁵In⁺, ¹⁷¹Yb⁺ and ⁸⁷Sr clocks at PTB are demonstrated. Measurements against Sr reveal a stability of the Al+ clock of 5.9 × 10⁻¹⁶ at 1s , which is one of the best single ion clock stabilities to date. All measured frequency ratios fall within the 10⁻¹⁸ uncertainty regime. The ratio between Sr/Al⁺ exhibits a significant deviation of more than 8.6𝜎 from published measurement data, underscoring the importance of inter-institutional frequency ratio measurements to verify the uncertainty budgets and reproducibility of optical clocks.
Atomic clocks occupy a unique position in measurement science, exhibiting higher accuracy than any other measurement standard and underpinning six out of seven base units in the SI system. By exploiting higher resonance frequencies, optical atomic clocks now achieve greater stability and lower frequency uncertainty than existing primary standards. Here, we report frequency ratios of the 27^{27}27Al+^++, 171^{171}171Yb and 87^{87}87Sr optical clocks in Boulder, Colorado, measured across an optical network spanned by both fiber and free-space links. These ratios have been evaluated with measurement uncertainties between 6×10−186\times10^{-18}6×10−18 and 8×10−188\times10^{-18}8×10−18, making them the most accurate reported measurements of frequency ratios to date. This represents a critical step towards redefinition of the SI second and future applications such as relativistic geodesy and tests of fundamental physics.
We report on a frequency ratio measurement of a 199Hg^{199}\mathrm{Hg}199Hg-based optical lattice clock referencing a 87Sr^{87}\mathrm{Sr}87Sr-based clock. Evaluations of lattice light shift, including atomic-motion-dependent shift, enable us to achieve a total systematic uncertainty of $7.2\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$ for the Hg clock. The frequency ratio is measured to be ${\ensuremath{\nu}}_{\mathrm{Hg}}/{\ensuremath{\nu}}_{\mathrm{Sr}}=2.629\text{ }314\text{ }209\text{ }898\text{ }909\text{ }60(22)$ with a fractional uncertainty of $8.4\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$, which is smaller than the uncertainty of the realization of the International System of Units (SI) second, i.e., the SI limit.
Direct comparison of multi-ion optical clocks based on 40^{40}40Ca+^++ and 88^{88}88Sr+^++
We report the first direct frequency comparison between two multi-ion optical clocks based on the S1/2_{1/2}1/2​ to D5/2_{5/2}5/2​ transition in \Ca and \Sr ions. Using linear chains of up to nine \Ca ions and six \Sr ions, we demonstrate improved stability as a function of the number of ions that are contributing to the laser frequency stabilization servo. The measured joint fractional frequency stability of the two clocks reaches 1.37(12)×10−151.37(12)\times 10^{-15}1.37(12)×10−15 at one second, placing an upper bound on the same stability of one of the clocks at 9.6(8)×10−169.6(8)\times 10^{-16}9.6(8)×10−16 in one second. We measured the frequency ratio of the two clocks to be RSr/Ca=1.082076536381896986(18)R_{\text{Sr/Ca}}=1.082076536381896986(18)RSr/Ca​=1.082076536381896986(18), where the systematic uncertainty is primarily limited by the room temperature blackbody radiation. Our direct measurement represents an order of magnitude improvement compared to existing indirect frequency ratio measurements. Furthermore, by combining our results with recent absolute frequency measurements of the \Sr transition, referenced to a primary frequency standard, we refined the absolute frequency of the \Ca transition to νCa+=411042129776400.21(4)ν_{\text{Ca}^+}=411042129776400.21(4)νCa+​=411042129776400.21(4) Hz, reducing its uncertainty by a factor of three. This study presents the first direct comparison between two multi-ion optical clocks, highlighting their significant potential for future applications in fundamental physics tests, geodesy, and precision metrology.