Overcoming the thermal-noise limit of microwave measurements by precooling with an active cold load

Anonymous

Published Dec 23, 2024• 3 citations
arXivHTMLPDF

Abstract

We introduce a method, which we here name ``active pre-cooling'' (APC), for removing, just prior to performing a measurement, a large fraction of the deleterious thermal photons that would otherwise occupy the electromagnetic modes of a microwave cavity or some alternative form of radio-frequency resonator. The removal is achieved by temporarily over-coupling the cavity's modes to an active cold load (ACL). We report a room-temperature bench-top demonstration of the method, where this load takes the form of the input of a commercial low-noise amplifier (LNA). No isolator is inserted between the LNA's input and the cavity's coupling port. The noise temperature of a monitored microwave mode drops to 123 K. Upon incorporating our pre-coolable cavity into a time-resolved (tr-) EPR spectrometer, a commensurate improvement in the signal-to-noise ratio is observed, corresponding to a factor-of-5 speed up over a conventional tr-EPR measurement at room temperature for the same precision and/or sensitivity. Modeling indicates the feasibility, for realistic mode quality factors and couplings, of cooling the room-temperature cavity's modes down to a few tens of K, and for this coldness to last several tens of microseconds, whilst the cavity and its contents are optimally interrogated by a microwave tone or pulse sequence. The method thus provides a simple and generally applicable approach to improving the sensitivity and/or read-out speed in pulsed and time-resolved EPR spectroscopy, quantum detection and other radiometric measurements. It provides a first-stage cold reservoir (of microwave photons) for other, deeper cooling methods to work from and, through the use of cryogenic ACLs (realized, in the first instance, as the inputs of existing cryogenic low-noise amplifiers), the method could itself be directly extended to lower temperature regimes.

DOI
10.1103/bjky-3yb3

Related papers

Room-temperature quantum devices: innovations in active pre-cooling protocol, spin-phonon coupling devices, and stable molecular spins

Kuan‐Cheng Chen

The feasibility of active pre-cooling (APC) by temporarily over-coupling microwave cavity modes to a low-noise amplifier acting as an active cold load was demonstrated, offering an effective method to reduce thermal photon occupancy without cryogenic cooling. The potential of the organic radical BDPA as a tunable qubit medium with remarkably long spin-lattice and spin coherence relaxation times was assessed, comparing favorably with state-of-the-art molecular qubit media. A high-quality surface acoustic wave device integrated with pentacene-deposited molecular thin films was developed to achieve heterogeneous optically-detected spin acoustic resonance (HODSAR), enabling efficient zero-field spin control at room temperature via spin-phonon coupling. Through the construction of specialized experimental setups, the effectiveness of these techniques was demonstrated, facilitating pre-cooling of cavity modes, enhancing coherence times in qubit media, and achieving precise spin manipulation. While APC offers advantages over methods like masers and anti-masers, challenges such as the need for fast, low-loss microwave switches remain. The BDPA radicals showed significant promise for spin ensemble, and the SAW-based HODSAR device provides a scalable platform for spin manipulation in quantum technologies.

•Nov 4, 2024•0 citations

Bench-Top Cooling of a Microwave Mode Using an Optically Pumped Spin Refrigerator

Hao Wu, Shamil Mirkhanov, Wern Ng +1 more

We experimentally demonstrate the temporary removal of thermal photons from a microwave mode at 1.45 GHz through its interaction with the spin-polarized triplet states of photo-excited pentacene molecules doped within a p-terphenyl crystal at room temperature. The crystal functions electromagnetically as a narrowband cryogenic load, removing photons from the otherwise room-temperature mode via stimulated absorption. The noise temperature of the microwave mode dropped to 50_{-32}^{+18} K (as directly inferred by noise-power measurements), while the metal walls of the cavity enclosing the mode remained at room temperature. Simulations based on the same system's behavior as a maser (which could be characterized more accurately) indicate the possibility of the mode's temperature sinking to ∼10 K (corresponding to ∼140 microwave photons). These observations, when combined with engineering improvements to deepen the cooling, identify the system as a narrowband yet extremely convenient platform-free of cryogenics, vacuum chambers, and strong magnets-for realizing low-noise detectors, quantum memory, and quantum-enhanced machines (such as heat engines) based on strong spin-photon coupling and entanglement at microwave frequencies.

arXiv:2104.00268•Physical Review Letters•Jul 29, 2021•28 citations

Low-Noise Amplification, Detection and Spectroscopy of Ultra-Cold Systems in RF Cavities

M. H. S. Bukhari, Zahoor H. Shah

The design and development of a cryogenic Ultra-Low-Noise Signal Amplification (ULNA) and detection system for spectroscopy of ultra-cold systems are reported here for the operation in the 0.5 - 4 GHz spectrum of frequencies (the “L” and “S” microwave bands). The design is suitable for weak RF signal detection and spectroscopy from ultra-cold systems confined in cryogenic RF cavities, as entailed in a number of physics, physical chemistry and analytical chemistry applications, such as NMR/NQR/EPR and microwave spectroscopy, Paul traps, Bose-Einstein Condensates (BEC’s) and cavity Quantum Electrodynamics (cQED). Using a generic Low-Noise Amplifier (LNA) architecture for a GaAs enhancement mode High-Electron Mobility FET device, our design has especially been devised for scientific applications where ultra-low-noise amplification systems are sought to amplify and detect weak RF signals under various conditions and environments, including cryogenic temperatures, with the least possible noise susceptibility. The amplifier offers a 16 dB gain and a 0.8 dB noise figure at 2.5 GHz, while operating at room temperature, which can improve significantly at low temperatures. Both dc and RF outputs are provided by the amplifier to integrate it in a closed-loop or continuous-wave spectroscopy system or connect it to a variety of instruments, a factor which is lacking in commercial LNA devices. Following the amplification stage, the RF signal detection is carried out with the help of a post-amplifier and detection system based upon a set of Zero-Bias Schottky Barrier Diodes (ZBD’s) and a high-precision ultra-low noise jFET operational amplifier. The scheme offers unique benefits of sensitive detection and very-low noise amplification for measuring extremely weak on-resonance signals with substantial low- noise response and excellent stability while eliminating complicated and expensive heterodyne schemes. The LNA stage is fully capable to be a part of low-temperature experiments while being operated in cryogenic conditions down to about 500 mK.

Modern Instrumentation•Jan 1, 2016•4 citations

Quantum noise squeezing at microwave frequencies

R. Movshovich, B. Yurke, P. G. Kaminsky +3 more

We report the observation of zero-point noise squeezing by means of a Josephson- junction parametric amplifier, which has been used previously to squeeze 4.2 K thermal noise1. To observe quantum noise squeezing, a number of improvements were made on the apparatus; these include the in stallation of two low-noise cryogenic high- electron-mobility-transistor (HEMT) amplifiers, which boosted the detector sensitivity an order of magnitude to 215 K. The Josephson- junction parametric amplifier was cooled to 30 mK and was operated in the degenerate mode with the signal carrier frequency at 19.16 GHz. At this frequency the vacuum noise floor is hv/2k = 0.46 K. In the deamplified quadrature a drop in the noise of 0.223 K was observed relative to the pump-off noise floor. The pump-off noise floor was established to be within 3.4 + 2.0% of the vacuum noise floor by varying the temperature of the cold termination at the parametric amplifier's input from 30 mK to 1K. A probe signal at the input of the first HEMT amplifier was used to establish that the detector system saturation was less than 7 × 10−4 dB. We have thus observed squeezing 47% ± 8% below the vacuum noise floor.

Annual Meeting Optical Society of America•Jan 1, 1990•0 citations

Device-Agnostic Microwave Noise Metrology for Nonlinear Cryogenic Quantum Devices

Andrea Celotto, Alessandro Alocco, Bernardo Galvano +6 more

Microwave devices capable of near-quantum-limited signal processing are essential components in the toolbox of solid-state quantum technologies. The manipulation and readout of single-photon microwave signals through amplifiers, mixers, isolators, etc. must fulfill strict requirements in terms of signal integrity to ensure reliable operation. These active microwave quantum devices operate in complex cryo-electronic setups. This poses challenges to their characterization, since all relevant figures of merit must be expressed at the reference planes of their ports. Even though cryogenic S-parameter calibration is non-trivial, metrological approaches are converging toward rigorous methods. Furthermore, preserving signal integrity must be quantified via absolute noise levels at the ports of the Device Under Test (DUT), requiring an absolute power reference. In this work, we present an in situ noise metrology protocol based on substituting a controllable noise source for the DUT. We motivate this choice by showing that placing the noise source at the DUT input impacts the separability of the calibration from the DUT characteristics. Our proposed architecture combines Planck spectroscopy using a Variable Temperature Stage with Short-Open-Load-Reciprocal scattering-parameter calibration, so that noise and scattering quantities are referred to the same cryogenic reference planes. In this configuration, the readout-chain calibration is separated from the internal dynamics of the DUT. As a demanding use case, we apply the protocol to a Josephson Traveling Wave Parametric Amplifier and extract its gain and input-referred added noise under pump conditions activating multimode nonlinear behavior. This illustrates how our device-agnostic protocol supports portable noise characterization of nonlinear cryogenic microwave devices.

quant-pharXiv:2605.28808•May 28, 2026•1 citation

Observation of the fundamental Nyquist noise limit in an ultra-high Q-factor cryogenic bulk acoustic wave cavity

Maxim Goryachev, Eugene N. Ivanov, Frank J. van Kann +2 more

Thermal Nyquist noise fluctuations of high-Q bulk acoustic wave cavities have been observed at cryogenic temperatures with a DC superconducting quantum interference device amplifier. High Q modes with bandwidths of few tens of milliHz produce thermal fluctuations with a signal-to-noise ratio of up to 23 dB. The estimated effective temperature from the Nyquist noise is in good agreement with the physical temperature of the device, confirming the validity of the equivalent circuit model and the non-existence of any excess resonator self-noise. The measurements also confirm that the quality factor remains extremely high (Q > 108 at low order overtones) for very weak (thermal) system motion at low temperatures, when compared to values measured with relatively strong external excitation. This result represents an enabling step towards operating such a high-Q acoustic device at the standard quantum limit.

arXiv:1410.4293•Applied Physics Letters•Oct 13, 2014•41 citations
We introduce a method, which we here name ``active pre-cooling'' (APC), for removing, just prior to performing a measurement, a large fraction of the deleterious thermal photons that would otherwise occupy the electromagnetic modes of a microwave cavity or some alternative form of radio-frequency resonator. The removal is achieved by temporarily over-coupling the cavity's modes to an active cold load (ACL). We report a room-temperature bench-top demonstration of the method, where this load takes the form of the input of a commercial low-noise amplifier (LNA). No isolator is inserted between the LNA's input and the cavity's coupling port. The noise temperature of a monitored microwave mode drops to 123 K. Upon incorporating our pre-coolable cavity into a time-resolved (tr-) EPR spectrometer, a commensurate improvement in the signal-to-noise ratio is observed, corresponding to a factor-of-5 speed up over a conventional tr-EPR measurement at room temperature for the same precision and/or sensitivity. Modeling indicates the feasibility, for realistic mode quality factors and couplings, of cooling the room-temperature cavity's modes down to a few tens of K, and for this coldness to last several tens of microseconds, whilst the cavity and its contents are optimally interrogated by a microwave tone or pulse sequence. The method thus provides a simple and generally applicable approach to improving the sensitivity and/or read-out speed in pulsed and time-resolved EPR spectroscopy, quantum detection and other radiometric measurements. It provides a first-stage cold reservoir (of microwave photons) for other, deeper cooling methods to work from and, through the use of cryogenic ACLs (realized, in the first instance, as the inputs of existing cryogenic low-noise amplifiers), the method could itself be directly extended to lower temperature regimes.
The feasibility of active pre-cooling (APC) by temporarily over-coupling microwave cavity modes to a low-noise amplifier acting as an active cold load was demonstrated, offering an effective method to reduce thermal photon occupancy without cryogenic cooling. The potential of the organic radical BDPA as a tunable qubit medium with remarkably long spin-lattice and spin coherence relaxation times was assessed, comparing favorably with state-of-the-art molecular qubit media. A high-quality surface acoustic wave device integrated with pentacene-deposited molecular thin films was developed to achieve heterogeneous optically-detected spin acoustic resonance (HODSAR), enabling efficient zero-field spin control at room temperature via spin-phonon coupling. Through the construction of specialized experimental setups, the effectiveness of these techniques was demonstrated, facilitating pre-cooling of cavity modes, enhancing coherence times in qubit media, and achieving precise spin manipulation. While APC offers advantages over methods like masers and anti-masers, challenges such as the need for fast, low-loss microwave switches remain. The BDPA radicals showed significant promise for spin ensemble, and the SAW-based HODSAR device provides a scalable platform for spin manipulation in quantum technologies.
We experimentally demonstrate the temporary removal of thermal photons from a microwave mode at 1.45 GHz through its interaction with the spin-polarized triplet states of photo-excited pentacene molecules doped within a p-terphenyl crystal at room temperature. The crystal functions electromagnetically as a narrowband cryogenic load, removing photons from the otherwise room-temperature mode via stimulated absorption. The noise temperature of the microwave mode dropped to 50_{-32}^{+18} K (as directly inferred by noise-power measurements), while the metal walls of the cavity enclosing the mode remained at room temperature. Simulations based on the same system's behavior as a maser (which could be characterized more accurately) indicate the possibility of the mode's temperature sinking to ∼10 K (corresponding to ∼140 microwave photons). These observations, when combined with engineering improvements to deepen the cooling, identify the system as a narrowband yet extremely convenient platform-free of cryogenics, vacuum chambers, and strong magnets-for realizing low-noise detectors, quantum memory, and quantum-enhanced machines (such as heat engines) based on strong spin-photon coupling and entanglement at microwave frequencies.
The design and development of a cryogenic Ultra-Low-Noise Signal Amplification (ULNA) and detection system for spectroscopy of ultra-cold systems are reported here for the operation in the 0.5 - 4 GHz spectrum of frequencies (the “L” and “S” microwave bands). The design is suitable for weak RF signal detection and spectroscopy from ultra-cold systems confined in cryogenic RF cavities, as entailed in a number of physics, physical chemistry and analytical chemistry applications, such as NMR/NQR/EPR and microwave spectroscopy, Paul traps, Bose-Einstein Condensates (BEC’s) and cavity Quantum Electrodynamics (cQED). Using a generic Low-Noise Amplifier (LNA) architecture for a GaAs enhancement mode High-Electron Mobility FET device, our design has especially been devised for scientific applications where ultra-low-noise amplification systems are sought to amplify and detect weak RF signals under various conditions and environments, including cryogenic temperatures, with the least possible noise susceptibility. The amplifier offers a 16 dB gain and a 0.8 dB noise figure at 2.5 GHz, while operating at room temperature, which can improve significantly at low temperatures. Both dc and RF outputs are provided by the amplifier to integrate it in a closed-loop or continuous-wave spectroscopy system or connect it to a variety of instruments, a factor which is lacking in commercial LNA devices. Following the amplification stage, the RF signal detection is carried out with the help of a post-amplifier and detection system based upon a set of Zero-Bias Schottky Barrier Diodes (ZBD’s) and a high-precision ultra-low noise jFET operational amplifier. The scheme offers unique benefits of sensitive detection and very-low noise amplification for measuring extremely weak on-resonance signals with substantial low- noise response and excellent stability while eliminating complicated and expensive heterodyne schemes. The LNA stage is fully capable to be a part of low-temperature experiments while being operated in cryogenic conditions down to about 500 mK.
We report the observation of zero-point noise squeezing by means of a Josephson- junction parametric amplifier, which has been used previously to squeeze 4.2 K thermal noise1. To observe quantum noise squeezing, a number of improvements were made on the apparatus; these include the in stallation of two low-noise cryogenic high- electron-mobility-transistor (HEMT) amplifiers, which boosted the detector sensitivity an order of magnitude to 215 K. The Josephson- junction parametric amplifier was cooled to 30 mK and was operated in the degenerate mode with the signal carrier frequency at 19.16 GHz. At this frequency the vacuum noise floor is hv/2k = 0.46 K. In the deamplified quadrature a drop in the noise of 0.223 K was observed relative to the pump-off noise floor. The pump-off noise floor was established to be within 3.4 + 2.0% of the vacuum noise floor by varying the temperature of the cold termination at the parametric amplifier's input from 30 mK to 1K. A probe signal at the input of the first HEMT amplifier was used to establish that the detector system saturation was less than 7 × 10−4 dB. We have thus observed squeezing 47% ± 8% below the vacuum noise floor.
Microwave devices capable of near-quantum-limited signal processing are essential components in the toolbox of solid-state quantum technologies. The manipulation and readout of single-photon microwave signals through amplifiers, mixers, isolators, etc. must fulfill strict requirements in terms of signal integrity to ensure reliable operation. These active microwave quantum devices operate in complex cryo-electronic setups. This poses challenges to their characterization, since all relevant figures of merit must be expressed at the reference planes of their ports. Even though cryogenic S-parameter calibration is non-trivial, metrological approaches are converging toward rigorous methods. Furthermore, preserving signal integrity must be quantified via absolute noise levels at the ports of the Device Under Test (DUT), requiring an absolute power reference. In this work, we present an in situ noise metrology protocol based on substituting a controllable noise source for the DUT. We motivate this choice by showing that placing the noise source at the DUT input impacts the separability of the calibration from the DUT characteristics. Our proposed architecture combines Planck spectroscopy using a Variable Temperature Stage with Short-Open-Load-Reciprocal scattering-parameter calibration, so that noise and scattering quantities are referred to the same cryogenic reference planes. In this configuration, the readout-chain calibration is separated from the internal dynamics of the DUT. As a demanding use case, we apply the protocol to a Josephson Traveling Wave Parametric Amplifier and extract its gain and input-referred added noise under pump conditions activating multimode nonlinear behavior. This illustrates how our device-agnostic protocol supports portable noise characterization of nonlinear cryogenic microwave devices.
Thermal Nyquist noise fluctuations of high-Q bulk acoustic wave cavities have been observed at cryogenic temperatures with a DC superconducting quantum interference device amplifier. High Q modes with bandwidths of few tens of milliHz produce thermal fluctuations with a signal-to-noise ratio of up to 23 dB. The estimated effective temperature from the Nyquist noise is in good agreement with the physical temperature of the device, confirming the validity of the equivalent circuit model and the non-existence of any excess resonator self-noise. The measurements also confirm that the quality factor remains extremely high (Q > 108 at low order overtones) for very weak (thermal) system motion at low temperatures, when compared to values measured with relatively strong external excitation. This result represents an enabling step towards operating such a high-Q acoustic device at the standard quantum limit.