COMPARATIVE TEST OF TWO METHODS OF QUANTUM EFFICIENCY ABSOLUTE MEASUREMENT BASED ON SQUEEZED VACUUM DIRECT DETECTION

I. N. Agafonov, Maria V. Chekhova, Alexander N. Penin, Georgy O. Rytikov, O. A. Shumilkina, T. Sh. Iskhakov

Published Jan 1, 2011• 7 citations
arXivHTMLPDF

Abstract

We realize and test in experiment a method recently proposed for measuring absolute quantum efficiency of analog photodetectors. Similar to the traditional (Klyshko) method of absolute calibration, the new one is based on the direct detection of two-mode squeezed vacuum at the output of a traveling wave OPA. However, in the new method, one measures the difference-photocurrent variance rather than the correlation function of photocurrents (number of coincidences), which makes the technique applicable for high-gain OPA. In this work we test the new method versus the traditional one for the case of photon-counting detectors where both techniques are valid.

DOI
10.1142/s0219749911007265

Related papers

Absolute calibration of photodetectors: photocurrent multiplication versus photocurrent subtraction

I. N. Agafonov, Maria V. Chekhova, T. Sh. Iskhakov +3 more

We report testing of the new absolute method of photodetector calibration based on the difference-signal measurement for two-mode squeezed vacuum by comparison with the traditional absolute method based on coincidence counting. Using low-gain parametric downconversion, we have measured the quantum efficiency of a counting detector by both methods. The difference-signal method was adapted for the counting detectors by taking into account the dead-time effect.

arXiv:1012.3248•Optics Letters•Apr 4, 2011•20 citations

Measurement of the Quantum Efficiency of Analog Detectors in the Parametric Down-Conversion Field

T. I. Novikova, A.A. Leontyev, G. Kh. Kitaeva

Approaches to the standard-free calibration of the quantum efficiency of a wide class of analog detectors based on the measurement of statistical characteristics of fields generated by parametric down-conversion are analyzed. General expressions are obtained for the noise reduction factor of the difference photocurrent and for the covariance of photocurrents in the signal and idler channels that take into account the possibility of strong fluctuations in the amplitudes of single-photon response functions of the detectors used. It is shown that the measurement of the noise reduction factor of the difference photocurrent using detectors that cannot operate in the photon counting mode is on its own insufficient to directly characterize the level of two-mode squeezing in the down-conversion field and to determine the quantum efficiency of photosensitive elements without additional calibration procedures. A method to determine the quantum efficiency of such detectors based on measuring the dependence of the normalized covariance of photocurrents on the parametric gain is proposed.

Journal of Experimental and Theoretical Physics Letters•Sep 1, 2022•3 citations

Elementary responses of analog photodetectors and absolute calibration of their efficiency

Daniil A. Safronenkov, Pavel A. Prudkovskii, A. V. Osipenkov +1 more

A modified Klyshko method for standardless calibration of analog photodetector efficiencies is suggested and experimentally implemented. The method is based on the measurement of the number of incident photons using correlation properties of biphotons generated under spontaneous parametric downconversion, and on the determination of the mean number of discrete elementary responses of a photodetector by deconvoluting the statistical distribution of the output photocurrents. As an example, the generalized efficiency of an analog photomultiplier is measured as the ratio of the average number of elementary current responses to the average number of incident photons. The statistical properties of the elementary responses of an analog photodetector depend not only on the detector itself, but also on the parameters of the subsequent electronic circuit. Nevertheless, it has been shown that the generalized efficiency determined by this way allows one to calibrate the photon fluxes by the analog photodetector without involvement of any reference sources or detectors.

Journal of Applied Physics•May 8, 2025•3 citations

Measuring the absolute photodetection efficiency using photon number correlations

Michael Lindenthal, Johannes Kofler

We present two methods for determining the absolute detection efficiency of photon-counting detectors directly from their singles rates under illumination from a nonclassical light source. One method is based on a continuous variable analog to coincidence counting in discrete photon experiments, but it does not actually rely on high detector time resolutions. The second method is based on difference detection, which is a typical detection scheme in continuous variable quantum optics experiments. Since no coincidence detection is required with either method, they are useful for detection efficiency measurements of photodetectors with detector time resolutions far too low to resolve coincidence events.

arXiv:quant-ph/0606041•Applied Optics•Aug 20, 2006•10 citations

Measurement of quantum efficiency of avalanche photodetectors based on quantum two-photon field

A. Czitrovszky, Alexander V. Sergienko

A pair of rigorously correlated photons (biphotons) from parametric down conversion makes it possible to use them for measuring the absolute value of the quantum efficiency of the optical detectors, both in photon counting and in analog regimes, that does not use any calibrated standard light sources. Using this method the quantum efficiency of EG&G avalanche photodetectors was determined.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Aug 19, 1998•3 citations

Modified Klyshko method for an analog detector calibration

Pavel A. Prudkovskii, Daniil A. Safronenkov, G. Kh. Kitaeva

We extend the absolute quantum efficiency (QE) calibration method, previously used only for photon counters, to analog detectors with a high dispersion of single-photon responses. Our approach is demonstrated on a reference-free measurement of a photomultiplier tube (PMT) cathode QE with 4% relative uncertainty. It involves not only measuring the biphoton field correlation function, but also a special approximation of the distribution of detector readings in order to determine the average number of photoelectrons and the average charge corresponding to a single photoelectron. Results of the calibration were verified in an independent experiment; the numbers of incident photons detected by the PMT and a single-photon counter are in good agreement.

Optics Letters•Sep 2, 2022•10 citations
We realize and test in experiment a method recently proposed for measuring absolute quantum efficiency of analog photodetectors. Similar to the traditional (Klyshko) method of absolute calibration, the new one is based on the direct detection of two-mode squeezed vacuum at the output of a traveling wave OPA. However, in the new method, one measures the difference-photocurrent variance rather than the correlation function of photocurrents (number of coincidences), which makes the technique applicable for high-gain OPA. In this work we test the new method versus the traditional one for the case of photon-counting detectors where both techniques are valid.
Approaches to the standard-free calibration of the quantum efficiency of a wide class of analog detectors based on the measurement of statistical characteristics of fields generated by parametric down-conversion are analyzed. General expressions are obtained for the noise reduction factor of the difference photocurrent and for the covariance of photocurrents in the signal and idler channels that take into account the possibility of strong fluctuations in the amplitudes of single-photon response functions of the detectors used. It is shown that the measurement of the noise reduction factor of the difference photocurrent using detectors that cannot operate in the photon counting mode is on its own insufficient to directly characterize the level of two-mode squeezing in the down-conversion field and to determine the quantum efficiency of photosensitive elements without additional calibration procedures. A method to determine the quantum efficiency of such detectors based on measuring the dependence of the normalized covariance of photocurrents on the parametric gain is proposed.
A modified Klyshko method for standardless calibration of analog photodetector efficiencies is suggested and experimentally implemented. The method is based on the measurement of the number of incident photons using correlation properties of biphotons generated under spontaneous parametric downconversion, and on the determination of the mean number of discrete elementary responses of a photodetector by deconvoluting the statistical distribution of the output photocurrents. As an example, the generalized efficiency of an analog photomultiplier is measured as the ratio of the average number of elementary current responses to the average number of incident photons. The statistical properties of the elementary responses of an analog photodetector depend not only on the detector itself, but also on the parameters of the subsequent electronic circuit. Nevertheless, it has been shown that the generalized efficiency determined by this way allows one to calibrate the photon fluxes by the analog photodetector without involvement of any reference sources or detectors.
We present two methods for determining the absolute detection efficiency of photon-counting detectors directly from their singles rates under illumination from a nonclassical light source. One method is based on a continuous variable analog to coincidence counting in discrete photon experiments, but it does not actually rely on high detector time resolutions. The second method is based on difference detection, which is a typical detection scheme in continuous variable quantum optics experiments. Since no coincidence detection is required with either method, they are useful for detection efficiency measurements of photodetectors with detector time resolutions far too low to resolve coincidence events.
We extend the absolute quantum efficiency (QE) calibration method, previously used only for photon counters, to analog detectors with a high dispersion of single-photon responses. Our approach is demonstrated on a reference-free measurement of a photomultiplier tube (PMT) cathode QE with 4% relative uncertainty. It involves not only measuring the biphoton field correlation function, but also a special approximation of the distribution of detector readings in order to determine the average number of photoelectrons and the average charge corresponding to a single photoelectron. Results of the calibration were verified in an independent experiment; the numbers of incident photons detected by the PMT and a single-photon counter are in good agreement.